Zeiss LSM-900 confocal Airyscan microscope
Roger Gaudry R-421Transmitted lightInterference Contrast (DIC)FluorescenceLaser scanning confocalLight sources
12V 100W halogen lamp for transmitted light
X-Cite 120LED mini for visible fluorescence
| Expand |
|---|
| title | X-Cite 120LED mini complete specifications |
|---|
|
Emission peak (nm) | Power (mW) |
|---|
334 | 5 |
5x/0.15 Ph1 Air WD 5.30 N-AchroPlan 420931-9911
10x/0.25 Ph1 WD TBD N-AchroPlan 420941-991120x/0.80 Air WD 0.61 Plan ApoChromat 420650-9901
40x/1.4 Oil WD TBD Plan ApoChromat 420762-990063x/1.40 Oil WD 0.19 Plan ApoChromat 420782-9900
40x/0.95 Air WD TBD Plan ApoChromat 420660-9970 Variable coverslip 0.13-0.21| Expand |
|---|
| title | Complete lens specifications |
|---|
|
Position | Nom | Marque | Nom complet | Identifiant | Grossissement | Ouverture numérique | Immersion | Type | Distance de travail (mm) | Transmittance (% [nm]) | Technique | Épaisseur du couvre-objet (mm) |
|---|
2 | 20x/0.80 Air | Zeiss | 20x/0.8 DIC II Plan-Apochromat W0.8x1/36" | 440640-9903-000 | 20x | 0.8 | Air | Plan Apochromat | 0.55 | >90% [410-800] | BF, DIC, Fluo | 0.17 |
4 | 63x/1.40 Huile | Zeiss | 63x/1.4 DIC III Plan-Apochromat M27 | 420782-9900-000 | 63x | 1.4 | Plan Apochromat | 0.19 | >80% [440-710] | BF, DIC, Fluo | 0.17 | BF: Bright-field
DIC: Interference contrast
38 GFP49 DAPI64 mPlumDIC AnalyzerEmptyEmpty| Expand |
|---|
| title | Complete filter specifications |
|---|
|
Position | Nom | Marque | Identifiant | Filtre d'excitation | Miroir dichroïque | Filtre d'émission | Commentaire | 1 | DAPI Filter Set 49 | Zeiss | 488049-9901-000 | 365/50 [340-390] | 395LP | 445/50 [420-470] | 2 | GFP Filter Set 38 | Zeiss | 000000-1031-346 | 470/40 [450-490] | 495LP | 525/50 [500-550] | FT 495 | BP 525/50 | | Tabs Page |
|---|
| id | User Guide |
|---|
| title | User Guide |
|---|
|
| UI Expand |
|---|
|
Remove the dust cover from the microscopeTurn on the computer (#1)Turn on the System (#2) and Components (#3) switches in the rack on the left of the microscope
Turn on the laser key (#4) in the rack on the left of the microscope
Use your UdM credentials to log in to Windows| Note |
|---|
When using for the first time, it is necessary to import the microscope-specific parameters BEFORE starting the software. See the First Use section below.
|
Start the Zen software
| UI Expand |
|---|
|
When using for the first time, it is necessary to import the microscope-specific parameters into the software. This procedure is usually carried out during the training session.However, it is also possible to use it to reset the software if it is not displayed correctly, for example.| Note |
|---|
Please note, this procedure will delete all your experiment protocols and restore the software to its original settings. |
- If open, close the Zen software and wait for it to close completely (up to 30 seconds)
- On the Desktop open the Documentation folder
- Double-click Settings for LSM800
- A script will run and a black window will appear briefly
- You can then reopen the Zen software
| UI Expand |
|---|
|
This procedure puts the microscope in a safe configuration and performs a focus calibration. At the end of this procedure the microscope will be ready for acquisition.
