Zeiss Axio-Observer Z1 inverted microscope

Roger Gaudry Building, Room R-421
Simple Microscope usage price

Access upon request to the platform managerConsult the

Link in New Window
iconfalse
linkTextfees structure
hrefhttps://pharmacologie-physiologie.umontreal.ca/wp-content/uploads/sites/38/2017/05/tarification-2017.pdf
target_blank
and
Link in New Window
iconfalse
linkTextaccess policies
hrefhttps://pharmacologie-physiologie.umontreal.ca/ressources/plateforme-de-microscopie/politique-dacces/
target_blank
of the
Link in New Window
iconfalse
linkTextPhamarcology and Physiology Department Microscopy Platform
hrefhttps://pharmacologie-physiologie.umontreal.ca/ressources/plateforme-de-microscopie/
target_blank
.

Instrument awarded to Dr. Audrey Claing and Jean-Philippe Gratton by the Canadian Foundation for Innovation (CFI)

  • Applications
    • Bright-field
    • Phase contrast
    • Fluorescence
  • Light sources

    • LED lamp for transmitted light

    • X-Cite 120Q 100 W (350~600nm) for fluorescenceColibri

Développer
titleX-Cite complete specifications


Emission peak (nm)Power (mW)
3347
36545
40534
43643
54637
57926

X-Cite 120Q Brochure (pdf)
X-Cite 120Q quick start guide (pdf)
X-Cite 120Q user guide (pdf)
Comparison between mercury and X-Cite emission spectra (Source)

  • Objectives

    1. 2.5x/0.085 Air WD 8.8

    2. Empty
    3. 10x/0.25 Air Ph1 WD 6.5

    4. Empty
    5. 20x/0.5 Air Ph2 WD 2.0
    6. 40x/0.95 Air WD 0.25
Développer
titleObjectives complete specifications


PositionNameBrandFull nameIDMagnificationNumerical ApertureImmersionTypeWorking distance (mm)Transmittance
(% [nm])
TechniqueCover glass thickness (mm)
120x/0.5 AirZeiss20x/0.50 Ph2
EC Plan-Neofluar
M27
420351-991020x2.0AirPlan Neofluar2.0Not AvailableBF, PhC, Fluo0.17
2

Empty












3

40x/0.95

Zeiss40x/0.95
Plan-Apochromat Corr
M27
420660-9970

40x

0.95

Air

Plan ApoChromat0.25>80% [400-840]BF, Fluo

0.13 - 0.21
Correction ring

4Empty










52.5x/0.085 AirZeiss2.5x/0.085
EC Plan-Neofluar
M27 
420320-99022.5x0.085AirPlan Neofluar8.8>95% [400-750]BF, Fluo0.17
610x/0.25 AirZeiss10x/0.25 Ph1
N-Achroplan
M27 
420941-991110x0.25AirAchroPlan

6.5

Not availableBF, PhC, Fluo0.17


  • Filter cubes
    1. DAPI
    2. GFP
    3. Rhodamine
    4. DHE (dihydroethidium)
    5. Cy5
Développer
titleFilters complete specifications


PositionNameBrandIDExcitation filterDichroic mirrorEmission filterComments
1DAPI
Filter Set 49
Zeiss

488049-9901

365/50
[325-375]
395LP445/50
[420-470]

2GFP
Filter Set 13
Zeiss488013-0000

470/20
[460-480]

495LP

517/25
[505-530]

3Rhodamine
Filter Set 43
Zeiss

000000-1114-101

545/25
[533-567]
570LP605/70
[570-640]

4DHECustom

Custom

500/50
[475-525]

540LP580/20
[570-590]
Undefined specifications
Best guess values
5Cy5
Filter Set 50
Zeiss488050-9901640/30
[625-655]
660LP

690/50
[665-715]


