urlhttps://wiki.umontreal.ca/display/Microscopie/Nikon+Eclipse+E600


Tabs Container
idInstrument Tabs
titleNikon Eclipse E600
directionhorizontal


Tabs Page
idDescription
titleDescription

Nikon Eclipse E-600 upright microscope

Roger Gaudry Building, Room N-620
Simple Microscope usage price

  • Applications

    • Bright-field

    • Phase Contrast

    • Polarized light

    • Fluorescence

    • Color camera
  • Light sources

    • Halogen lamp for 30 W transmitted lightMercury lamp 100 W (350~600nm) for fluorescence 

    • Lumencor SOLA V-nIR

Développer
titleMercury lamp Lumencor SOLA V-nIR complete specifications


36545405344364354637579
Peak emission Emission peak (nm)Power (mW)
39540
334430735
47515
51020
55020
58020
64026

HBO Mercury lamp emission spectra (Source)
Comparison HBO Mercury vs XCite Metal Halide Lamps (Source)
OSRAM HBO 100W/2 Mercury lamp manual (pdf)

20
73010

Lumencor SOLA manual.pdf
Lumencor SOLA datasheet.pdf
Lumencor SOLA recommended filter sets.pdf

  • Objectives

    Objectives

    • 4x/0.13 Air WD 17.1

    • 10x/0.3 Air Ph1 WD 16.0

    • 20x/0.5 Air DIC 2.1
    • 40x/0.75 Air Ph2 WD Ph2 WD 0.72

    • 60x/0.85 Air WD 0.753
    • 100x/1.3 Oil Ph3 WD 0.2
Développer
titleObjectives complete specifications


PositionNameBrandFull nameIDMagnificationNumerical ApertureImmersionTypeWorking distance (mm)Transmittance
(% [nm])
TechniqueCover glass thickness (mm)
14x/0.13 AirNikon4x/0.3 Air Plan Fluor0.13AirPlan Fluor17.1


BF, Fluo-
210x/0.3 AirNikon10x/0.3 Air Plan Fluor Ph1 DLL10x0.3AirPlan Fluor16

>75% [400-800]
Max % @500nm

BF, Pol, PhC, Fluo0.17
3

20x/0.5

Nikon20x/0.5 Air Plan Fluor DIC M0.5

Air

Plan Fluor2.1
BF, Fluo0.17
4

40x/0.75 Air

Nikon40x/0.75 Air Plan Fluor Ph2 DLL40x

0.75

Air

Plan Fluor0.72>80% [400-750]
Max % @500nm
BF, Pol, PhC, Fluo0.17
560x/0.85 AirNikon60x/0.85 Air Plan Fluor0.85AirPlan Fluor0.30
BF, FluoAjustable 0.11-0.23
63

100x/1.3 Oil Oil

Nikon100x/1.3 Oil Plan Fluor Ph3 DLL

100x

1.3
Remarque
Oil

 

Plan Fluor0.2>75% [400-800]
Max % @500nm
BF, Pol, PhC, Fluo0.174Empty5Empty6Empty


  • Filter cubes

    • DAPI/Hoechst/AMCA

    • FITC/EGFP

    • TRITC/Rhodamine

    • TexasRed

    • Empty
    • Empty
Développer
titleFilters complete specifications


PositionNameBrandIDExcitation FilterDichroic mirrorEmission FilterComments
1DAPI/Hoechst/AMCAChromaCube set 31000v2350/50x
[325-375]
400LP460/50m
[435-485]

2FITC/EGFPChromaCube set 41001

480/40x
[420-500]

505LP

535/50m
[510-560]

3TRITC/RhodamineChroma

Cube set 41002c

545/30x
[530-560]
570LP620/60m
[590-650]

4Texas RedChroma

Cube set 41004

560/55x
[532-588]

595LP645/75m
[607-682]

5Empty





6Empty






  • Detector

    • Color CMOS Nikon DS-Ri2 4908 x 3264 pixels,14-bit, 6 images/s at full frame


Tabs Page
idUser Guide
titleUser Guide


Mercury lamp must be on for at least 30 min before being turned off and vice-versa

press the mode button on the lumencor remote

  • If transmitted light is required, turn on the halogen lamp on the right side of the microscope (#4#3)

  • Log in Windows using your UdM credentials

  • Start-up NIS-Elements
  • UI Expand
    expandedtrue
    titleStart-up
    1. Turn on the computer (#1)
    2. Turn on the microscope power bar (#2)
    3. If fluorescence is required, turn on the mercury lamp power supply unit (#3A) and press ignition (#3B)

    Avertissement
    Info

    The first time you use the instrument, you need to import the microscope settings into the software. To do this follow the instructions Software setup.



    UI Expand
    titleShutdown
    1. Save your data
    2. Close NIS-Elements
    3. Transfer your data to the D: drive (Data Storage) or to your external drive and delete it from the local C: drive
    4. Turn off the computer
    5. If oil objective was used, clean it with lens cleaner and lens paper (not Kimwipes)
    6. If fluorescence was used, turn off the mercury lamp power supply unit (#3A)
    7. If transmitted light was used, turn off the halogen lamp on the right side of the microscope (#4)
    8. Turn off the microscope power bar (#2)
    9. Wait until the lamps have cooled and cover Cover the microscope
    Remarque
    titleImportant Reminders
    • Take back your samples including ones in the microscope
    • Leave the microscope and the working area clean
    • Mercury lamp must be on for at least 30 min before being turned off and vice-versa



    UI Expand
    titleStorage management
    • Files can be saved temporarily (during acquisition) on the local C: drive (desktop)
    • At the end of each session, copy your data to your external drive and delete it from the local C: drive
    • You can store your files on the D: drive (Data Storage). If you do, please create a folder per laboratory using the principal investigator last name. Within, create one folder per user (Firstname_Lastname).
    Remarque

    In any case, your files should be removed from the C: drive.



