Comparaison des versions

Légende

  • Ces lignes ont été ajoutées. Ce mot a été ajouté.
  • Ces lignes ont été supprimées. Ce mot a été supprimé.
  • La mise en forme a été modifiée.
Commentaire: Added objectives 4x 20x and 60x

Inclure page
_head-en
_head-en

Button Hyperlink
iconbuild
titleFrançais
typeprimary
urlhttps://wiki.umontreal.ca/display/Microscopie/Nikon+Eclipse+E600


Tabs Container
idInstrument Tabs
titleNikon Eclipse E600
directionhorizontal


Tabs Page
idDescription
titleDescription


Nikon Eclipse E-600 upright microscope

Roger Gaudry Building, Room N-620

  • Applications

    • Bright-field

    • Phase Contrast

    • Polarized light

    • Fluorescence

    • Color camera
  • Light sources

    • Halogen lamp for 30 W transmitted light

    • Mercury lamp 100 W (350~600nm) for fluorescence 

Développer
titleMercury lamp complete specifications


Peak emission (nm)Power (mW)
3347
36545
40534
43643
54637
57926

HBO Mercury lamp emission spectra (Source)
Comparison HBO Mercury vs XCite Metal Halide Lamps (Source)
OSRAM HBO 100W/2 Mercury lamp manual (pdf)

  • Objectives

    • 10x/0.3 Air Ph1 WD 16

    • 40x/0.75 Air Ph2 WD 0.75

    • 100x/1.3 Oil Ph3 WD 0.2

Développer
titleObjectives complete specifications


PositionNameBrandFull nameIDMagnificationNumerical ApertureImmersionTypeWorking distance (mm)Transmittance
(% [nm])
TechniqueCover glass thickness (mm)
110x/0.3 AirNikon10x/0.3 Air Plan Fluor Ph1 DLL
10x0.3AirPlan Fluor16>75% [400-800]
Max % @500nm
BF, Pol, PhC, Fluo0.17
2

40x/0.75 Air

Nikon40x/0.75 Air Plan Fluor Ph2 DLL
40x

0.75

Air

Plan Fluor0.72>80% [400-750]
Max % @500nm
BF, Pol, PhC, Fluo0.17
3

100x/1.3 Oil

Nikon100x/1.3 Oil Plan Fluor Ph3 DLL

100x

1.3


Remarque
Oil


Plan Fluor0.2>75% [400-800]
Max % @500nm
BF, Pol, PhC, Fluo0.17
4Empty










5Empty










6Empty











  • Filter cubes

    • DAPI/Hoechst/AMCA

    • FITC/EGFP

    • TRITC/Rhodamine

    • TexasRed

Développer
titleFilters complete specifications


PositionNameBrandIDExcitation FilterDichroic mirrorEmission FilterComments
1DAPI/Hoechst/AMCAChromaCube set 31000v2350/50x
[325-375]
400LP460/50m
[435-485]

2FITC/EGFPChromaCube set 41001

480/40x
[420-500]

505LP

535/50m
[510-560]

3TRITC/RhodamineChroma

Cube set 41002c

545/30x
[530-560]
570LP620/60m
[590-650]

4Texas RedChroma

Cube set 41004

560/55x
[532-588]

595LP645/75m
[607-682]

5Empty





6Empty






  • Detector

    • Color CMOS Nikon DS-Ri2 4908 x 3264 pixels,14-bit, 6 images/s at full frame


Tabs Page
idUser Guide
titleUser Guide


UI Expand
expandedtrue
titleStart-up
  1. Turn on the computer (#1)
  2. Turn on the microscope power bar (#2)
  3. If fluorescence is required, turn on the mercury lamp power supply unit (#3A) and press ignition (#3B)

    Avertissement

    Mercury lamp must be on for at least 30 min before being turned off and vice-versa


  4. If transmitted light is required, turn on the halogen lamp on the right side of the microscope (#4)

  5. Log in Windows using your UdM credentials

  6. Start-up NIS-Elements
Info

The first time you use the instrument, you need to import the microscope settings into the software. To do this follow the instructions Software setup.



UI Expand
titleShutdown
  1. Save your data
  2. Close NIS-Elements
  3. Transfer your data to the D: drive (Data Storage) or to your external drive and delete it from the local C: drive
  4. Turn off the computer
  5. If oil objective was used, clean it with lens cleaner and lens paper (not Kimwipes)
  6. If fluorescence was used, turn off the mercury lamp power supply unit (#3A)

    Avertissement

    Mercury lamp must be on for at least 30 min before being turned off and vice-versa


  7. If transmitted light was used, turn off the halogen lamp on the right side of the microscope (#4)
  8. Turn off the microscope power bar (#2)
  9. Wait until the lamps have cooled and cover the microscope
Remarque
titleImportant Reminders
  • Take back your samples including ones in the microscope
  • Leave the microscope and the working area clean



UI Expand
titleStorage management
  • Files can be saved temporarily (during acquisition) on the local C: drive (desktop)
  • At the end of each session, copy your data to your external drive and delete it from the local C: drive
  • You can store your files on the D: drive (Data Storage). If you do, please create a folder per laboratory using the principal investigator last name. Within, create one folder per user (Firstname_Lastname).
Remarque

In any case, your files should be removed from the C: drive.



