  • Détector
    • Zeiss AxioCam MR R3 CCD Camera 1388 x 1040 pixels, 12-bit, 13 images/s at full resolution, detector size 8.9 mm x 6.7 mm
Tabs Container
idInstrument Tabs
titleZeiss Axio-Observer Z1
directionhorizontal


X-Cite 120Q Brochure (pdf)
X-Cite 120Q quick start guide (pdf)
X-Cite 120Q user guide (pdf)
Comparison between mercury and X-Cite emission spectra (Source)

Tabs Page
idDescription
titleDescription

Zeiss Axio-Observer Z1 inverted microscope

Roger Gaudry Building, Room R-421

Access upon request to the platform managerConsult the

Link in New Window
iconfalse
linkTextfees structure
hrefhttps://pharmacologie-physiologie.umontreal.ca/wp-content/uploads/sites/38/2017/05/tarification-2017.pdf
target_blank
and
Link in New Window
iconfalse
linkTextaccess policies
hrefhttps://pharmacologie-physiologie.umontreal.ca/ressources/plateforme-de-microscopie/politique-dacces/
target_blank
of the
Link in New Window
iconfalse
linkTextPhamarcology and Physiology Department Microscopy Platform
hrefhttps://pharmacologie-physiologie.umontreal.ca/ressources/plateforme-de-microscopie/
target_blank
.

Simple Microscope usage price

Instrument awarded to Dr. Audrey Claing and Jean-Philippe Gratton by the Canadian Foundation for Innovation (CFI)

  • Applications
    • Bright-field
    • Phase contrast
    • Fluorescence
  • Light sources

    • LED lamp for transmitted light

    • X-Cite 120Q 100 W (350~600nmColibri 7 (385/469/555/631) for fluorescence

      Développer
      title
X-Cite complete specifications
    • Full specifications of the Zeiss Colibri 7 light source

      Emission peak (nm)

      Power (mW)
334
    • 385/30
7
    • 150
365
    • 469/38
45
    • 110
405
    • 555/30
34
    • 40
436
    • 631/33
43
54637
57926
50

Information Zeiss Colibri 7


  • Objectives

    1. 2.5x/0.085 Air WD 8.8

    2. Empty
    3. 10x/0.25 Air Ph1 WD 6.5

    4. Empty
    5. 20x/0.5 Air Ph2 WD 2.0
    6. 40x/0.95 Air WD 0.25
      Développer
      titleObjectives complete specifications


      PositionNameBrandFull nameIDMagnificationNumerical ApertureImmersionTypeWorking distance (mm)Transmittance
      (% [nm])
      TechniqueCover glass thickness (mm)
      120x/0.5 AirZeiss20x/0.50 Ph2
      EC Plan-Neofluar
      M27
      420351-991020x2.0AirPlan Neofluar2.0Not AvailableBF, PhC, Fluo0.17
      2

      Empty












      3

      40x/0.95

      Zeiss40x/0.95
      Plan-Apochromat Corr
      M27
      420660-9970

      40x

      0.95

      Air

      Plan ApoChromat0.25>80% [400-840]BF, Fluo

      0.13 - 0.21
      Correction ring

      4Empty










      52.5x/0.085 AirZeiss2.5x/0.085
      EC Plan-Neofluar
      M27 
      420320-99022.5x0.085AirPlan Neofluar8.8>95% [400-750]BF, Fluo0.17
      610x/0.25 AirZeiss10x/0.25 Ph1
      N-Achroplan
      M27 
      420941-991110x0.25AirAchroPlan

      6.5

      Not availableBF, PhC, Fluo0.17

  • Filter cubes
    1. DAPI
    2. GFP
    3. Rhodamine
    4. DHE (dihydroethidium)
    5. Cy5
    6. Quadruple DAPI/GFP/Cy3/Cy5
      Développer
      titleFilters complete specifications


      PositionNameBrandIDExcitation filterDichroic mirrorEmission filterComments
      1DAPI
      Filter Set 49
      Zeiss

      488049-9901

      365/50
      [325-375]
      395LP445/50
      [420-470]

      2GFP
      Filter Set 13
      Zeiss488013-0000

      470/20
      [460-480]

      495LP

      517/25
      [505-530]

      3Rhodamine
      Filter Set 43
      Zeiss

      000000-1114-101

      545/25
      [533-567]
      570LP605/70
      [570-640]

      4DHECustom

      Custom

      500/50
      [475-525]

