| Nikon Eclipse E-600 upright microscopeRoger Gaudry Building, Room N-620 Simple Microscope usage price
Applications BrightfieldBright-field Phase Contrast Polarized light Fluorescence - Color camera
Light sources
Objectives Objectives:10x/0.3 Air Ph1 WD 16.0 - 20x/0.5 Air DIC 2.1
40x/0.75 Air Ph2 WD 0.72 - 60x/0.85 Air WD 0.3
- 100x/1.3 Oil Ph3 WD 0.2
Développer |
---|
title | Detailed Objectives complete specifications |
---|
|
Position | Name | Brand | ID | MagnificationFull name | Numerical Aperture | Immersion | Type | Working distance (mm) | Transmittance (% [nm]) | Technique | Cover glass thickness (mm) |
---|
1 | 4x/0.13 Air | Nikon | 4x/0.3 Air Plan Fluor | 0.13 | Air | Plan Fluor | 17.1 |
| BF, Fluo | - | 2 | 10x/0.3 Air | Nikon | 10x/0.3 Air Plan Fluor Ph1 DLL | 10x | 0.3 | Air | Plan Fluor | 16 | >75% [400-800] Max % @500nm | BF, Pol, PhC, Fluo | 0.17 | 3 | | Nikon | 20x/0.5 Air Plan Fluor DIC M | 0.5 | | Plan Fluor | 2.1 |
| BF, Fluo | 0.17 | 4 | 40x/0.75 Air | Nikon | 40x/0.75 Air Plan Fluor Ph2 DLL | 40x | 0.75 | Air | Plan Fluor | 0.72 | >80% [400-750] Max % @500nm | BF, Pol, PhC, Fluo | 0.17 | | 5 | 60x/0.85 Air | Nikon | 60x/0.85 Air Plan Fluor | 0.85 | Air | Plan Fluor | 0.30 |
| BF, Fluo | Ajustable 0.11-0.23 | 63 | | Nikon | 100x/1.3 Oil Plan Fluor Ph3 DLL | 100x | 1.3 | | Plan Fluor | 0.2 | >75% [400-800] Max % @500nm
| BF, Pol, PhC, Fluo | 0.17 | 4 | Empty | 5 | Empty | 6 | Empty |
|
Filter cubes DAPI/Hoechst/AMCA FITC/EGFP TRITC/Rhodamine TexasRed - Empty
- Empty
Développer |
---|
title | Detailed Filters complete specifications |
---|
|
|
|
Tabs Page |
---|
id | User Guide |
---|
title | User Guide |
---|
|
UI Expand |
---|
expanded | true |
---|
title | StartupStart-up |
---|
| - Turn on the computercomputer (#1)
- Turn on the microscope power barbar (#2)
If fluorescence is required, press the mode button on the lumencor remote - Mercury lamp must be on for at least 30 min before being turned off and vice-versa
If transmitted light is required, turn turn on the mercury lamp power supply unit (3A) and press ignition (3B) Avertissement |
---|
halogen lamp on the right side of the microscope (#3) Log in Windows using your UdM credentials - Start-up NIS-Elements
Info |
---|
The first time you log in the computeruse the instrument, you need to import the microscope settings into the software. To do this follow the instructions Nikon Eclipse E600-enSoftware setup. |
|
UI Expand |
---|
| - Save your data
- Close NIS-Elements
- Transfer your data to the D: drive (Data Storage) or to your external drive and delete it from the local C: driveGet your samples from the microscope
- Turn off the computer
- If oil objective was used, clean the oil objective it with lens cleaner and lens paperIf fluorescence was used, turn off the mercury lamp power supply unit (3A (not Kimwipes)
- Turn off the microscope power bar (2#2)
- Wait until the lamps are cool and cover the Cover the microscope
Remarque |
---|
| - Take back your samples including ones in the microscope
- Leave the microscope and the working area cleanMercury lamp must be on for at least 30 min before being turned off and vice-versa
|
|
UI Expand |
---|
title | Storage Managementmanagement |
---|
| - Files can be saved temporarily (during acquisition) on the local C: drive (desktop)
- At the end of each session, copy your data to your external drive and delete it from the local C: drive
- You can store your files on the D: drive (Data Storage). If you do, please create a folder per laboratory using the principal investigator last name. Within, create one folder per user (Firstname_Lastname).