UI Expand
titleSoftware setup

The first time you use the instrument, you need to import the microscope settings into the software. You will usually do this during the training session.
This procedure can also be performed if something is not working properly and if you want to reset the software to its original settings.

Remarque

This process will delete all experiment protocols and reset the software to the original settings for this specific microscope.

  1. If open, close NIS-Elements and wait until it is completely closed (up to 30 seconds)
  2. On your Desktop open the Softwares folder
  3. Open NIS Settings Utility
  4. Click on the Import tab
  5. Click on Browse
  6. Navigate to your Desktop
  7. Select the file Nikon-E600 Settings.bin
  8. Click Select
  9. Select all items
  10. Click Import
  11. Click OK
  12. Close the NIS Settings
  13. You can now reopen NIS-Elements



Tabs Page
idLightpath
titleLightpath


The following schematics depict the light path for transmitted (bright-field, Phase Contrast, Polarized light) and reflected (fluorescence) lights
.

PDF
nameLightPath_Nikon E600.pdf


Tabs Page
idManuals
titleManuals


Available manuals


Tabs Page
idLog
titleLog


UI Expand
expandedtrue
titleTo do
  • Adjust focus drive jitter
  • Add a holder for polarized light analyzer
  • Bring new test slides


UI Expand
expandedtrue
title2022-09-26
  • Added 4x/0.13 Air
  • Added 20x/0.5 Air
  • Added 60x/0.85 Air
  • Updated E600 settings file


UI Expand
title2021-07-20
  • Replacement mercury lamp bulb HBO 1003W/2
  • 256 GB SSD added in workstation for OS
  • Windows 10 Installation
  • BIOS updated to v2.47
  • Camera Firmware updated to v2.11
  • NIS-Elements Basic Research v4.6 64-bits installed
  • Creation NIS Elements Settings
  • FITC and TxRed cubes excitation and emission filters slightly damaged. Dichroic OK


UI Expand
title2018-04-04
  • Nikon Preventive Maintenance
  • Adjustment and cleaning Stage Nosepiece, condenser, filter turret, focus drive, shutters
  • 100x slightly damaged but not the lens
  • Fluorescence cubes damaged: FITC, TexasRed
  • Focus drive has a jitter




Tabs Page
idTechnical Datasheet
titleTechnical Datasheet

Stand

  • Nikon Eclipse E600W upright Serial 725540

Light sources

  • Transmitted light Halogen 12V 100W Model C-LPSerial 01875599
  • Reflected light
    • Power supply Nikon Mercury lamp Type C SHG1 Serial D12139
    • Bulb housing model LH M100C-1 Serial 039326

Condenser

  • Condenser Universal C-CU Serial 081901
  • Lens Dry NA 0.9
  • Filters: Empty, Ph1, Ph2, Ph3, DICM, DICH, Empty

Objectives

  • 10x/0.3 Plan Fluor Ph1 DLL WD 16 Air
  • 40x/0.75 Plan Fluor Ph2 DLL WD 0.72 Air with PF40 Wollaston prism
  • 100x/1.3 Plan Fluor Ph3 DLL WD 0.2 Oil with PF/PA 100 Oil Wollaston prism
  • Empty
  • Empty
  • Empty

Stage

  • Manual Stage

Filters

  • DAPI/Hoechst/AMCA

  • FITC/EGFP

  • TRITC/Rhodamine

  • TexasRed

Detector

  • Color Camera Nikon DS-Ri2 Serial 701234

Workstation

  • HP Z440 Workstation
  • Intel Xeon E5-1630 v3 @ 3.7GHz
  • RAM 32 GB DDR4 2133 MHz  (4 x 8 GB)
  • OS 256 GB SSD 550 MBs
  • 2 TB HD Data Storage 150 MBs
  • Video Card NVIDIA Quadro K620 2 GB DDR3 dedicated memory
  • Monitor HP Z24i display 24' 1920 x 1200

Consumables


Tabs Page
idFAQ
titleTroubleshooting & FAQ


Troubleshooting

UI Expand
titleNIS-Elements shows an error: Camera Driver...

This happens when NIS-Elements does not find the camera. It is usually because the camera is not powered on.

  • Turn off NIS-Elements
  • Turn on the camera (both the microscope power bar and the camera need to be ON
    (the camera is usually ON, the switch on the top of the camera)
  • Check the camera is correctly connected to the computer via a USB cable
  • Turn on NIS-Elements


FAQ

UI Expand
titleCan I use this microscope to look at cells in a dish?

No. This is an upright microscope designed to observe specimen mounted between a slide and a 0.17 mm thick coverslip.





Tabs Container
idDemo Image
titleNikon Eclipse E-600
directionhorizontal


Tabs Page
idDemo Image
titleDemo Image


Cells courtesy of Benoit Bessette and Monique Vasseur (Biochemistry)
Images courtesy of Dr Shirley Campbell and Emilie Fiola-Masson (Pharmacology)




Inclure page
Plateformes:_foot-en
Plateformes:_foot-en