      540LP580/20
      [570-590]
      Undefined specifications
      Best guess values
      5Cy5
      Filter Set 50
      Zeiss488050-9901640/30
      [625-655]
      660LP

      690/50
      [665-715]


      6
    1. Quadruple
      DAPI/GFP/Cy3/Cy5
      FS90 HE LED

      489090-9110-000


      QBS 405 + 493 + 575 + 653

      QBP 425/30+514/30+592/25+709/100

      Excitation filters included in the light source FS90 HE LED
  • Detector
Empty
Tabs Page
idUser Guide
titleUser Guide
The X-Cite lamp must be on for at least 30 min before being turned off and vice-versa
  • for the incubation module (#2A) and open the CO2 cylinder (#2B)
  • Turn on the power bar at the left of the computer monitor (#3)

  • Press the microscope start button located on the rear left of the microscope (#4)

  • Log in Windows using your UdM credentials
  • UI Expand
    Start-up
    expandedtrue
    titleStartup
    1. Remove the dust cover from the microscope

    2. Turn on the computer (#1)

      Turn
    3. If necessary, turn on the microscope power bar (#2)
    4. Press the power witch on the left side of the microscope (#3)
    5. If fluorescence is required, turn on the X-Cite lamp (#4A) and open the diaphragm (#4B)

    Avertissement
    Remarque

    When using for the first time, it is necessary to import the microscope-specific parameters BEFORE starting the software. See the First Use section below.

  • Start -up Zen Bluethe Zen software
  • UI Expand
    titleFirst Use

    When using for the first time, it is necessary to import the microscope-specific parameters into the software. This procedure is usually carried out during the training session.However, it is also possible to use it to reset the software if it is not displayed correctly, for example.

    Remarque

    Please note, this procedure will delete all your experiment protocols and restore the software to its original settings.

    1. If open, close the Zen software and wait for it to close completely (up to 30 seconds)
    2. On the Desktop open the Documentation folder
    3. Double-click Settings for Axio-Osberver Z1
    4. A script will run and a black window will appear briefly
    5. You can then reopen the Zen software
    Shutdown
    UI Expand
    titleLoading samples

    This procedure puts the microscope in a safe configuration and performs a focus calibration. At the end of this procedure the microscope will be ready for acquisition.

    UI Expand
    titleFocus Calibration Z

    On the microscope touch screen:

    1. Press Home>Load Position to lower the stage to its lowest position
    2. Press Set Work Position to store this position
    3. If necessary, move the focus slightly up to remove the “Lower Z limit reached” message displayed on the touchscreen
    4. Press Home>Microscope>Turret>Objectives>10x to select the 10x objective
    5. If asked, tap Done to remove the oil lens cleaning warning
    6. Press Home>Microscope>XYZ>Position>Z-Position>Set zero>Auto to perform focus calibration
    7. Press OK to start the focus calibration procedure
    8. Wait a few seconds for the calibration to be completed
    Remarque

    Once calibrated, the focus can be found at Z = 1.5 mm). The Z value can be found on the microscope touch screen Home>Z-Position

    Info
    The first time you use the instrument, you need to import the microscope settings into the software. To do this follow the instructions Software setup.

    UI Expand
    titleFirst focus
    Avertissement
    1. Save your data
    2. Close Zen Blue
    3. Transfer your data to the D: drive (Data Storage) or to your external drive and delete it from the local C: drive
    4. Turn off the computer
    5. If fluorescence was used, turn off the X-Cite lamp (#4A)
    6. Turn off the microscope power bar (#2)
    7. Cover the microscope
    titleImportant

    Make sure to calibrate the focus before performing the first focus.

    On the microscope touch screen:

    1. Press Home>Microscope>Turret>Objectives
    2. Press 10x to select the 10x lens
      Info

      The 10x objective is the safest because it has the longest working distance (6.5 mm). The sample will appear perfectly sharp long before the lens approaches it. It is recommended to always first focus with the safest lens. The objectives are para-focal, focusing with the safest objective will then allow you to easily find your sample with another objective.

    3. Press Home>Load Position to lower the stage to its lowest position
    4. Press Set Work Position to store this position
    5. If necessary, move the focus slightly up to remove the “Lower Z limit reached” message displayed on the touchscreen
    6. Place the test slide on the microscope stage with the coverslip toward the objective
      Remarque
      titleImportant

      Always use the test slide to perform the first focus.