6Empty






  • Détector
    • Zeiss AxioCam MR R3 CCD Camera 1388 x 1040 pixels, 12-bit, 13 images/s at full resolution, detector size 8.9 mm x 6.7 mm
Tabs Page
idUser Guide
titleUser Guide
The X-Cite lamp must be on for at least 30 min before being turned off and vice-versa
  • Log in Windows using your UdM credentials
  • Start-up Zen Blue
  • The first time you use the instrument, you need
    UI Expand
    Start-up
    expandedtrue
    titleStartup
    1. Remove the dust cover from the microscope

    2. Turn on the computer (#1)

    3. Turn on the

      microscope

      power bar at the bak f the computer monitor (#2)

    4.  (#4B
    5. Press the

      power witch

      microscope start button located on the rear left

      side

      of the microscope (#3

      )
    6. If fluorescence is required, turn on the X-Cite lamp (#4A) and open the diaphragm
    7. )

    Avertissement
    Info
    Remarque

    When using for the first time, it is necessary to import the microscope

    settings into
    1. -specific parameters BEFORE starting the software.

    To do this follow the instructions Software setup.
    1. See the First Use section below.

    2. Startup Zen Blue
  • Save your data
  • Close Zen Blue
  • Transfer your data to the D: drive (Data Storage) or to your external drive and delete it from the local C: drive
  • Turn off the computer
  • If fluorescence was used, turn off the X-Cite lamp (#4A)

    UI Expand
    titleShutdown
    Avertissement

    The X-Cite lamp must be on for at least 30 min before being turned off and vice-versa

  • Turn off the microscope power bar (#2)

  • Cover the microscope
  • Remarque
    titleImportant Reminders
    • Take back your samples including ones in the microscope
    • Leave the microscope and the working area clean
    First Use

    When using for the first time, it is necessary to import the microscope-specific parameters into the software. This procedure is usually carried out during the training session.However, it is also possible to use it to reset the software if it is not displayed correctly, for example.

    Remarque

    Please note, this procedure will delete all your experiment protocols and restore the software to its original settings.

    1. If open, close the Zen Blue software and wait for it to close completely (up to 30 seconds)
    2. On the Desktop open the Documentation folder
    3. Double-click Settings for Axio-Osberver Z1
    4. A script will run and a black window will appear briefly
    5. You can then reopen the Zen Blue software

    UI Expand
    titleStorage management
    • Files can be saved temporarily (during acquisition) on the local C: drive (desktop)
    • At the end of each session, copy your data to your external drive and delete it from the local C: drive
    • You can store your files on the D: drive (Data Storage). If you do, please create a folder per laboratory using the principal investigator last name. Within, create one folder per user (Firstname_Lastname).
    Remarque

    In any case, your files should be removed from the C: drive.

    UI Expand
    titleSoftware setupShutdown
    1. Save your data
    2. Close Zen Blue
    3. Transfer your data to the D: drive (Data Storage) or to your external drive and delete it from the local C: drive
    4. Turn off the microscope power bar (#2)

    5. Turn off the computer
    6. Cover the microscope
    Remarque
    titleImportant Reminders
    • Take back your samples including ones in the microscope
    • Leave the microscope and the working area clean

    The first time you use the instrument, you need to import the microscope settings into the software. You will usually do this during the training session.
    This procedure can also be performed if something is not working properly and if you want to reset the software to its original settings.

    Remarque

    This process will delete all experiment protocols and reset the software to the original settings for this specific microscope.

  • If open, close Zen Blue and wait until it is completely closed (up to 30 seconds)
  • On your Desktop open the Documentation folder
  • Double click on Zen Settings for Axio-Observer Z1
  • You can now reopen Zen Blue
    Tabs Page
    idLightpath
    titleLightpath


    The following schematics depict the light path for transmitted (bright-field and Phase Contrast) and reflected (fluorescence) lights.

