| Tabs Page |
|---|
| DescriptionLight sourcesObjectives- 10x/0.45
- 20x/0.75
- 40x/0.6
- 60x/0.95
| Expand |
|---|
| title | Objective specifications |
|---|
| | Position | Name | Brand | Full name | Identifier | Working distance (mm) | Transmittance (% [nm]) | Techniques | Cover glass thickness (mm) |
|---|
| 1 | 10x/0.45 Air | Nikon | 10x/0.45 Air Plan Apo | MRD00101 | 4.0 |
| BF, p-PhC, p-DIC, Fluo | 0.17 |
|---|
| 2 | 20x/0.75 Air | Nikon | 20x/0.75 Air Plan Apo DIC N2 | MRD00201 | 1.0 |
| BF, p-PhC, p-DIC, Fluo | 0.17 |
|---|
| 3 | | Nikon | 40x/0.6 Plan Fluor ELWD | MRH08430 | 2.7-3.7 |
| BF, p-PhC, p-DIC, Fluo | 0.17 |
|---|
| 4 | 60x/0.95 Air | Nikon | 60x/0.95 Air Plan Apo | MRD00600 | 0.15 |
| BF, p-PhC, p-DIC, Fluo | 0.17 |
|---|
BF: Bright-field p-PhC: Pseudo-phase contrast p-DIC: Pseudo-Differetial interference contrast |
FiltersDAPI GFP Cy3 Cy5 - Cy5.5
| Expand |
|---|
| title | Complete specifications |
|---|
| | Position | Name | Brand | Identifier | Polychroic (Transmission) | Emission filter | Effective bandwith (nm) |
|---|
| 1 | DAPI | TBD | TBD | BP 420/20 [430-462] | 455/50 [430-480] | [430-480] |
|---|
| 2 | GFP | TBD | TBD | BP 446/32; 524/42; 600/36; 732/137 | 525/20 [515-535] | [515-535] |
|---|
| 3 | Cy3 | TBD | TBD | BP 446/32; 524/42; 600/36; 732/137 | 605/52 [579-631] | [582-618] |
|---|
| 4 | Cy5 | TBD | TBD | BP 446/32; 524/42; 600/36; 732/137 | 707/72 [671-743] | [671-743] |
|---|
| 5 | Cy5.5 | TBD | TBD | BP 446/32; 524/42; 600/36; 732/137 | 720/60 [690-750] | [690-750] |
|---|
|
Detector| Expand |
|---|
| title | Complete specifications |
|---|
| Camera | PCO Edge 5.5 |
|---|
Sensor Type | sCMOS |
|---|
Sensor Category
| Monochrome |
|---|
Nb Pixels
| 5.5 M |
|---|
Pixel Layout | 2560 x 2160 |
|---|
Pixel size | 6.5 um
|
|---|
Sensor size
| 16.6 mm x 14.0 mm |
|---|
Sensor diameter
| 21.8 mm
|
|---|
Bit depth
| 16-bit |
|---|
| Speed at full resolution | 100 images/s
|
|---|
Max QE
| 60 % at 600 nm |
|---|
| Reading noise | 1.0 e⁻ |
|---|
Cooling
| Forced air +7C |
|---|
Dark Current
| 0.6 e⁻/pixel/sec |
|---|
Full well capacity
| 30 000 e-
|
|---|
Dynamic Range
| 1:30000 |
|---|
Interface
| Dual Camera Link PCIe |
|---|
Mount
| C-mount |
|---|
|
|
| Tabs Page |
|---|
| User Guide| UI Expand |
|---|
| If not already done, turn on the microscope (#1) (the switch is not easily accessible and is located on the back of the right side of the device) If not already done, turn on the computer (#2) and log in to Windows using your UdeM credentials Start INCell Analyzer 6000
|
| UI Expand |
|---|
| During this procedure, you will: Once completed, your sample will be ready for acquisition. | UI Expand |
|---|
| - Within the software, click Eject to open the access door
- Insert your sample in the space provided for this purpose, respecting the orientation indicated
Click Load to close the access doors | Note |
|---|
For use with the 60x objective it is absolutely necessary to use a multi-well plate with a glass bottom. The glass thickness should be 0.17mm. We recommend the following plates: |
|
| UI Expand |
|---|
| When using the instrument for the first time, it is necessary to define the type of plate used. This procedure is usually carried out during the training session but can also be performed if you are using a different multi-well plate. During this procedure, you will: Create a new plate map Measure the A1 well offset Measure and compute the inter-well distance Measure the plate bottom height and thickness
| Tip |
|---|
If you need help, feel free to ask for support. We are always happy to help. |
| UI Expand |
|---|
| title | Create a new plate map |
|---|
| - In the software, select from the menu Application>Plate\Slide Manager...
- Click on the New Plate\Slide icon
- Select the number of wells in your plate (6, 24, 96 etc...)