    7. If necessary, move the stage so that the sample is centered on the objective

    On the computer:

    1. Open the Zen software
    2. In the Locate tab, select BF or the desired fluorescence (GFP, DsRed, DAPI, etc…) to activate the configuration
    3. Adjust the focus with the main dial while looking through the eyepieces until the image is perfectly sharp

      Remarque

      Once calibrated, the focus can be found at Z =  1.5mm). The Z value can be found on the microscope touch screen Home>Z-Position


    4. In the Locate tab, select Off to turn off the illumination


    UI Expand
    titleSeconday focus
    Avertissement
    titleImportant

    First focus with the safest lens before selecting another lens and continuing with secondary focus.


    UI Expand
    titleFocusing with air objectives

    After performing the first focus, on the microscope touch screen:

    1. Press Home>Microscope>Turret>Objectives
    2. Press 20x or 40x to select the desired lens

      Info

      The 40x objective is the best Air objective because it has the greatest number of optical corrections (Plan Apochromat) and the largest numerical aperture (0.95). It offers a lateral resolution of 420nm at a wavelength of 550nm.


    3. Adjust the focus with the precision dial while looking through the eyepieces until the image is perfectly sharp
    4. Your sample is ready for acquisition!


    UI Expand
    titleFocusing with oil lenses

    This microscope does not have oil immersion objectives. However, if there were, the procedure would be as follows:

    After performing the first focus, on the microscope touch screen:

    1. Press Home>Microscope>Turret>Objectives
    2. Press 63x Oil, 100x Oil (1.3) ou 100x Oil (1.4) to select the desired lens. The microscope will automatically lower the stage so that the sample is accessible.

      Info

      The 63x objective is the best oil objective because it has the most optical corrections (Plan Apochromat) and the largest numerical aperture (1.4). It offers a lateral resolution of 240nm at a wavelength of 550nm.


    3. Place a drop of oil on your sample
    4. Press Done. The microscope will automatically return the sample to its original position

    In Zen Blue software:

    1. In the Locate tab, select BF or the desired fluorescence (GFP, DsRed, DAPI, etc…) to activate the configuration
    2. Adjust the focus with the precision dial while looking through the eyepieces until the image is perfectly sharp
    3. In the Locate tab, select Off to turn the illumination off
    4. Your sample is ready for acquisition!
    Remarque
    titleImportant Reminders
  • Take back your samples including ones in the microscope
  • Leave the microscope and the working area clean
  • The X-Cite lamp must be on for at least 30 min before being turned off and vice-versa
    UI Expand
    titleStorage management
    • Files can be saved saved temporarily (during acquisition) on the local C: drive (desktop)
    • At the end of each session, copy your data to your external drive and delete it from the local C: drive
    • You can store your files on the D: drive (Data Storage). If you do, please create a folder per laboratory using the principal investigator last name. Within, create one folder per user (Firstname_Lastname).
    Remarque

    In any case, your files should be removed from the C: drive.

    UI Expand
    titleSoftware setupShutdown
    1. Save your data
    2. Close the Zen software
    3. Transfer your data to the D: drive (Data Storage) or to your external drive and delete it from the local C: drive
    4. If used, turn off the incubation module power strip (#2A) and close the CO2 cylinder (#2B)

    5. Select the 10x objective and press load position to bring the objectives to the bottom position

    6. Turn off the microscope power bar (#3)

    7. Turn off the computer
    8. Cover the instrument with the protective dust cover
    Remarque
    titleImportant Reminders
    • Take back your samples including ones in the microscope
    • Leave the microscope and the working area clean

    The first time you use the instrument, you need to import the microscope settings into the software. You will usually do this during the training session.
    This procedure can also be performed if something is not working properly and if you want to reset the software to its original settings.

    Remarque

    This process will delete all experiment protocols and reset the software to the original settings for this specific microscope.

  • If open, close Zen Blue and wait until it is completely closed (up to 30 seconds)
  • On your Desktop open the Documentation folder
  • Double click on Zen Settings for Axio-Observer Z1
  • You can now reopen Zen Blue
    Tabs Page
    idLightpath
    titleLightpath

    The following schematics depict the light path for transmitted (bright-field and Phase Contrast) and reflected (fluorescence) lights.

