Nikon Eclipse E-600 upright microscope

Roger Gaudry Building, Room N-620
Simple Microscope usage price

  • Applications

    • BrightfieldBright-field

    • Phase Contrast

    • Polarized light

    • Fluorescence

    • Color camera
  • Light sources

    • Halogen lamp for 30 W transmitted light

    • Mercury lamp (350~600nm) for fluorescence 

    • Lumencor SOLA V-nIR

Développer
titleLumencor SOLA V-nIR complete specifications


Emission peak (nm)Power (mW)
39540
43035
47515
51020
55020
58020
64020
73010

Lumencor SOLA manual.pdf
Lumencor SOLA datasheet.pdf
Lumencor SOLA recommended filter sets.pdf

  • Objectives

    • 4x/0.13 Air WD 17.1

    Objectives:
    • 10x/0.3 Air Ph1 WD 16.0

    • 20x/0.5 Air DIC 2.1
    • 40x/0.75 Air Ph2 WD 0.72

    • 60x/0.85 Air WD 0.3
    • 100x/1.3 Oil Ph3 WD 0.2
Développer
titleDetailed Objectives complete specifications


PositionNameBrandIDMagnificationFull nameNumerical ApertureImmersionTypeWorking distance (mm)Transmittance
(% [nm])
TechniqueCover glass thickness (mm)
14x/0.13 AirNikon4x/0.3 Air Plan Fluor0.13AirPlan Fluor17.1


BF, Fluo-
210x/0.3 AirNikon10x/0.3 Air Plan Fluor Ph1 DLL10x0.3AirPlan Fluor16

>75% [400-800]
Max % @500nm

BF, Pol, PhC, Fluo0.17
3

20x/0.5

Nikon20x/0.5 Air Plan Fluor DIC M0.5

Air

Plan Fluor2.1
BF, Fluo0.17
4

40x/0.75 Air

Nikon40x/0.75 Air Plan Fluor Ph2 DLL40x

0.75

Air

Plan Fluor0.72>80% [400-750]
Max % @500nm
BF, Pol, PhC, Fluo0.17
560x/0.85 AirNikon60x/0.85 Air Plan Fluor0.85AirPlan Fluor0.30
BF, FluoAjustable 0.11-0.23
63

100x/1.3 Oil Oil

Nikon100x/1.3 Oil Plan Fluor Ph3 DLL

100x

1.3
Remarque
Oil

 

Plan Fluor0.2>75% [400-800]
Max % @500nm
BF, Pol, PhC, Fluo0.174Empty5Empty6Empty


  • Filter cubes

    • DAPI/Hoechst/AMCA

    • FITC/EGFP

    • TRITC/Rhodamine

    • TexasRed

    • Empty
    • Empty
Développer
titleDetailed Filters complete specifications


PositionCube nameNameBrandIDExcitation FilterDichroic mirrorEmission FilterComments
1DAPI/Hoechst/AMCAChromaCube set 31000v2350/50x 
[325-375]
400LP460/50m 
[435-485]

2FITC/EGFPChromaCube set 41001

480/40x 
[420-500]

505LP

535/50m 
[510-560]

3TRITC/RhodamineChroma

Cube set 41002c

545/30x 
[530-560]
570LP620/60m 
[590-650]

4Texas RedChroma

Cube set 41004

560/55x 
[532-588]

595LP645/75m 
[607-682]

5Empty





6Empty






  • Detector

    • Color CMOS Nikon DS-Ri2 4908 x 3264 pixels,14-bit, 6fps 6 images/s at full frame


Tabs Page
idUser Guide
titleUser Guide


UI Expand
expandedtrue
titleStartupStart-up
  1. Turn on the computercomputer (#1)
  2. Turn on the microscope power barbar (#2)
  3. If fluorescence is required, press the mode button on the lumencor remote

  4. If transmitted light is required, turn  turn on the mercury lamp power supply unit (3A) and press ignition (3B)

    Avertissement

    Mercury lamp must be on for at least 30 min before being turned off and vice-versa

    halogen lamp on the right side of the microscope (#3)

  5. Log in Windows using your UdM credentials

  6. Start-up NIS-Elements
Info

The first time you log in the computeruse the instrument, you need to import the microscope settings into the software. To do this follow the instructions Nikon Eclipse E600-enSoftware setup.



