idDescription
titleDescription

GE

InCell

INCell Analyzer 6000

hight-content screening fully motorized inverted microscope

High content microscope

JA Bombardier Building, Room 3129
Advanced Microscope Tier 2 usage price

Instrument awarded to Dr. Steve Michnick by the Canadian Foundation for Innovation (CFI)

  • Applications
    • Transmitted light, Bright-field
    • Pseudo phase contrast and DIC
    • Fluorescence
    • High-throughput imaging
  • Light sources

    • LED for transmitted light

    • Toptica iChrome MLE for fluorescence

      Développer
      titleToptica iChrome complete specifications
      SourcePolychroicExcitation wavelength (nm)Compatible fluorophoresNominal Power
      (mW)
      405nm390/40[370-410]DAPI, Hoechst50
      488nm

      482/18

      [473-491]FITC, GFP, YFP

      25

      561nm

      564/9

      [559-569]

      Cy3, DsRed, TxRed

      30
      642nm640/14[632-647]

      Cy5, Cy5.5

      50



  • Objectives

    1. 10x/0.45 Air WD 4.0
    2. 20x/0.75 Air WD 1.0
    3. 40x/0.6 Air WD 2.7-3.7
    4. 60x/0.95 Air WD 0.15

      Développer
      titleObjective specifications


      PositionNameBrandFull nameIdentifierWorking distance (mm)Transmittance
      (% [nm])
      TechniquesCover glass thickness (mm)
      110x/0.45 AirNikon

      10x/0.45 Air
      Plan Apo
      M25x0.75

      MRD001014.0>80% [TBD-TBD]BF, p-PhC, p-DIC, Fluo0.17
      2

      20x/0.75 Air

      Nikon20x/0.75 Air
      Plan Apo
      DIC N2
      M25x0.75
      MRD002011.0>80% [TBD-TBD]BF, p-PhC, p-DIC, Fluo0.17
      3

      40x/0.6 Air

      Nikon40x/0.6 Air
      Plan Fluor ELWD
      M25x0.75
      MRH084302.7-3.7>80% [TBD-TBD]BF, p-PhC, p-DIC, Fluo0.17
      460x/0.95 Air Nikon60x/0.95 Air 
      Plan Apo
      M25x0.75
      MRD006000.15>80% [TBD-TBD]BF, p-PhC, p-DIC, Fluo0.17

      BF: Bright-field
      p-PhC: Pseudo-phase contrast
      p-
      DIC: Pseudo-Differetial interference contrast

  • Filters
    1. DAPI

    2. GFP

    3. Cy3

    4. Cy5

    5. Cy5.5

      Développer
      titleFilter specifications


      PositionNameBrandIdentifier

      Polychroic (Transmission)

      Emission filterEfective bandwith (nm)
      1DAPITBDTBDBP 420/20
      [430-462]
      455/50
      [430-480]
      [430-480]
      2GFPTBDTBD

      BP 446/32; 524/42; 600/36; 732/137

      525/20
      [515-535]
      [515-535]
      3Cy3TBD

      TBD

      BP 446/32; 524/42; 600/36; 732/137605/52
      [579-631]
      [582-618]
      4

      Cy5

      TBD

      TBD

      BP 446/32; 524/42; 600/36; 732/137707/72
      [671-743]
      [671-743]
      5

      Cy5.5

      TBDTBDBP 446/32; 524/42; 600/36; 732/137720/60
      [690-750]
      [690-750]



  • Detector
    • sCMOS Monochrome 2560 x 2160 pixels, 16-bit, pixel 6.5um x 6.5um, sensor 16.6mm x 14.0mm


Tabs Page
idUser Guide
titleUser Guide
UI Expand
expandedtrue
titleStartup
  1. Turn on the microscope (#1) – (if needed) 

    Info
    titleNote

    The switch is not easily accessible and is located on the back of the right side of the device.


