Light sourcesEmission (nm) | Nominal power (mW) |
---|
|
Objectives- 4x/0.1 Air
- 10x/0.30 Ph1 Air
- 20x/0.45 Ph1 Air
- 40x/0.95 Air
- 60x/1.4 Ph3 Oil
- 100x/1.45 Oil
Position | Name | Brand | Full name | Identifier | Magnification | Numerical aperture | Immersion | Working distance (mm) | Transmittance (% [nm]) | Applications | Cover glass thickness (mm) |
---|
1 | 4x/0.10 Air | Nikon | 4x/0.1 Plan | MRL00042 | 4x | 0.1 | Air | 30 | >80% [400-900] | BF, Fluo | - | 2 | 10x/0.30 Ph1 Aur | Nikon | 10x/0.30 Plan Fluor Ph1 DL | MRH20101 MRH20105 | 10x | 0.3 | Air | 15.2 | >80% [440-700] | BF, PhC, Fluo | 1.2 ! | 3 | 20x/0.45 Ph1 Air | Nikon | 20x/0.45 Plan Fluor ELWD Ph1 ADL | MRH48230 MRH48220
| 20x | 0.45 | Air | 7.4 | >80% [350-870] | BF, PhC, Fluo | 0-2 | 4 | 40x/0.95 Air | Nikon | 40x/0.95 Plan Apo DIC M/N2 | MRD00400 MRD70470 | 40x | 0.95 | Air | 0.14 | >80% [440-760] | BF, Fluo | 0.11-0.23 | 5 | 60x/1.4 Ph3 Oil | Nikon | 60x/1.4 Plan Apo Ph3 DM | MRD31600 | 60x | 1.4 | Oil | 0.21 | | BF, PhC, Fluo | 0.17 | 6 | 100x/1.45 Oil | Nikon | 100x/1.45 Plan Apo Lambda OFN25 DIC N2 | MRD01905 MRD71970
| 100x | 1.45 | Oil | 0.13 | >80% [460-720] | BF, Fluo | 0.17 | - | 40x/0.60 Ph2 Air | Nikon | Plan Fluor 40x/0.60 Ph2 DIC M ELWD 0-2 | MRH48430 | 40x | 0.6 | Air | 2.7-3.7 | >80% [350-850] | BF, PhC, Fluo | 0-2 | - | 40x/0.60 Air | Nikon | S Plan Fluor 40x/0.6 ELWD DIC N1 | MRH08430 | 40x | 0.6 | Air | 2.8-3.6 | >80% [350-850] | BF, Fluo | 0-2 | - | 40x/0.90 Air | Nikon | S Fluor 40x/0.90 DIC M/N2 | NA
| 40x | 0.9 | Air | 0.3 | NA
| BF, Fluo | 0.11-0.23 | - | 60x/0.70 Air | Nikon | Plan Fluor 60x/0.70 ELWD DIC M/N1 | MRH08630 | 40x | 0.7
| Air | 1.5-2.1 | >80% [380-880] | BF, Fluo | 0.5-1.5 | - | 60xA/1.40 Oil | Nikon | Plan Apo 60xA/1.40 Oil DIC H | MRD71670 | 60x | 1.4 | Oil | 0.21 | >80% [480-700] | BF, Fluo | 0.17 | - | 60x/1.40 Oil | Nikon | Plan Apo VC 60x/1.40 Oil DIC N2 | MRD71670 | 60x | 1.4 | Oil | 0.13 | >80% [480-700] | BF, Fluo | 0.17 | - | 100x/0.85 Air | Nikon | L Plan 100x/0.85 ODN 25 | MUE35900
| 100x | 0.85 | Air | 0.85 | |
| 0-0.7 | - | 100x/1.40 Oil | Nikon | Plan Apo 100x/1.4 Oil DIC H | MRD71970 | 100x | 1.4 | Oil | 0.13 | >80% [480-720] | BF, Fluo | 0.17 |
BF: Bright-field PhC: Phase Contrast DIC: Differential Interference Contrast Fluo: Fluorescence |
FiltersDichroic DAPI\FITC\TRITC Dichroic CFP\YFP\Cy5 Dichroic CFP\YFP\mCherry - DAPI
- Cy3n
- Cy5.5
Also available CFP FITC YFP Cy3.5 mCherry Cy5
Filter cubesExcitation filters
Emission filters
|
Detector |
- If not already done, turn on the computer (#1) and use your UdeM credentials to log in to Windows
- Remove the cover from the microscope
- Turn on the X-Cite exacte light source (#2)
- Turn on the microscope power bar (#3)
- When using the software for the first time, it is necessary to import the microscope-specific settings before launching the acquisition software. Please refer to the First Use section below for instructions.