    PDF
    nameLightPath_Zeiss_Z1.pdf


    Tabs Page
    idManuals
    titleManuals


    Available manuals


    Tabs Page
    idLog
    titleLog
    UI Expand
    titleTo do
    • Check stage tilt
    • Ask for Colibri Remote
    • Ask for fluorescence backlight from LED
    UI Expand
    title2022-05-09
    • Computer replacement
    • Added Colibri
    • Added startup procedure
    • Parafocality and paracentrality
    UI Expand
    title2022-03-17
    • Added complete description
    UI Expand
    title2021-09-27
    • Added to wiki
    Tabs Page
    idTechnical Datasheet
    titleTechnical Datasheet

    Stand

    • Zeiss Axio-Observer Z1 inverted  Serial: 3851001242 Part Number: 431007-9902-000
      System ID: 1024979772
    • Camera adapter Model 60N-C, 1", 1x, Model: 426114

    Light sources

    • Transmitted LED light
      • TBD Filters
    • X-Cite 120Q Serial: TBD

    Condenser

    • Manual condenser Product number: TBD, Serial: TBD
    • Lens NA 0.35 WD 70 mm Part Number: 424241

    • Filter turret 6 positions manual

      1. H

      2. Ph0
      3. Ph1

      4. Ph2

      5. DIC

      6. DUC

    Objectives

    • 2.5x/0.085 Air WD 8.8

    • Empty
    • 10x/0.25 Air Ph1 WD 6.5

    • Empty
    • 20x/0.5 Air Ph2 WD 2.0
    • 40x/0.95 Air WD 0.25

    Stage

    • Motorized stage Marzhauser Sensotech, Part number 432903-9011-000, #14 07 132052; 90-76-200-0820
    • Remote control joystick
    • Inserts
      • Slide combo
      • 6-well plate
      • 35 mm dish
      • Multi-well plat

    Filters

    1. DAPI Filter Set 49
    2. GFP Filter Set 13
    3. Rhodamine Filter Set 43
    4. DHE (dihydroethidium)
    5. Cy5 Filter Set 50
    6. Empty

    Detector

    • Zeiss AxioCam MR R3 CCD Camera 1388 x 1040 pixels, 12-bit, 13 images/s at full resolution, detector size 8.9 mm x 6.7 mm. Model: r3.1 Part Number: 426509-9901-000. Serial: 1 22 12 5537

    Workstation

    • Fujitsu Esprimo P920 E90+
    • Intel Core i5-4670 @ 3.4 GHz
    • RAM 32 GB DDR3 1600 MHz ECC (4 x 8 GB)
    • OS 500 GB SSD 550 MB/s
    • 2 TB HD Data Storage (2 x 1 TB spanned volume) 110 MB/s
    • Video Card AMD FirePro V4900 1 GB DDR5 dedicated memory
    • Monitor TBD display TBD' 1920 x 1200
    • Software Zen Blue 2.0

    Incubation

    • Pecon incubation
    • TBD

    Consumables

    • CO2 Tank
    • N2 Tank
    • Liquid Light Guide
    • 100 W Mercury lamp


    Tabs Page
    idFAQ
    titleTroubleshooting & FAQ

    Troubleshooting

    UI Expand
    titleI don't see any fluorescence!

    This microscope is motorized for most of its components, yet the X-Cite light source is manual. Ensure the X-Cite diaphragm (#4B) fully opened (high position).

    1. On the X-Cite lamp
    2. Turn the intensity diaphragm (#4B) up

    FAQ

    UI Expand
    titleCan I use this microscope to look at cell in a dish?
    • Yes. This is an inverted microscope designed to look at specimen in a dish or a multi-well plate
    • The objectives are optimized to image through thin glass bottom multi-well plates
    • You may also image specimen mounted between a slide and a 0.17mm thick coverslip
    • For long timelapse, be aware of photo-toxicity
    UI Expand
    titleCan I use this microscope to perform timelapse experiments?

    Yes, but... This microscope has an incubation module to maintain temperature, humidity and gas. Yet it does not have a Definite focus which can maintain focus throughout time. Therefore, it is possible to loose the focus over long period.