Nikon Ti2-E inverted microscope

Roger Gaudry Building, Room R-619
Advanced Microscope Tier 1 usage price

Instrument awarded to Dr Paradis-Bleau the Canadian Foundation for Innovation (CFI)
Advanced Microscope Tier 1 usage price

Applications

in 2016

  • Transmitted light, Bright-field
  • Phase contrast
  • Polarized light, DIC

  • Fluorescence

  • Live imaging
  • Time-lapse
  • Timelapse imaging

Image Modified


Tabs Container
idInstrument Tabs
titleNikon Ti2-E
directionhorizontal
Tabs Page
idDescription
titleDescription

Light sources

  • LED lamp for transmitted light

  • Lumencor SpectraX for fluorescence

Expand
title
Lumencor SpectraX complete
Complete specifications
SourceID
Filter type
Excitation wavelengths (nm)Compatible fluorophoresNominal power (mW)
Violet

395/25

Bandpass

[382-407]

DAPI, Hoechst

295
Blue

440/20

Bandpass

[430-450]CFP256
Cyan

470/24

Bandpass

[458-482]FITC, GFP196
Teal

510/25

Bandpass

[497-522]YFP62
Green/Yellow550/15
Bandpass

[542-557]

TRITC, Cy3

260

Green/Yellow (Storage)

575/25
Bandpass
[562-587]

mCherry

310
Red

640/30

Bandpass

[625-655]

Cy5

231

Objectives

20x/0.5 Air Ph1 WD 2.1
  1. 60x/1.4 Oil DIC
WD 0.13
  1. 100x/1.45 Oil Ph3
WD 0.13
  1. 100x/1.45 Oil
DIC WD 0.13
  • 4x/0.2 Air WD 20
  • 20x/0.75 Air
    1. DIC
    WD 1.0
    Expand
    title
    Objectives complete
    Complete specifications
    PositionNameBrandFull nameIdentifierMagnificationNumerical apertureImmersionTypeWorking distance (mm)Transmittance
    (% [nm])
    Techniques
    ApplicationsCover glass thickness (mm)
    14x/0.2
    Air
    Nikon
    20x
    4x/0.
    5 Air Ph1Nikon

    20x/0.5 Air Plan Fluor Ph1

    MRH10201
    2
     Plan Apo Lambda

    MRD00045

    4x0.2AirPlan Apo Lambda20
    2.1750, Pol, PhC
    , Fluo0.17
    4
    2
    60x
    20x/
    1.4 Oil DIC
    0.5
    Air Ph1
    Nikon
    60x

    20x/

    1.4 Oil Plan Apo Lambda DIC N2MRD016050.13

    0.5 Ph1
     Plan Fluor

    MRH1020120x0.5Air Plan Fluor2.1>80% [400-750
    >80% [475-725
    ]BF, Pol,
    DIC
    PhC, Fluo0.17
    5
    3
    100x
    20x/
    1.45 Oil Ph3
    0.75
    Air DIC
    Nikon
    100x
    20x/
    1.45 Oil
    0.75 DIC
    Plan Apo Lambda
    Ph3MRD31905

    DIC N2
    MRD00205

    20x

    0.75

    Air

    Plan Apo Lambda

    1.0

    0.13475750
    950]BF, Pol,
    PhC
    DIC, Fluo0.17
    6
    4
    100x

    60x/1.

    45

    4
    Oil DIC

    Nikon
    100x
    60x/1.
    45
    4 Oil
    Plan

     Plan Apo Lambda
    DIC N2
    MRD0160560x1.4Oil
    MRD01905750
    725]BF, Pol, DIC, Fluo0.17
    5
    100x/1
    4x/0.2 AirNikon4x/0.2 Air
    .45
    Oil Ph3
    Nikon100x/1.45 Oil
    Plan Apo Lambda
    Ph3
    MRD31905100x1.45Oil
    20
    Plan Apo Lambda

    MRD00045

    0.13>80% [
    4001000
    750]BF, Pol, PhC, Fluo0.17
    3
    6
    20x
    100x/
    0.75 Air
    1.45
    Oil DIC
    Nikon
    20x
    100x/
    0.75 Air
    1.45 Oil
    Plan Apo Lambda
    DIC N2
    MRD01905
    MRD00205
    100x1.45OilPlan Apo Lambda0.13>80% [
    400950
    750]BF, Pol, DIC, Fluo0.17