- Start NIS-Elements
|
When using the acquisition software for the first time, it is necessary to import the microscope-specific parameters into the software. This procedure is usually carried out during the training session. However, it is also possible to use it to reset the software if it is not displayed correctly, for example. Please note, this procedure will delete all your experiment protocols and restore the software to its original settings. |
- If open, close NIS-Elements and wait until it is completely closed
- On the Desktop open the folder Softwares
- Open NIS Settings Utility
- Click on the Import tab
- Click on Browse
- Navigate to your Documents
- Select the file NIS Settings for Nikon TE2000E.bin
- Click Select
- Select all items
- Click Import
- Click OK
- Close NIS Settings Utility
- You can then open NIS-Elements
|
This procedure puts the microscope in a safe configuration to load your sample. At the end the microscope will be ready for acquisition. - If not already done, in NIS-Elements select 4x to select the 4x objective
The 4x objective is the safest because it has the longest working distance (30mm). The sample will appear perfectly sharp long before the lens approaches it. It is recommended to always first focus with the safest lens. The objectives are parafocal, focusing with the safest objective will then allow you to easily find your sample with another objective. The 10x objective is also safe because its working distance is 15.2 mm. |
- If not done already, press Escape to lower the objective to the lowest position
- Place the test slide on the microscope stage with the coverslip toward the objective
Always use the test slide to perform the first focus. |
- If necessary, move the stage so that the sample is centered on the objective
- In NIS-Elements, select the optical configuration BF vis or the desired fluorescence (DAPI vis, ...) to activate the configuration
- Adjust the focus with the main dial while looking through the eyepieces until the image is perfectly sharp
The focus can be found at Z = mm). The Z value can be found in the NIS Software XYZ Position tab. |
- In NIS-Elements, select the optical configuration Off to turn off the illumination
|
First focus with the safest objective before selecting another lens and continuing with secondary focus. |
- In NIS-Elements, click 10x, 20x or 40x to select the desired objective
The 40x objective is the best Air objective because it has the greatest number of optical corrections (Plan Apochromat) and the largest numerical aperture (0.95). |
- In NIS-Elements, select the optical configuration BF vis or the desired fluorescence (DAPI vis, ...) to activate the configuration
- Adjust the focus with the precision dial while looking through the eyepieces until the image is perfectly sharp
- Select the optical configuration Off to turn off the illumination
- Your sample is ready for acquisition!
|
After performing the first focus, in NIS-Elements:
- In NIS, click Escape, to lower the objectives
Click 60x Oil or 100x Oil to select the desired objective. The 100x objective has the greatest resolution (NA 1.45 vs 1.4) |
- Place a single drop of oil on the objective
- In NIS-Elements, click again Escape to return the objective to its original position
- Select the optical configuration BF vis or the desired fluorescence (DAPI vis, ...) to activate the configuration
- Adjust the focus with the precision dial while looking through the eyepieces until the image is perfectly sharp
- Select the optical configuration Off to turn off the illumination
- Your sample is ready for acquisition!
|
|
|
- Files can be saved temporarily (during acquisition) to local C: drive (desktop)
- At the end of each session, copy your data to your external drive and delete it from local C: drive
- You can store your files on the D: drive (Data Storage). If you do, please create one folder per laboratory using the principal investigator's last name. Inside, create a folder per user using the following nomenclature (First Name_Last Name).