  2. Turn on the computer (#2)
  3. Use your UdeM credentials to log in to Windows
  4. Start the software IN Cell Analyzer 6000

UI Expand
titleSet up

Within the software IN Cell Analyzer:

  1. Click Eject to open the access door
  2. Insert your sample in the space provided for this purpose, respecting the orientation indicated
  3. Click Load to close the access doors

    Info
    titleFirst use

    During the training you will follow the First use protocol


    Remarque

    For use with the 60x objective it is absolutely necessary to use a multi-well plate with a glass bottom. The glass thickness should be 0.17mm. We recommend the following plates:

    • Link in New Window
      linkTextGreiner SensoPlate plus 96-well plate #655891
      hrefhttps://shop.gbo.com/en/row/products/bioscience/microplates/sensoplate-glass-bottom-plates/bs-sensoplate-plus/655891.html
    • Link in New Window
      linkTextNunc Optical CVG 96-well 164588
      hrefhttps://www.thermofisher.com/order/catalog/product/164588


UI Expand
titleFirst use

When using the microscope for the first time, it is necessary to define the type of plate used. You will usually do this during the training session .This procedure can also be performed if you are using a different multi-well plate.



UI Expand
titleCreate a new plate

In the INCell Analyzer software:

  1. Select from the menu Application>Plate\Slide Manager...
  2. Click on the New Plate\Slide icon
  3. Select the number of wells in your plate (6, 24, 96 etc...)
  4. Adjust the following settings:
    1. Name: Your-Name_Your-Plate
    2. Plate height, bottom thickness and well volume usually provided in the product technical sheet by the manufacturer
    3. The material used plastic or glass
    4. If necessary adjust the number of rows and columns as well as the shape of the well (round or rectangular)
  5. Click OK
  6. Close the Plate/Slide Manager window


UI Expand
titleMeasure A1 well offset

In the INCell Analyzer software:

  1. if necessary, click on the Dashboard > Objective Lens select the 10x objective
  2. In the channel list, click the + button and add a transmitted light channel (brightfield)
  3. To the right of the plate layout, click the Setup Preview Imaging button
  4. Draw a rectangle around well A1
  5. Click on preview (on the acquisition panel at the top right of the screen)
  6. Wait for the preview to be generated
  7. On the plate map, click on the real center of well A1 to center this well on the objective
  8. In the Dashboard menu select Plate/Slide

  9. Click Edit Plate then Define Upper Left Well

  10. Click Yes

  11. Check that the yellow circle matches the edges of the well
  12. Click Yes

UI Expand
titleCompute the inter-well distance

In the INCell Analyzer software:

  1. if necessary, click on the Dashboard > Objective Lens select the 10x objective
  2. In the channel list, click the + button and add a transmitted light channel (brightfield)
  3. To the right of the plate layout, click the Setup Preview Imaging button
  4. Draw a rectangle around the bottom right well
  5. Click on preview (on the acquisition panel at the top right of the screen)
  6. Wait for the preview to be generated
  7. On the plate map, click on the real center of well A1 to center this well on the objective
  8. In the Dashboard menu select Plate/Slide

  9. Click Edit Plate then Define bottom right Well

  10. Click Yes

  11. The inter-well distance will then be automatically calculated
  12. Check that the yellow circles matches the edges of the wells
  13. Click Yes

UI Expand
titleMeasure the bottom height and thickness of the plate bottom

In the INCell Analyzer software:

  1. Click on the Dashboard
  2. In Objective Lens select the 10x objective
  3. On the plate plan, click on the center of a well containing a sample to center this well on the objective
  4. In the Dashboard menu select Plate/Slide
  5. Click Verify LAF
  6. If the 2 detected peaks (blue vertical lines) correspond to the measured peaks (black curve)
    1. Click Apply Measured Parameters
    2. Click OK
    3. Click Yes
  7. If the 2 detected peaks (blue vertical lines) do not correspond to the measured peaks (black curve)
    1. Change the bottom thickness value and repeat the operation
  8. Close the Laser Autofocus Plate/Slide Verification window


UI Expand
titleStorage management
  • Files can be saved temporarily (during acquisition) to local C: drive (desktop)
  • At the end of each session, copy your data to your external drive and delete it from local C: drive
  • You can store your files on the D: drive (Data Storage). If you do, please create one folder per laboratory using the principal investigator's last name. Inside, create a folder per user using the following nomenclature (First Name_Last Name).
Remarque
titleImportant

In any case, do not store your files on the C: drive.