    • Filters

      1. DAPI

      2. GFP/FITC (CFP)

      3. Cy5

      4. DAPI/GFP/Cy3/Cy5 (requires Cy3 filter in Lumencor SpectraX Green/Yellow position)

      5. DIC Analyzer
      6. CFP/YFP/mCherry (requires mCherry filter in Lumencor SpectraX Green/Yellow position)

    Expand
    titleFilters complete specifications


    PositionNameBrandIDExcitation FilterDichroic mirrorEmission FilterComments
    1DAPINikonDAPI-U HQ395/25x
    [383-408]
    303258
    425LP

    460/50m
    [435-485]
    307914

    C-FL-C DAPI-U HQ
    2GFPSemrockGFP-4050B-000

    466/40x
    [446-486]

    495LP

    525/50m
    [500-550]
    Nikon ID 96372
    3Cy5Semrock

    Cy5-5070A

    618/50x
    [5-6]
    652LP698/70
    [6-6]
    Nikon ID 96376 M351081
    4

    DAPI/GFP/Cy3/Cy5

    Semrock

    C182279

    None

    409/493/573/652432/515/595/68177074160 Custom Quad C182279.
    Excitation filters are in the Lumencor SpectraX light source
    5

    DIC Analyzer

    NikonTi2-C-DICACLNot ApplicableNot ApplicableNot ApplicableUsed for DIC imaging
    6

    CFP\YFP\mCherry XT

    SemrockC1997767None459/526/596

    475/543/702

    345131

    77074656 Excitation filters are in the Lumencor SpectraX light source


    • Detector

      • Hamamatsu ORCA Flash V2 C11440-22CU CMOS Monochrome Camera 2048 x 2048 pixels, 16-bit, 30 images/s at full frame


    Tabs Page
    idUser Guide
    titleUser Guide


    UI Expand
    expandedtrue
    titleStartup
    1. Turn on the computer (#1)
    2. Remove the cover from the microscope
    3. Turn on the microscope power bar (#2)
    4. If incubation is required, turn on the Okolab incubation module (#3A), the Lauda water bath (#3B) and open the CO2 (#3C) and N2 (#3D) tanks

      Warning

      Make sure the humidifier and the water bath are clean and properly filled with distilled water.


    5. Use your UdeM credentials to log in to Windows
    6. Start the software NIS-Elements
    Info

    The first time you use the instrument, you need to import the microscope settings into the software. To do this follow the instructions First use protocol.



    UI Expand
    titleShutdown
    1. Save your data
    2. Close NIS-Elements software
    3. Transfer your data to the D: drive (Data Storage) or to your external drive and delete it from the local C: drive
    4. Turn off the computer
    5. If oil objectives were used, clean it with lens cleaner and lens paper (not Kimwipes)
    6. If incubation was used, turn off the Okolab incubation module (#3A),Lauda water bath (#3B) and close the CO2 (#3C) and N2 (#3D) tanks
    7. Wait until the SpectaX has cooled down and turn off the microscope power bar (#2)
    8. Cover the microscope
    Note
    titleImportant Reminders
    • Collect your samples, especially those in the microscope
    • Leave the microscope and workspace clean



    UI Expand
    titleStorage management
    • Files can be saved temporarily (during acquisition) to local C: drive (desktop)
    • At the end of each session, copy your data to your external drive and delete it from local C: drive
    • You can store your files on the D: drive (Data Storage). If you do, please create one folder per laboratory using the principal investigator's last name. Inside, create a folder per user using the following nomenclature (First Name_Last Name).
    Note
    titleImportant

    In any case, do not store your files on the C: drive.




    UI Expand
    titleFirst use

    Anchor
    FirstUse
    FirstUse
    When using the microscope for the first time, you need to import the microscope settings into the software. You will usually do this during the training session.
    This procedure can also be performed if something is not working properly and if you want to reset the software to its original settings.

    Note

    This process will delete all experiment protocols and reset the software to the original settings for this specific microscope.

