Comparaison des versions

Légende

  • Ces lignes ont été ajoutées. Ce mot a été ajouté.
  • Ces lignes ont été supprimées. Ce mot a été supprimé.
  • La mise en forme a été modifiée.

Inclure page
_head-en
_head-en

Button Hyperlink
iconbuild
titleFrançais
typeprimary
urlhttps://wiki.umontreal.ca/display/Microscopie/Zeiss+Axio-Imager+Z2


Tabs Container
idInstrument Tabs
titleZeiss Axio-Imager Z2
directionhorizontal


Tabs Page
idDescription
titleDescription

Zeiss Axio-Imager Z2 fully motorized upright microscope

Roger Gaudry B-301

  • See

    Link in New Window
    iconfalse
    linkTextdetails
    hrefhttps://biochimie.umontreal.ca/plateformes-scientifiques-bmm/microscopie/microscope-zeiss-axio-imager-z2/
    target_blank
    on the
    Link in New Window
    iconfalse
    linkTextBiochemistry Department Microscopy Platform
    hrefhttps://biochimie.umontreal.ca/plateformes-scientifiques-bmm/microscopie/
    target_blank

  • Training and booking available upon request to the corresponding the platform manager

  • Consult the
    Link in New Window
    iconfalse
    linkTextfees structure
    hrefhttps://biochimie.umontreal.ca/wp-content/uploads/sites/37/2021/02/Tarifs_2021-22_pour_utilisation_des_plateformes_BMM.pdf
    target_blank
    and
    Link in New Window
    iconfalse
    linkTextaccess policies
    hrefhttps://biochimie.umontreal.ca/wp-content/uploads/sites/37/20182021/10/Reglements_microscopes_septoct_20182021.pdf
    target_blank
    of the
    Link in New Window
    iconfalse
    linkTextBiochemistry Department Microscopy Platform
    hrefhttps://biochimie.umontreal.ca/plateformes-scientifiques-bmm/microscopie/
    target_blank


Tabs Page
idLog
titleLog


UI Expand
titleTo do


UI Expand
title2021-10-18
  • X-Cite Liquid light guide replaced (Transmittance was 75% remaining)
  • Fluorescence Light bulb replaced (was 2593 hours)
  • Power output at the sample using 20x objective GFP cube 488nm is 116mW
  • Objective parafocality adjusted
  • Objective Focus speed was changed 10x: 7; 20x: 6; 4-x: 5; 63x: 3; 100x-1.3: 3; 100x-1.4: 3.
  • Condenser lens cleaned (oil)
  • Data storage OK


UI Expand
title2021-09-27
  • Added to wiki



Tabs Page
idFAQ
titleTroubleshooting & FAQ


UI Expand
titleTroubleshooting



UI Expand
titleFAQ

Can I use this microscope to look at cell in a dish?

  • Yes. This is an inverted microscope designed to look at specimen in a dish or a multi-well plate
  • The objectives are optimized to image through thin glass bottom multi-well plates
  • You may also image specimen mounted between a slide and a 0.17mm thick coverslip





Tabs Container
idDemo Image
directionhorizontal


Tabs Page
idDemo Image
titleDemo Image





Inclure page
Plateformes:_foot-en
Plateformes:_foot-en