UI Expand
titleShutdown
  1. Collect your samples
  2. Click Load to close the access door
  3. Close the INCell Analyzer software
  4. Transfer your data to disk D: (Data Storage) or to your external hard drive and delete it from local disk C:
  5. Turn off the computer
Remarque
titleImportant Reminders
  • Collect your samples, especially those in the microscope
  • Leave the microscope and workspace clean
Tabs Page
idLight path
titleLight path

The following diagrams allow you to follow the light path in transmitted light (bright field) and in reflected light (fluorescence). These diagrams will be available soon.

View file
namePlate_Diagram.pdf
height250


_blank
Tabs Page
idManuals
titleManuals

Available manuals

Roger Gaudry D-505

  • See

    Link in New Window
    iconfalse
    linkTextdetails
    hrefhttps://biochimie.umontreal.ca/plateformes-scientifiques-bmm/microscopie/microscope-confocal-a-haut-debit-in-cell-analyzer-6000-de-ge/
    target_blank
    on the
    Link in New Window
    iconfalse
    linkTextBiochemistry Department Microscopy Platform
    hrefhttps://biochimie.umontreal.ca/plateformes-scientifiques-bmm/microscopie/
    target_blank

  • Training and booking available upon request to the corresponding the platform manager

  • Consult the
    Link in New Window
    iconfalse
    linkTextfees structure
    hrefhttps://biochimie.umontreal.ca/wp-content/uploads/sites/37/2021/02/Tarifs_2021-22_pour_utilisation_des_plateformes_BMM.pdf
    target_blank
    and
    Link in New Window
    iconfalse
    linkTextaccess policies
    hrefhttps://biochimie.umontreal.ca/wp-content/uploads/sites/37/2018/10/Reglements_microscopes_sept_2018.pdf
    target_blank
    of the
Link in New Window
iconfalse
linkTextBiochemistry Department Microscopy Platform
hrefhttps://biochimie.umontreal.ca/plateformes-scientifiques-bmm/microscopie/
target


Tabs Page
idLog
titleLog
UI Expand
titleTo do

UI Expand
title2022-12

Because of an issue with the liquid handling motor, the liquid handling arm as been replaced by a standard fixed arm for the transmitted light LED. Therefore liquid handling is no longer available.

UI Expand
title2021-10-26

Y motor not resting when reaching the transmitted light position: This time the issue is slightly different: Y motor is homing properly during the initialization procedure but instead of resting it keeps pushing toward the home position and after a brief instant the instrument stops and the red light shows up. If the liquid handler Y motor is disabled the initialization proceed properly and the system is usable at the exception of the brightfield of course


UI Expand
title2021-09-27
  • Added to wiki


Tabs Page
idTechnical Datasheet
titleTechnical Datasheet

Instrument Serial W24318-1493815 BK02029

Service Password apiAWL


Tabs Page
idFAQ
titleTroubleshooting & FAQ

Troubleshooting

UI Expand
Troubleshooting
titleThe microscope does not show the usual green light, what should I do?

Usually, this microscope displays a green light when operational and an orange light when in use. A red light is on means the microscope is not functional. In this case please contact the platform manager.

FAQ

look at cell in a dish
UI Expand
titleFAQCan I use this microscope to
observe cells in culture?
NoYes.This a microscope designed to acquire images of specimen is designed for high-throughput acquisition of specimens in multi-well platesYou may also image . The microscope can control temperature but does not have an environmental chamber.You can also acquire images of specimen mounted between a slide and a 0.17mm thick coverslipcoverslip (thickness 0.17mm).
UI Expand
titleHow to turn off the microscope?

Usually the microscope remains on. However, it is best to turn it off if not in use for a long time (weeks). To do this:

  1. Navigate to the menu Application > Hardware >
  2. Click Shutdown Instrument
  3. Wait for the microscope to turn off
  4. The software closes automatically