    1. If open, close NIS-Elements and wait until it is completely closed (up to 30 seconds)
    2. On your Desktop open the Softwares folder
    3. Open NIS Settings Utility
    4. Click on the Import tab
    5. Click on Browse
    6. Navigate to your Desktop
    7. Select the file Nikon-Ti2 Settings.bin
    8. Click Select
    9. Select all items
    10. Click Import
    11. Click OK
    12. Close the NIS Settings
    13. You can now reopen NIS-Elements



    Tabs Page
    idLight path
    titleLight path


    The following schematics depict the light path for transmitted (bright-field and Phase Contrast) and reflected (fluorescence) lights.


    PDF
    nameLightPath_Nikon Ti2.pdf


    Tabs Page
    idManuals
    titleManuals


    Available manuals


    Tabs Page
    idLog
    titleLog
    UI Expand
    expandedtrue
    titleTo do
    • Buy 20x/0.75 DIC objective has no DIC prism
    • Condenser DIC prism is DIC N1 while objective requires N2
    • Check why at 4x Condenser gives a black shadow on the righ side of the image
    • NIS Objective Z offset
    • Migrate to Windows 11
    • Nitrogen tank replacement Praxair 1066 106723504
    • CO2 Tank replacement Praxair 1013 106722901
    UI Expand
    title2025-09-03 Maintenance
    • Fixed Stage insert bent
    • Cleaning all objective and filter cubes
    • Added XY Offset
    UI Expand
    title2025-08-30 Incubation maintenance
    • Fixed incubator tubing misplaced
    • Added better attachment for tubing on the incubation chamber
    • Checked incubation chamber for obstruction
    UI Expand
    title2024-03-20 Incubation maintenance
    • Fixed incubator tubing misplaced
    • Added better attachment for tubing on the incubation chamber
    • Checked incubation chamber for obstruction
    UI Expand
    expandedtrue
    title2023-11-29
    • Fixed condenser holder loose
    • 20x/0.5 Ph1 dirty or damaged likely dirty with varnished


    UI Expand
    title2022-09-22
    • Added network cable between Acquisition and Processing workstations
    • Added shortcuts in Desktop\Documentation


    UI Expand
    title2022-03-16
    • Clean 60x objective (nail polish on it !)
    • Fix damaged stage insert
    • Stage insert tilt adjustment
    • Objective XY offset adjustments


    UI Expand
    title2022-03-16
    • Installed new objective 4x/0.2 Plan Apo Lambda
    • Updated configuration file


    UI Expand
    title2022-02-21
    • Computer maintenance
    • Imaging test slides
    • Re-install DCAM API drivers v22.2.6391
    • Okolab incubation complete maintenance
      • Disassemble Okolab stage insert
      • Disconnect the pipes
      • Disconnect the Lauda water-bath
      • Clean everything with 10% acetic acid (excepted the plastic pipes and washers)
      • Rinse with regular tap water 2 times
      • Rinse with distilled water once
      • Dry, reassemble and reconnect
      • Refill with the Lauda water-bath with 3L of milliQ water
      • Set the Okolab temperature Control Mode to Chamber
      • Set the Okolab temperature to 65°C
      • Let run for 1h. This will sterilize the water.
      • Check for any leakage
      • Set the Okolab temperature Control Mode back to Sample
      • Set the Okolab temperature back to 37°C


    UI Expand
    title2021-12-16
    • Cleaning objectives
    • Control for incubation chamber liquid intrusion


    UI Expand
    title2021-11-30
    • CO2 Calibration for Okolab: Offset of 1%


    UI Expand
    title2021-11-05


    UI Expand
    title2021-09-16
    • 20x/0.75 DIC objective was dirty and has been cleaned
    • Printed and displayed a Memo about available air and oil objectives
    • Adjustment of camera angle
    • Objective calibration
    • Objective XY offset
    • Updated NIS settings
    • Added light path schematics to wikipedia


    UI Expand
    title2021-09-15
    • Okolab display a NAN message for the free thermal sensor
    • The probe is made of two wires that need to be in contact to measure the temperature properly

      • Turn off the Okolab module
      • Disconnect the free sensor probe
      • Cut out the damaged part
      • Carefully expose the two wires
      • Connect the two wires together