UI Expand
titleShutdown
  1. Save your data
  2. Close NIS-Elements
  3. Transfer your data to the D: drive (Data Storage) or to your external drive and delete it from the local C: driveGet your samples from the microscope
  4. Turn off the computer
  5. If oil objective was used, clean the oil objective it with lens cleaner and lens paperIf fluorescence was used, turn off the mercury lamp power supply unit (3A (not Kimwipes)
  6. Turn off the microscope power bar (2#2)
  7. Wait until the lamps are cool and cover the Cover the microscope
Remarque
titleImportant Reminders
  • Take back your samples including ones in the microscope
  • Leave the microscope and the working area clean
  • Mercury lamp must be on for at least 30 min before being turned off and vice-versa



UI Expand
titleStorage Managementmanagement
  • Files can be saved temporarily (during acquisition) on the local C: drive (desktop)
  • At the end of each session, copy your data to your external drive and delete it from the local C: drive
  • You can store your files on the D: drive (Data Storage). If you do, please create a folder per laboratory using the principal investigator last name. Within, create one folder per user (Firstname_Lastname).
Remarque

In any case, your files should be removed from the C: drive.



This process is required the first time you are using the instrument

. You will usually do

it

this during the training session.

It


This procedure can also be performed if something is not working properly

right or

and if you want to

refresh

reset the software

interface

to its original settings.

UI Expand
titleSetting-up NIS-ElementsSoftware setup

The first time you use the instrument, you need to import the microscope settings into the software

Ancre
Setting-up NIS-ElementsSetting-up NIS-Elements
Remarque

This process will delete all experiment protocols and restore the parameters for the microscopereset the software to the original settings for this specific microscope.

  1. If open, close NIS-Elements and wait until it is completely closed (up to 30 seconds)
  2. On your Desktop open the Softwares folder
  3. Open 
  4. Close NIS-Elements
  5. Wait until the complete closure of NIS-Elements
  6. Open the folder Desktop\Logiciels
  7. Open the software NIS Settings Utility
  8. Click on the Import tab
  9. Click on Browse
  10. Navigate to your desktop Desktop
  11. Select the file Nikon-E600 Settings.bin
  12. Click Select
  13. Select all items
  14. Click Import
  15. Click OK
  16. Close the NIS Settings
  17. Open You can now reopen NIS-Elements




Tabs Page
idLightpath
titleLightpath


Follow the schematics below for Transmitted light (Bright

The following schematics depict the light path for transmitted (bright-field, Phase Contrast, Polarized light) and

transmitted light

reflected (fluorescence) lights.

view-file

PDF
nameLightPath_Nikon E600.pdf
height
400


Tabs Page
idManuals
titleManuals


Available manuals


Tabs Page
idLog
titleLog


UI Expand
expandedtrue
titleTo do
  • Adjust focus drive jitter
  • Add a holder for polarized light analyzer
  • Bring new test slides


UI Expand
expandedtrue
title2018-04-042022-09-26
  • Added Lumencor SOLA v-nIR
  • Added Fluorescent light collimator


UI Expand
expandedtrue
title2022-09-26
  • Added 4x/0.13 Air
  • Added 20x/0.5 Air
  • Added 60x/0.85 Air
  • Updated E600 settings file
  • Nikon Preventive Maintenance
  • Adjustment and cleaning Stage Nosepiece, condenser, filter turret, focus drive, shutters
  • 100x slightly damaged but not the lens
  • Fluorescence cubes damaged: FITC TexasRed
  • Focus drive has a jitter


UI Expand
title2021-07-20
  • Replacement mercury lamp bulb HBO 1003W/2
  • 256 GB SSD added in workstation for OS
  • Windows 10 Installation
  • BIOS updated to v2.47
  • Camera Firmware updated to v2.11
  • NIS-Elements Basic Research v4.6 64-bits installed
  • Creation NIS Elements Settings
  • FITC and TxRed cubes excitation and emission filters slightly damaged. Dichroic OK


UI Expand
title2018-04-04
  • Nikon Preventive Maintenance
  • Adjustment and cleaning Stage Nosepiece, condenser, filter turret, focus drive, shutters
  • 100x slightly damaged but not the lens
  • Fluorescence cubes damaged: FITC, TexasRed
  • Focus drive has a jitter




  • Manual Stage

Filters

Tabs Page
idTechnical Datasheet
titleTechnical Datasheet

Stand

  • Nikon Eclipse E600 E600W upright Serial 725540

Light sources

  • Transmitted light Halogen 12V 100W Serial Model C-LPSerial 01875599
  • Reflected light