- Adjust the following settings:
- Name: Your-Name_Your-Plate
- Plate height, bottom thickness and well volume are usually provided in the product technical sheet by the manufacturer
- The material used plastic or glass
- If necessary adjust the number of rows and columns as well as the shape of the well (round or rectangular)
- Click OK
- Close the Plate/Slide Manager window
|
| UI Expand |
|---|
| title | Measure A1 well offset |
|---|
| - In the software, if necessary, click on the Dashboard > Objective Lens> 10x to select the 10x objective
- In the channel list, click the + button and add a transmitted light channel (brightfield)
- To the right of the plate layout, click the Setup Preview Imaging button
- Draw a rectangle around well A1
- Click on preview (on the acquisition panel at the top right of the screen)
- Wait for the preview to be generated
- On the plate map, click on the real center of well A1 to center this well on the objective
In the Dashboard menu select Plate/Slide Click Edit Plate then Define Upper Left Well Click Yes - Check that the yellow circle matches the edges of the well
Click Yes
|
| UI Expand |
|---|
| title | Compute the inter-well distance |
|---|
| - In the software, if necessary, click on the Dashboard > Objective Lens> 10x to select the 10x objective
- In the channel list, click the + button and add a transmitted light channel (brightfield)
- To the right of the plate layout, click the Setup Preview Imaging button
- Draw a rectangle around the bottom right well
- Click on preview (on the acquisition panel at the top right of the screen)
- Wait for the preview to be generated
- On the plate map, click on the real center of bottom right Well to center this well on the objective
In the Dashboard menu select Plate/Slide Click Edit Plate then Define bottom right Well Click Yes - The inter-well distance will then be automatically calculated
- Check that the yellow circles matches the edges of the wells
Click Yes
|
| UI Expand |
|---|
| title | Measure the bottom height and thickness of the plate bottom |
|---|
| This diagram explains how the Laser Autofocus measure and estimate the position of the sample. | View file |
|---|
| name | Plate_Diagram.pdf |
|---|
| height | 250 |
|---|
|
- In the software, if necessary, click on the Dashboard > Objective Lens> 10x to select the 10x objective
- On the plate map, click on the center of a well containing a sample to center this well on the objective
- In the Dashboard menu select Plate/Slide
- Click Verify LAF
- If the 2 detected peaks (blue vertical lines) correspond to the measured peaks (black curve)
- Click Apply Measured Parameters
- Click OK
- Click Yes
- If the 2 detected peaks (blue vertical lines) do not correspond to the measured peaks (black curve)
- Change the bottom thickness value and repeat the operation until the peak are detected successfully
- Close the Laser Autofocus Plate/Slide Verification window
|
|
|
| UI Expand |
|---|
| - Files can be saved temporarily (during acquisition) to local C: drive (desktop)
- At the end of each session, copy your data to your external drive and delete it from local C: drive
- You can store your files on the D: drive (Data Storage). If you do, please create one folder per laboratory using the principal investigator's last name. Inside, create a folder per user using the following nomenclature (First Name_Last Name).
| Note |
|---|
In any case, do not store your files on the C: drive. |
|
| UI Expand |
|---|
| - Click Eject to open the access door
- Collect your samples
- Click Load to close the access door
- Select the 10x objective
- Close the software
- Transfer your data to the D: drive (Data Storage) or to your external drive and delete it from the local C: drive
- Turn off the computer
| Note |
|---|
- Collect your samples, especially those in the microscope
- Leave the microscope and workspace clean
|
|
|
| Tabs Page |
|---|
| Log
| UI Expand |
|---|
| title | 2025-11-23 Migrated to windows 11 |
|---|
| Migrated to windows 11 |
| UI Expand |
|---|
| title | 2022-12 Liquid handling removal |
|---|
| Because of an issue with the liquid handling motor, the liquid handling arm as been replaced by a standard fixed arm for the transmitted light LED. Therefore liquid handling is no longer available. |
| UI Expand |
|---|
| title | 2021-10-26 Y motor not resting |
|---|
| Y motor not resting when reaching the transmitted light position: This time the issue is slightly different: Y motor is homing properly during the initialization procedure but instead of resting it keeps pushing toward the home position and after a brief instant the instrument stops and the red light shows up. If the liquid handler Y motor is disabled the initialization proceed properly and the system is usable at the exception of the brightfield of course |
| UI Expand |
|---|
| title | 2021-09-27 Added to Wiki |
|---|
| Added to wiki |
|
| Tabs Page |
|---|
| Technical DatasheetInstrument Serial W24318-1493815 BK02029 Service Password apiAWL Workstation |
| Tabs Page |
|---|
| title | Troubleshooting & FAQ |
|---|
| Troubleshooting| UI Expand |
|---|
| title | The microscope does not show the usual green light, what should I do? |
|---|
| Usually, this microscope displays a green light when operational and an orange light when in use. A red light ON means the microscope is not functional. In this case please contact the platform manager. |
FAQ| UI Expand |
|---|
| title | Can I use this microscope to observe cells in culture? |
|---|
| | Yes. This microscope is designed for high-throughput acquisition of specimens in multi-well plates. The microscope can control temperature but does not have an environmental chamber.You can also acquire images of specimen mounted between a slide and a coverslip (thickness 0.17mm). |
| UI Expand |
|---|
| title | How to turn off the microscope? |
|---|
| Usually the microscope remains on. However, it is best to turn it off if not in use for a long time (weeks). To do this: - Navigate to the menu Application > Hardware >
- Click Shutdown Instrument
- Wait for the microscope to turn off
- The software closes automatically
|
|
|