    UI Expand
    title2021-09-13
    • Added label on gas bottles
    • Added line mark on humidifier


    UI Expand
    title2021-07-20
    • 512 GB SSD installed for OS
    • Windows 10 Installation
    • BIOS updated to v2.47



    Tabs Page
    idTechnical Datasheet
    titleTechnical Datasheet

    Stand

    • Nikon Ti2-E inverted  Serial 540156 System 170110-Sys-006287

    Light sources

    • Transmitted LED light
      • ND32 filter
      • IR filter
      • Manual Polarizer
    • Lumencor SpectraX 6-NII-SE Serial 9409

    Condenser

    • Motorized condenser Ti2-C-TC-E Serial 519097
    • Lens LWD NA 0.52

    • Filter turret 7 motorized positions

      1. Empty

      2. Empty
      3. Ph1

      4. Ph3

      5. Shutter

      6. Empty

      7. DIC N1

    Objectives

    • 20x/0.5 Air Ph1 WD 2.1
    • 60x/1.4 Oil DIC WD 0.13
    • 100x/1.45 Oil Ph3 WD 0.13
    • 100x/1.45 Oil DIC WD 0.13
    • 4x/0.2 Air WD 20
    • 20x/0.75 Air DIC WD 1.0

    Stage

    • Motorized stage Ti2 SHU compatible Serial 127808
    • Remote control joystick Ti2-S-JS Serial 127976
    • Inserts
      • Multi-well plate Ti2-S-HW with tilt adjustment (no incubation)
      • Combo slide 3cm dish  Ti2-S-HU with tilt adjustment (no incubation)
      • Okolab H101-CellASIC Frame with perfusion ports
        • 1 x 35mm petri dish 1x35-M + cover
        • 2 Chamber Slide 2xGS-M+ cover
        • 1 Multi-well for oil objectives MW-OIL
        • 6-well plate 6MW+ cover
        • 1 CellASIC + cover

    Filters

    1. DAPI Cube Ex 383-408 DAPI-U DM 425 BA 435-485
    2. GFP Cube Semrock 96372 M349727 17
    3. Cy5 Cube Semrock 96376 M351081 8
    4. 77074160 Custom Quad C182279 Polychroic and quad bandpass emitter for use with the following single bandpass filters: ET395/25x, ET470/24x, ET550/15x, ET640/30x
    5. DIC Analyser Ti2-C-DICACL
    6. C197767 7707\4656 CFP\YFP\mCherry XT

    Detector

    • Hamamatsu ORCA Flash V2 C11440-22CU CMOS Monochrome Camera 2048 x 2048 pixels, 16-bit, 30fps at full resolution Serial 101081

    Workstation

    • HP Z440 Workstation
    • Intel Xeon E5-1620 v4 @ 3.5GHz
    • RAM 32 GB DDR4 2400 MHz ECC (4 x 8 GB)
    • OS 500 GB SSD 530 MB/s
    • 4 TB HD Data Storage (2 x 2 TB spanned volume) 130 MB/s
    • Video Card nVidia GTX 1080 8GB DDR5 dedicated memory
    • Monitor HP Z24i display 24' 1920x1200
    • Software NIS-Elements AR v5.02

    Incubation

    • Okolab BoldLine Temperature unit Serial 284-1058 H101 T Unit BL
    • Okolab BoldLine CO2/O2 Unit  0-10/1-18 Serial 088-1102
    • Okolab OkoTouch Serial 118-224
    • Lauda water-bath Model Eco RE415 S LCK 4910 Serial LCK-4910-16-0006 Okolab 1322-1006 2017-05-30

    Consumables

    • CO2 Tank
    • N2 Tank
    • Liquid Light Guide


    Tabs Page
    idFAQ
    titleTroubleshooting & FAQ


    Troubleshooting

    UI Expand
    titleChamber does not reach the desired temperature (or is very slow to warm up)

    This can happen when the liquid running through the chamber is not flowing properly.