Condenser

  • Condenser Universal C-CU Serial 081901
  • Lens Dry NA 0.9
  • Filters: Empty, Ph1, Ph2, Ph3, DICM, DICH, Empty

Objectives

  • 4x/0.13
  • 10x/0.3 Plan Fluor Ph1 DLL WD 16 Air
  • 20x/
  • 40x/0.75 Plan Fluor Ph2 DLL WD 0.72 Air with PF40 Wollaston prism
  • 60x/0.85
  • 100x/1.3 Plan Fluor Ph3 DLL WD 0.2 Oil with PF/PA 100 Oil Wollaston prismEmpty

Stage

  • Empty
  • Empty
  • Développer
    titleDetailed specifications
    PositionNameBrandIDMagnificationNumerical ApertureImmersionTypeWorking distance (mm)Transmittance
    (% [nm])
    TechniqueCover glass thickness (mm)
    110x/0.3 AirNikon10x/0.3 Air Plan Fluor Ph1 DLL10x0.3AirPlan Fluor16>75% [400-800]BF, Pol, PhC, Fluo0.17
    2

    40x/0.75 Air

    Nikon40x/0.75 Air Plan Fluor Ph2 DLL40x

    0.75

    Air

    Plan Fluor0.72>80% [400-750]BF, Pol, PhC, Fluo0.17
    3

    100x/1.3 Oil

    Nikon100x/1.3 Oil Plan Fluor Ph3 DLL

    100x

    1.3
    Remarque
    Oil
    Plan Fluor0.2>75% [400-800]BF, Pol, PhC, Fluo0.174Empty5Empty6Empty

    Turret

    • DAPI/Hoechst/AMCA

    • FITC/EGFP

    • TRITC/Rhodamine

    • TexasRed

    Développer
    titleDetailed specifications
    PositionCube nameBrandIDExcitation FilterDichroic mirrorEmission Filter1DAPI/Hoechst/AMCAChromaCube set 31000v2350/50x [325-375]400LP460/50m [435-485]2FITC/EGFPChromaCube set 41001

    480/40x [420-500]

    505LP

    535/50m [510-560]3TRITC/RhodamineChroma

    Cube set 41002c

    545/30x [530-560]570LP620/60m [590-650]
    4Texas RedChroma

    Cube set 41004

    560/55x [532-588]

    595LP645/75m [607-682]5Empty6Empty

    Detector

    • Color Camera Nikon DS-Ri2 Serial 701234

    Workstation

    • HP Z440 Workstation
    • Intel Xeon E5-1630 v3 @ 3.7GHz
    • RAM 32 GB DDR4 2133 MHz  (4 x 8 GB)
    • OS 256GB 256 GB SSD 550 MBs
    • 2TB 2 TB HD Data Storage 150 MBs
    • Video Card NVIDIA Quadro K620 2 GB DDR3 dedicated memory
    • Monitor HP Z24i display 24' 1920x12001920 x 1200

    Consumables

    Mercury lamp bulb OSRAM HBO 1003W/2 100W


    Tabs Page
    idFAQ
    titleTroubleshooting & FAQ

    Troubleshooting

    UI Expand
    titleNIS-Elements shows an error: Camera Driver...

    This happens when NIS-Elements does not find the camera. It is usually because the camera is not powered on.

    • Turn off NIS-Elements
    • Turn on the camera (both the microscope power bar and the camera need to be ON
      (the camera is usually ON, the switch on the top of the camera)
    • Check the camera is correctly connected to the computer via a USB cable
    • Turn on NIS-Elements

    FAQ

    UI Expand
    titleCan I use this microscope to look at cells in a dish?

    No. This is an upright microscope designed to observe specimen mounted between a slide and a 0.17 mm thick coverslip.





    Nikon Eclipse E-600Nikon Eclipse E600 MicroscopeImage Removed
    Tabs Container
    idDemo Image
    Tabs Container
    idInstrument Tabs
    titleNikon Eclipse E600
    directionhorizontal
    Tabs Page
    id
    titleNikon Eclipse E-600
    direction
    horizontal


    Image Removed
    Tabs Page
    idDemo Image
    titleDemo Image


    Image Removed

    Image Added

    Cells courtesy of Benoit Bessette and Monique Vasseur (Biochemistry)
    Images courtesy of Dr Shirley Campbell and Emilie Fiola-Masson (Pharmacology)

    Tabs Page
    idNikon Eclipse E-600 at Roger Gaudry N-620
    titleN-620




    Inclure page
    Plateformes:_foot-en
    Plateformes:_foot-en