Remarque |
---|
In any case, your files should be removed from the C: drive. |
|
UI Expand |
---|
title | Setting-up NIS-Elements | Software setup |
---|
| The first time you use the instrument, you need to import the microscope settings into the software Ancre |
---|
| Setting-up NIS-Elements | Setting-up NIS-Elements | This process is required the first time you are using the instrument. You will usually do it this during the training session. It This procedure can also be performed if something is not working properly right or and if you want to refresh reset the software interfaceto its original settings. Remarque |
---|
This process will delete all experiment protocols and restore the parameters for the microscopereset the software to the original settings for this specific microscope. |
- If open, close NIS-Elements and wait until it is completely closed (up to 30 seconds)
- On your Desktop open the Softwares folder
- Open
- Close NIS-Elements
- Wait until the complete closure of NIS-Elements
- Open the folder Desktop\Logiciels
- Open the software NIS Settings Utility
- Click on the Import tab
- Click on Browse
- Navigate to your desktop Desktop
- Select the file Nikon-E600 Settings.bin
- Click Select
- Select all items
- Click Import
- Click OK
- Close the NIS Settings
- Open You can now reopen NIS-Elements
|
Tabs Page |
---|
id | Lightpath |
---|
title | Lightpath |
---|
|
Follow the schematics below for Transmitted light (Bright The following schematics depict the light path for transmitted (bright-field, Phase Contrast, Polarized light) and transmitted light reflected (fluorescence) lights.view-file PDF |
---|
name | LightPath_Nikon E600.pdf |
---|
|
| height | 400
---|
Tabs Page |
---|
| Available manuals _-_._-__._-__._-__.__.__. |
Tabs Page |
---|
|
UI Expand |
---|
| - Adjust focus drive jitter
- Add a holder for polarized light analyzer
- Bring new test slides
|
UI Expand |
---|
expanded | true |
---|
title | 2022-09-26 |
---|
| - Added Lumencor SOLA v-nIR
- Added Fluorescent light collimator
|
UI Expand |
---|
expanded | true |
---|
title | 2022-09-26 |
---|
| - Added 4x/0.13 Air
- Added 20x/0.5 Air
- Added 60x/0.85 Air
- Updated E600 settings file
|
UI Expand |
---|
| - Replacement mercury lamp bulb HBO 1003W/2
- 256 GB SSD added in workstation for OS
- Windows 10 Installation
- BIOS updated to v2.47
- Camera Firmware updated to v2.11
- NIS-Elements Basic Research v4.6 64-bits installed
- Creation NIS Elements Settings
- FITC and TxRed cubes excitation and emission filters slightly damaged. Dichroic OK
|
UI Expand |
---|
| - Nikon Preventive Maintenance
- Adjustment and cleaning Stage Nosepiece, condenser, filter turret, focus drive, shutters
- 100x slightly damaged but not the lens
- Fluorescence cubes damaged: FITC
| TxRed- , TexasRed
- Focus drive has a
| jutter
|
Tabs Page |
---|
id | Technical Datasheet |
---|
title | Technical Datasheet |
---|
| Stand E600 - E600W upright Serial 725540
Light sources- Transmitted light Halogen 12V 100W
Serial - Model C-LPSerial 01875599
- Reflected light
Power supply Nikon Mercury lamp Type C SHG1 Serial D12139 Bulb housing model LH M100C-1 Serial 039326Condenser- Condenser Universal C-CU Serial 081901
- Lens Dry NA 0.9
- Filters: Empty, Ph1, Ph2, Ph3, DICM, DICH, Empty
Objectives- 4x/0.13
- 10x/0.3 Plan Fluor Ph1 DLL WD 16 Air
- 20x/
1- 40x/0.75 Plan Fluor Ph2 DLL WD 0.72 Air
(Transmission curve) - with PF40 Wollaston prism
- 60x/0.85
- 100x/1.3 Plan Fluor Ph3 DLL WD 0.2 Oil
(Transmission curve) - with PF/PA 100 Oil Wollaston prism
EmptyStageEmpty Empty Turret | Position | Cube name | Brand | ID | Excitation Filter | Dichroic mirror | Emission Filter |
---|
Filters1 | Chroma | Cube set 31000v2 | 350/50x400LP | 460/50m | 2 | Chroma | Cube set 41001480/40x | 505LP | 535/50m | 3 | Chroma | Cube set 41002c | 545/30x | 570LP | 620/60m | 4 | Texas Red | Chroma | Cube set 41004 | 560/55x | 595LP | 645/75m | Detector5 | Empty | 6 | Empty | Detector - Camera Nikon DS-Ri2 Serial 701234
Workstation- HP Z440 Workstation
- Intel Xeon E5-1630 v3 @ 3.7GHz
- RAM 32 GB DDR4 2133 MHz (4 x 8 GB)
- OS
256GB 2TB - 2 TB HD Data Storage 150 MBs
- Video Card NVIDIA Quadro K620 2 GB DDR3 dedicated memory
- Monitor HP Z24i display 24'
1920x1200ConsumablesMercury lamp bulb OSRAM HBO 1003W/2 100W 12V
Tabs Page |
---|
id | FAQ |
---|
title | Troubleshooting & FAQ |
---|
| Troubleshooting UI Expand |
---|
title | NIS-Elements shows an error: Camera Driver... |
---|
| This happens when NIS-Elements does not find the camera. It is usually because the camera is not powered on. - Turn off NIS-Elements
- Turn on the camera (both the microscope power bar and the camera need to be ON
(the camera is usually ON, the switch on the top of the camera) - Check the camera is correctly connected to the computer via a USB cable
- Turn on NIS-Elements
|
FAQ UI Expand |
---|
title | Can I use this microscope to look at cells in a dish? |
---|
| No. This is an upright microscope designed to observe specimen mounted between a slide and a 0.17 mm thick coverslip. |
|
|