| UI Expand |
|---|
|
On the microscope touch screen: - Press Home>Load Position to lower the stage to its lowest position
- Press Set Work Position to store this position
- If necessary, move the focus slightly up to remove the “Lower Z limit reached” message displayed on the touchscreen
- Press Home>Microscope>Control>Objectives>5x to select the 5x objective
- If asked, tap Done to remove the oil lens cleaning warning
- Press Home>Microscope>XYZ>Position>Z-Position>Set zero>Auto to perform focus calibration
- Press OK to start the focus calibration procedure
- Wait a few seconds for the calibration to be completed
| Note |
|---|
Once calibrated, the focus can be found at Z = 1.6 mm). The Z value can be found on the microscope touch screen Home>Z-Position |
|
| UI Expand |
|---|
|
| Warning |
|---|
|
Make sure to calibrate the focus before performing the first focus. |
On the microscope touch screen:
Press Home>Microscope>Turret>ObjectivesPress 5x to select the 5x lens
| Info |
|---|
The 5x objective is the safest because it has the longest working distance (TBD mm). The sample will appear perfectly sharp long before the lens approaches it. It is recommended to always first focus with the safest lens. The objectives are para-focal, focusing with the safest objective will then allow you to easily find your sample with another objective. |
- Inverted microscope
- Airyscan super-resolution with joint deconvolution
- Point scanning confocal
- Brightfield
- Phase contrast
- Interference Contrast (DIC)
- Fluorescence
- Targeted photo-modulation FRAP
- Incubation
- Timelapse imaging
- Deconvolution
- Extended Depth of Focus
Image AddedClick to enlarge
Image AddedClick to enlarge
| Tabs Container |
|---|
|
| Tabs Page |
|---|
| DescriptionLight sources| Expand |
|---|
| title | Complete specifications |
|---|
| Emission (nm) | Nominal Power (mW) | Measured Power (mW) |
|---|
405 | 5 |
| 488 | 10 |
| 561 | 10 |
| 640 | 5 |
|
|
- X-Cite 120LED mini for visible fluorescence
| Expand |
|---|
| title | Complete specifications |
|---|
| Filter (nm) | Emission Peak (nm) | Nominal Power (mW) | Measured Power (mW) |
|---|
DAPI | 365 |
|
| GFP | 470 |
|
| mPlum | 587 |
|
|
Image Added
| View file |
|---|
| name | XCite_120LEDmini_Datasheet.pdf |
|---|
| height | 250 |
|---|
|
|
Objectives- 5x/0.15 Ph1 Air
- 10x/0.25 Ph1 Air
- 20x/0.80 Air
- 40x/0.95 Air
- 40x/1.4 Oil
- 63x/1.4 Oil
| Expand |
|---|
| title | Complete specifications |
|---|
| Position | Name | Brand | Full name | Identifier | Magnification | Numerical aperture | Immersion | Type | Working distance (mm) | Transmittance (% [nm]) | Applications | Coverglass thickness (mm) |
|---|
1 | 5x/0.15 Air | Zeiss | 5x/0.15 Ph1 N-AchroPlan | 420931-9911-000 | 5x | 0.15 | Air | N-AchroPlan | 12.0 | Not Available | BF, PhC, Fluo | 0.17 | 2 | 10x/0.25 Air | Zeiss | 10x/0.25 Ph1 N-AchroPlan | 420941-9911-000 | 10x | 0.25 | Air
| N-AchroPlan | 6.5 | Not Available | BF, PhC, Fluo | 0.17 | 3 | 20x/0.8 Air
| Zeiss | 20x/0.8 Plan-Apochromat | 420650-9901-000 | 20x | 0.8 | Air | Plan Apochromat | 0.55
| >90% [400-800] | BF, DIC, Fluo | 0.17 | 4 | 40x/0.95 Air | Zeiss | 40x/0.95 Plan-Apochromat | 420660-9970-000
| 40x | 0.95 | Air | Plan Apochromat | 0.25 | >80% [410-820] | BF, DIC, Fluo | 0.13-0.21 | 5 | 40x/1.4 Oil | Zeiss | 40x/1.4 Plan-Apochromat | 420762-9900-000 | 40x | 1.4 | Oil | Plan Apochromat | 0.13 | >80% [500-850] | BF, DIC, Fluo | 0.17 | 6 | 63x/1.4 Oil
| Zeiss | 63x/1.4 Plan-Apochromat | 420782-9900-799 | 63x | 1.4 | Oil | Plan Apochromat | 0.19 | >80% [440-710] | BF, DIC, Fluo Super-Resolution Strehl ratio >90% | 0.17 |
|
Filters
- GFP
- DAPI
- mPlum
- DIC Analyzer
- -
- -
| Expand |
|---|
| title | Complete specifications |
|---|
| |
Detectors- 2x Gallium Arsenide Phosphid (GaAsP) Photomultiplier Tube (PMT)
- 1x transmitted light Electronically Switchable Illumination and Detection module (ESID)
- 1x Airyscan detector for 63x objective
| Expand |
|---|
| title | Complete specifications |
|---|
| Image Added
|
|
| Tabs Page |
|---|
| User Guide| UI Expand |
|---|
| - If not already done, turn on the computer (#1) and log in to Windows using your UdeM credentials
- Remove the dust cover from the microscope
- Turn on the System (#2) and Components (#3) switches in the rack below microscope
- Turn on the laser key (#4) in the rack below the microscope
- When using the instrument for the first time, it is necessary to import the microscope configuration into the software. See the First Use section below.