    PDF
    nameLightPath_Zeiss
    Axio-Observer Z1 Lightpath (pdf)
    _Z1.pdf


    Tabs Page
    idManuals
    titleManuals

    Available manuals


    Tabs Page
    idLog
    titleLog
    UI Expand
    titleTo do
    • Check stage tilt
    • Ask for Colibri Remote
    • Ask for fluorescence backlight from LED
    UI Expand
    title2024-02-22 Microscope Firmware update
    • Microscope Firmware update to add Colibri to the touchscreen
    • Zen 3.5 HotFix 10
    UI Expand
    title2022-05-09
    • Computer replacement
    • Added Colibri
    • Added startup procedure
    • Parafocality and paracentralityNothing
    UI Expand
    title2022-03-17
    • Added complete description
    UI Expand
    title2021-09-27
    • Added to wiki
    Tabs Page
    idTechnical Datasheet
    titleTechnical Datasheet

    Stand

    • Zeiss Axio-Observer Z1 inverted  Serial: 3851001242 Part Number: 431007-9902-000
      System ID: 1024979772
    • Camera adapter Model 60N-C, 1", 1x, Model: 426114

    Light sources

    • Transmitted LED light
    • TBD Filters
    • X-Cite 120Q Serial: TBDColibri 7 R(G/Y)B-UV 423052-9730-000 Serial 5440000661

    Condenser

    • Manual condenser Product number: TBD, Serial: TBD
    • Lens NA 0.35 WD 70 mm Part Number: 424241

    • Filter turret 6 positions manual

      1. H

      2. Ph0
      3. Ph1

      4. Ph2

      5. DIC

      6. DUC

    Objectives

    • 2.5x/0.085 Air WD 8.8

    • Empty
    • 10x/0.25 Air Ph1 WD 6.5

    • Empty
    • 20x/0.5 Air Ph2 WD 2.0
    • 40x/0.95 Air WD 0.25

    Stage

    • Motorized stage Marzhauser Sensotech, Part number 432903-9011-000, #14 07 132052; 90-76-200-0820
    • Remote control joystick
    • Inserts
      • Slide combo
      • 6-well plate
      • 35 mm dish
      • Multi-well plat

    Filters

    1. DAPI Filter Set 49
    2. GFP Filter Set 13
    3. Rhodamine Filter Set 43
    4. DHE (dihydroethidium) 424931
    5. Cy5 Filter Set 50Empty
    6. Multiband FS90 HE LED

    Detector

    • Zeiss AxioCam MR R3 CCD Camera 1388 x 1040 pixels, 12-bit, 13 images/s at full resolution, detector size 8.9 mm x 6.7 mm. Model: r3.1 Part Number: 426509-9901-000. Serial: 1 22 12 5537

    Workstation

    • Fujitsu Esprimo P920 E90+
    • Intel Core i5-4670 @ 3.4 GHz
    • RAM 32 GB DDR3 1600 MHz ECC (4 x 8 GB)
    • OS 500 GB SSD 550 MB/s
    • 2 TB HD Data Storage (2 x 1 TB spanned volume) 110 MB/s
    • Video Card AMD FirePro V4900 1 GB DDR5 dedicated memory
    • Monitor TBD display TBDMonitor LG Flatron E2711 27' 1920 x 12001080
    • Software Zen Blue 2.03.5 HASP YRMDZ 1798977001 1798977001

    Incubation

    • Pecon stage top incubationTBD

    Consumables

    • CO2 Tank
    • N2 Tank
    • Liquid Light Guide
    • Oil
    • Lens Cleaner100 W Mercury lamp


    Tabs Page
    idFAQ
    titleTroubleshooting & FAQ

    Troubleshooting

    UI Expand
    titleI don't see any fluorescence!see a high background in fluorescence

    The fluoresncece light source is a Colibri while the transmitted light is a LED. What happens is that the fluorescence illumination reflects and into the LED and give a high background. To solve this:

    • Tilt the transmitted light arm backward
      or
    • Stop the light from entering the transmitted LED by using a cardboard

    This microscope is motorized for most of its components, yet the X-Cite light source is manual. Ensure the X-Cite diaphragm (#4B) fully opened (high position).

  • On the X-Cite lamp
  • Turn the intensity diaphragm (#4B) up

    FAQ

    UI Expand
    titleCan I use this microscope to look at cell in a dish?
    • Yes. This is an inverted microscope designed to look at specimen in a dish or a multi-well plate
    • The objectives are optimized to image through thin glass bottom multi-well plates
    • You may also image specimen mounted between a slide and a 0.17mm thick coverslip
    • For long timelapse, be aware of photo-toxicity
    UI Expand
    titleCan I use this microscope to perform timelapse experiments?

    Yes, but... This microscope has an incubation module to maintain temperature, humidity and gas. Yet it does not have a Definite focus which can maintain focus throughout time. Therefore, it is possible to loose the focus over long period.


    Tabs Container