Comparaison des versions

Légende

  • Ces lignes ont été ajoutées. Ce mot a été ajouté.
  • Ces lignes ont été supprimées. Ce mot a été supprimé.
  • La mise en forme a été modifiée.

Inclure page
_head-en
_head-en

Button Hyperlink
iconbuild
titleFrançais
typeprimary
urlhttps://wiki.umontreal.ca/display/Microscopie/Nikon+Eclipse+E600




Tabs Container
idInstrument Tabs
titleDeconvolution and Image Analysis Workstation
directionhorizontal

Tabs Page
idDescription
titleDescription

HP Z840 Deconvolution and Image Analysis Workstation

Roger Gaudry R-619

  • Applications

    • 2D and 3D Deconvolution

    • Image processing

    • Image Analysis

    • Quantification

  • Processor

    • 2 x Intel Xeon E5-2623 v4 2.60 GHz 

  • RAM

    • 128 GB (8x16GB) ECC DDR4 2400 MHz

  • Drives

    • 512 GB NVMe at 2500 MB/s OS
    • 4 TB (2x2TB) at 170 MB/s spanned volume for storage
  • Video Card

    • NVidia GTX 1080 with 8GB DDR5 dedicated memory

  • Monitor

    • HP Z24n 24" IPS display native resolution 1920 x 1200

  • OS
    • Windows 10 21H1
  • Softwares
    • ImageJ

    • FIJI

    • Microsoft Office 2016

    • NIS-Elements AR v5.02 64-bits

    • NIS Elements Offline Deconvolution v4.11 64-bits


Tabs Page
idUser Guide
titleUser Guide


UI Expand
expandedtrue
titleStartup
  1. Turn on the computer
  2. Log in Windows using your UdM credentials
  3. Start-up NIS-Elements Offline Deconvolution


UI Expand
titleShutdown
  1. Save your data
  2. Close NIS-Elements
  3. Transfer your data to the D: drive (Data Storage) or to your external drive and delete it from the local C: drive


UI Expand
titleStorage Management
  • Files can be saved temporarily (during acquisition) on the local C: drive (desktop)
  • At the end of each session, copy your data to your external drive and delete it from the local C: drive
  • You can store your files on the D: drive (Data Storage). If you do, please create a folder per laboratory using the principal investigator last name. Within, create one folder per user (Firstname_Lastname).
Remarque

In any case, your files should be removed from the C: drive.




Tabs Page
idManuals
titleManuals

Manuals


Tabs Page
idLog
titleLog


UI Expand
expandedtrue
titleTo do
  • Bring a network cable to connect directly to Ti2 microscope workstation



UI Expand
title2021-07-20
  • Windows 10 UdeM Image installation
  • Change the 2nd network adapter configuration with fix IP to enable for computer to computer communication
  • Change sharing policies of public document folder



Tabs Page
idTechnical Datasheet
titleTechnical Datasheet

Stand

  • Nikon Eclipse E600 upright Serial 725540

Light sources

  • Transmitted light Halogen 12V 100W Serial 01875599
  • Reflected light
    • Power supply Nikon Mercury lamp Type C SHG1 Serial D12139
    • Bulb housing model LH M100C-1 Serial 039326

Condenser

  • Condenser Universal C-CU Serial 081901
  • Lens Dry NA 0.9
  • Filters: Empty, Ph1, Ph2, Ph3, DICM, DICH, Empty

Objectives

  • 10x/0.3 Plan Fluor Ph1 DLL WD 16 Air
  • 40x/0.75 Plan Fluor Ph2 DLL WD 0.72 Air with PF40 Wollaston prism
  • 100x/1.3 Plan Fluor Ph3 DLL WD 0.2 Oil with PF/PA 100 Oil Wollaston prism
  • Empty
  • Empty
  • Empty
Développer
titleObjectives complete specifications


PositionNameBrandIDMagnificationNumerical ApertureImmersionTypeWorking distance (mm)Transmittance
(% [nm])
TechniqueCover glass thickness (mm)
110x/0.3 AirNikon10x/0.3 Air Plan Fluor Ph1 DLL10x0.3AirPlan Fluor16>75% [400-800]BF, Pol, PhC, Fluo0.17
2

40x/0.75 Air

Nikon40x/0.75 Air Plan Fluor Ph2 DLL40x

0.75

Air

Plan Fluor0.72>80% [400-750]BF, Pol, PhC, Fluo0.17
3

100x/1.3 Oil

Nikon100x/1.3 Oil Plan Fluor Ph3 DLL

100x

1.3


Remarque
Oil


Plan Fluor0.2>75% [400-800]BF, Pol, PhC, Fluo0.17
4Empty









5Empty









6Empty










Filters

  • DAPI/Hoechst/AMCA

  • FITC/EGFP

  • TRITC/Rhodamine

  • TexasRed

Développer
titleFilters complete specifications


PositionCube nameBrandIDExcitation FilterDichroic mirrorEmission Filter
1DAPI/Hoechst/AMCAChromaCube set 31000v2350/50x [325-375]400LP460/50m [435-485]
2FITC/EGFPChromaCube set 41001

480/40x [420-500]

505LP

535/50m [510-560]
3TRITC/RhodamineChroma

Cube set 41002c

545/30x [530-560]570LP620/60m [590-650]
4Texas RedChroma

Cube set 41004

560/55x [532-588]

595LP645/75m [607-682]
5Empty




6Empty





Detector

  • Color Camera Nikon DS-Ri2 Serial 701234

Workstation

  • HP Z440 Workstation
  • Intel Xeon E5-1630 v3 @ 3.7GHz
  • RAM 32 GB DDR4 2133 MHz  (4 x 8 GB)
  • OS 256GB SSD 550 MBs
  • 2TB HD Data Storage 150 MBs
  • Video Card NVIDIA Quadro K620 2 GB DDR3 dedicated memory
  • Monitor HP Z24i display 24' 1920x1200

Consumables


Tabs Page
idFAQ
titleTroubleshooting & FAQ


UI Expand
expandedtrue
titleTroubleshooting

NIS-Elements shows an error: Camera Driver...

This happens when NIS-Elements does not find the camera. It is usually because the camera is not powered on.

  • Turn off NIS-Elements
  • Turn on the camera (the microscope power bar and the camera switch on the top of the camera)
  • Check the USB conexion between the camera and the computer
  • Turn on NIS-Elements


UI Expand
titleFAQ

Can I use this microscope to look at cell in a dish?

  • No. This is an upright microscope designed to look at specimen between a glass slide and a 0.17mm thick cover slip.





Tabs Container
idDemo Image
directionhorizontal


Image Removed
Tabs Page
idDemo Images
titleDemo Images


Cells courtesy of Benoit Bessette and Monique Vasseur (Biochemistry)

Images courtesy of Dr Shirley Campbell and Emilie Fiola-Masson (Pharmacology)