In any case, do not store your files on the C: drive. |
|
- Save your data
- If the oil objective was used, clean it with lens cleaner and lens paper (not Kimwipes)
- Select the lowest magnification objective and press Escape to place the objectives in a safe position
- Close NIS-Elements
- Transfer your data to the D: drive (Data Storage) or to your external drive and delete it from the local C: drive
- Turn off the computer
- Turn off the X-Cite exacte light (#2)
- Turn off the microscope power bar (#3)
- Cover the microscope
- Collect your samples, especially those in the microscope
- Leave the microscope and workspace clean
|
|
|
- Complete installation
- Complete cleaning
|
|
Stand- Nikon TE2000-E inverted
- Nikon T-HUBC Serial 750110
- Nikon PFS T-PFS 810060
- Nikon T-RCP Serial 750104
Light sources- Transmitted Halogen light Nikon TE2-PS100W Serial 505320 12V
- GIF Filter
- Diffusion
- Infra-red
- Neutral color
- Transmitted light shutter uniblitz VCM D1 Serial A607295
- X-Cite exacte Model XCT10A 200 W Seria;l XCT10A-0337
- EXFO Collimator 810-00030
- Prior Shutter HF201HT Serial 7332
- Prior 8 position motorized excitation filter wheel 61817
- Nikon Manual Neutral density filters ND8 ND4
- Manual Fluorescence field diaphragm
- Manual Fluorescence shutter
CondenserObjectivesStage- Prior Proscan 2 H30V4 Serial 62040
- Remote control joystick Prior CS152v2 Serial R0772128
- Inserts
- 1 slide
- Multi-well plate
- Adapter for 3cm dish
Filters Detector- Hamamtsu ORCA Flash V2 CMOS Model C11440-22CU Serial 002162 V2
- Monochrome Camera 2048 x 2048 pixels
- Pixel 6.5 um x 6.5 um
- Force air cooled at -10C
- Effective sensor area 13.312 mm (H) × 13.312 mm (V
- Dark current 0.06 electron/pixel/s
- Reading noise 1.6 electrons (r.m.s) at standard scan
- Dynamic range 1 : 18 000
- Spectral response >40% between 400-850nm. Max 80% at 550nm

Workstation- HP Z800 Workstation
- 2 x Intel Xeon X5650 @ 2.67GHz
- RAM 24 GB DDR3 666 MHz (6 x 4 GB)
- OS 500 GB SSD 530 MB/s
- 3 TB HD Data Storage (2 TB + 1 TB spanned volume) 130 MB/s
- Video Card AMD FirePro V5800 1 GB dedicated memory
- Monitor Dell 2407EFP-HC 1900x1200 60Hz
- Software NIS-Elements AR v4.20.02 Build 988 HASP ID 4683DFE6
- Core NIS-AR 2.3x STANDARD SUA Activation date 2010-07-07 SUA paid 2x SUA Expiration date 2013-08-07
- Modules SRM ,Filter Wheel, ND (6 dimensions), Shutter, Stage XY axis
Consumables- Liquid Light Guide
- Excite exacte bulb
- 12V halogen bulb
|
TroubleshootingThis can happen when the acquisition software is turned on before the camera has fully initialized. - Turn off NIS-Elements
- Turn off the camera
- Wait few seconds
- Turn the camera back on
- Wait until the camera has fully initialize (light at the back of the camera no longer blinks)
- Start NIS-Elements
|
FAQYes. This is an inverted microscope designed to look at specimen in a dish or a multi-well plate. The objectives are optimized to image through thin glass bottom multi-well plates. You may also image specimen mounted between a slide and a 0.17mm thick coverslip. For long timelapse, be aware of photo-toxicity. |
|
|