    • Pause your experiment
    • Turn off the Okolab environment controller 3A and 3B
    • Disconnect the blue end of the connection pipe (at the junction with the spring shape objective warmer)
    • Place the open end into a dish to collect liquid
    • Turn the Lauda water bath back ON (3B)
    • The liquid should flow quite rapidly, if not proceed as follow
      • Turn OFF the Lauda water bath (3B)
      • Take the 20 mL syringe located inside the drawer labelled Tubing
      • Connect it to the open end of the blue pipe
      • Use the syringe to blow pressurized air into the pipes to clear it out
      • Remove the syringe (and store it back into the Tubing drawer)
      • Test if the liquid is now flowing properly by turning the Lauda water bath back ON (3B)
        • If not repeat the procedure
        • If so, turn OFF the Lauda water-bath (3B)
        • Reconnect the blue and green pipes together
        • Turn the Okolab environment controller 3A and 3B back ON
    Warning

    Be careful not to touch your sample when performing these actions as it may displace your current acquisition


    Info
    titleHow the Okolab chamber works
    1. The liquid from the water bath enters the lead first via the red pipe

    2. Then it goes through the lead circuitry to keep it warm
    3. It exists via the unlabeled tube and enters the incubation chamber main body
    4. It goes through the circuitry of the main body and exits via the green labelled tube

    One must maintain a decent amount of liquid in the Lauda water bath to avoid bubble formation in the circuitry

    Because the circuitry inside the top lead and the incubation chamber main body is thin it can get clogged. The procedure above can solve this issue.




    UI Expand
    titleLiquid forming inside the incubation chamber

    This happens when the mixed-gas humidifier is overfilled. Bubbles created by the gas going through the humidifier can bring liquid into the gas feed line.

    • Turn off the Okolab module and the water bath
    • Remove your sample and store it properly
    • Dry the incubation chamber with a clean tissue
    • Carefully remove the cap of the humidifier glass bottle
    Warning

    Pay extra care when manipulating the humidifier bottle as it is made of glass and is very fragile

    • Remove distilled water from the humidifier
    Info

    Humidifier shouldn't be more than 2/3rd filled

    • Close the humidifier by replacing the cap
    • Turn on the Okolab module and the water bath



    UI Expand
    titleLiquid forming inside the incubation chamber but the humidifier is not overfilled

    This can happen during long experiments. The high humidity of the gas mixture condensates in the pipe between the humidifier and the incubation chamber.

    • Pause your experiment
    • Disconnect both ends of the yellow tube from the humidifier
    • Drain the tube from any liquid
    • Reconnect the yellow tube to the humidifier
    • You can use a syringe or gas pressure to blow out any liquid from the incubation chamber cover
    Warning

    Be careful not to touch your sample when performing this action

    • Wipe any liquid with a tissue
    • Reconnect the yellow tube to the incubation chamber 


    FAQ

    UI Expand
    titleCan I use this microscope to look at cell in a dish?
    • Yes. This is an inverted microscope designed to look at specimen in a dish or a multi-well plate
    • The objectives are optimized to image through thin glass bottom multi-well plates
    • You may also image specimen mounted between a slide and a 0.17mm thick coverslip
    • For long timelapse, be aware of photo-toxicity


    UI Expand
    titleHow can I warm up the incubation chamber faster?

    You may use a trick to warmup the incubation chamber faster. Using this method. you should reach a stable temperature within 30 minutes.

    Warning

    Do NOT proceed with your sample but with a blank control (dish with distilled water for example)

    • Place your blank control in the incubation chamber
    • Immerge the tip of the sample temperature probe in the blank
    • Set the Okolab temperature Control Mode to Chamber (Settings>Temperature>Control Mode>Chamber>Save)
    • Set the Okolab temperature to 50°C (Home>Temperature>50°C>Set)
    • Check the temperature of the sample (Menu Magnifier>Sample Temperature). It should take between 10 to 15 minutes for the blank to reach 30°C. 
    Info

    The Lauda water bath is faster to warm up water than to cool it down. Make sure to anticipate and not pass beyond the desired temperature.
    If your desired temperature is 37°C, you can stop the procedure when the blank has reached 30°C.

    When the blank has reach the desired temperature

    • Set the Okolab temperature back to 37°C (Home>Temperature>37°C>Set)
    • Set the Okolab temperature Control Mode to Sample (Settings>Temperature>Control Mode>Sample>Save)
    • Wait 10 to 20 minutes until the temperature stabilizes
    • Then you can safely replace the blank with your sample
    Image Removed