    UI Expand
    titleSoftware setup

    The first time you use the instrument, you need to import the microscope settings into the software. You will usually do this during the training session.
    This procedure can also be performed if something is not working properly and if you want to reset the software to its original settings.

    Remarque

    This process will delete all experiment protocols and reset the software to the original settings for this specific microscope.

    1. If open, close NIS-Elements and wait until it is completely closed (up to 30 seconds)
    2. On your Desktop open the Softwares folder
    3. Open NIS Settings Utility
    4. Click on the Import tab
    5. Click on Browse
    6. Navigate to your Desktop
    7. Select the file Nikon-E600 Settings.bin
    8. Click Select
    9. Select all items
    10. Click Import
    11. Click OK
    12. Close the NIS Settings
    13. You can now reopen NIS-Elements



    Tabs Page
    idLightpath
    titleLightpath


    The following schematics depict the light path for transmitted (bright-field, Phase Contrast, Polarized light) and reflected (fluorescence) lights
    .

    PDF
    nameLightPath_Nikon E600
    Lightpath (
    .pdf
    )


    Tabs Page
    idManuals
    titleManuals


    Available manuals


    Tabs Page
    idLog
    titleLog


    UI Expand
    expandedtrue
    titleTo do
    • Adjust focus drive jitter
    • Add a holder for polarized light analyzer
    • Bring new test slides


    UI Expand
    expandedtrue
    title2018-04-042022-09-26
    • Added Lumencor SOLA v-nIR
    • Added Fluorescent light collimator


    UI Expand
    expandedtrue
    title2022-09-26
    • Added 4x/0.13 Air
    • Added 20x/0.5 Air
    • Added 60x/0.85 Air
    • Updated E600 settings file
    • Nikon Preventive Maintenance
    • Adjustment and cleaning Stage Nosepiece, condenser, filter turret, focus drive, shutters
    • 100x slightly damaged but not the lens
    • Fluorescence cubes damaged: FITC TexasRed
    • Focus drive has a jitter


    UI Expand
    title2021-07-20
    • Replacement mercury lamp bulb HBO 1003W/2
    • 256 GB SSD added in workstation for OS
    • Windows 10 Installation
    • BIOS updated to v2.47
    • Camera Firmware updated to v2.11
    • NIS-Elements Basic Research v4.6 64-bits installed
    • Creation NIS Elements Settings
    • FITC and TxRed cubes excitation and emission filters slightly damaged. Dichroic OK


    UI Expand
    title2018-04-04
    • Nikon Preventive Maintenance
    • Adjustment and cleaning Stage Nosepiece, condenser, filter turret, focus drive, shutters
    • 100x slightly damaged but not the lens
    • Fluorescence cubes damaged: FITC, TexasRed
    • Focus drive has a jitter




    Tabs Page
    idTechnical Datasheet
    titleTechnical Datasheet

    Stand

    • Nikon Eclipse E600W upright Serial 725540

    Light sources

    • Transmitted light Halogen 12V 100W Model C-LPSerial 01875599
    • Reflected light

    Condenser

    • Condenser Universal C-CU Serial 081901
    • Lens Dry NA 0.9
    • Filters: Empty, Ph1, Ph2, Ph3, DICM, DICH, Empty

    Objectives

    • 4x/0.13
    • 10x/0.3 Plan Fluor Ph1 DLL WD 16 Air
    • 20x/
    • 40x/0.75 Plan Fluor Ph2 DLL WD 0.72 Air with PF40 Wollaston prism
    • 60x/0.85
    • 100x/1.3 Plan Fluor Ph3 DLL WD 0.2 Oil with PF/PA 100 Oil Wollaston prism
    • Empty
    • Empty
    • Empty

    Stage

    • Manual Stage

    Filters

    • DAPI/Hoechst/AMCA

    • FITC/EGFP

    • TRITC/Rhodamine

    • TexasRed

    Detector

    • Color Camera Nikon DS-Ri2 Serial 701234

    Workstation

    • HP Z440 Workstation
    • Intel Xeon E5-1630 v3 @ 3.7GHz
    • RAM 32 GB DDR4 2133 MHz  (4 x 8 GB)
    • OS 256 GB SSD 550 MBs
    • 2 TB HD Data Storage 150 MBs
    • Video Card NVIDIA Quadro K620 2 GB DDR3 dedicated memory
    • Monitor HP Z24i display 24' 1920 x 1200

    Consumables


    Tabs Page
    idFAQ
    titleTroubleshooting & FAQ

    Troubleshooting

    UI Expand
    titleNIS-Elements shows an error: Camera Driver...

    This happens when NIS-Elements does not find the camera. It is usually because the camera is not powered on.

    • Turn off NIS-Elements
    • Turn on the camera (both the microscope power bar and the camera need to be ON
      (the camera is usually ON, the switch on the top of the camera)
    • Check the camera is correctly connected to the computer via a USB cable
    • Turn on NIS-Elements

    FAQ

    UI Expand
    titleCan I use this microscope to look at cells in a dish?

    No. This is an upright microscope designed to observe specimen mounted between a slide and a 0.17 mm thick coverslip.





    Tabs Container
    idDemo Image
    titleNikon Eclipse E-600
    directionhorizontal


    Tabs Page
    idDemo Image
    titleDemo Image


    Image RemovedImage Added

    Cells courtesy of Benoit Bessette and Monique Vasseur (Biochemistry)
    Images courtesy of Dr Shirley Campbell and Emilie Fiola-Masson (Pharmacology)




    Inclure page
    Plateformes:_foot-en
    Plateformes:_foot-en