- Start Zen
|
| UI Expand |
|---|
| When using the instrument for the first time, it is necessary to import the microscope configuration into the software. This procedure is usually carried out during the training session. However, it can also be used to reset the software if it does not display correctly, for example. Running this procedure will erase all your experiment protocols and reset the software to its original settings (ask for support if you are not sure).
If Zen is open, close it and wait until it has completely shut down (this may take up to 30 seconds) On the Desktop, open the Softwares folder Double-click Zen Settings for LSM 900 A script will run and a black window will appear briefly When the message Settings for Zen have been imported successfully appears, click OK to close it You can now open Zen
|
| UI Expand |
|---|
| During this procedure, you will: - Set the microscope to a safe configuration
- Perform a calibration
- Load your sample
- Find and adjust the focus
Once completed, your sample will be ready for acquisition. | UI Expand |
|---|
| This step is required to calibrate the microscope in XY and Z. Performing this calibration will significantly reduce the time needed to locate and focus on your sample. - If not already done, select the lowest-magnification objective. On the microscope touch screen Home > Microscope > Control > Objectives > 5x
- In Zen, once it has started a calibration dialog should appear. Simply click Calibrate Now.
The microscope will lower the objectives, perform an XY calibration first, followed by a Z calibration, and then return to its original position.
| Tip |
|---|
Once calibrated, the focus is typically found at Z = 1.7 mm for a microscope slide. The Z position can be viewed on the microscope touch screen under Home > Z-Position, as well as in Zen within the Focus tab, located on the right side of the screen. |
| Expand |
|---|
| title | The calibration dialog did not appear... |
|---|
| The system may already be calibrated. This can occur if a previous user calibrated the system and left it on. | UI Expand |
|---|
| title | Verifying if the system is already calibrated |
|---|
| In Zen, within the Focus tab, located on the right side of the screen: Click Load to lower the objective - The Z position is indicated below Current
- If Z position value is less than 100 um the system is calibrated
On the microscope touch screen: Lower the objective by pressing Home > Load Position Press Set Work Position to store this position If necessary, slightly adjust the focus upward to clear the Lower Z Limit Reached message displayed on the touch screen - The Z position value is indicated
- If the value is less than 0.1 mm then the system is calibrated
|
| UI Expand |
|---|
| title | Manual calibration with the software |
|---|
| In Zen, within the Stage tab, located on the right side of the screen: - If not already done, check the Show All option
At the bottom of the tab click Calibrate A warning message will show up click Continue The microscope will lower the objectives and perform a XY stage calibration and then return to its original position.
In Zen, within the Focus tab, located on the right side of the screen: - If not already done, check the Show All option
At the bottom of the tab click Calibrate A warning message will show up click Continue The microscope will then perform a Z calibration and then return to its original position
XY and Z calibrations are now complete |
| UI Expand |
|---|
| title | Manual calibration with the microscope touch screen |
|---|
| On the microscope touch screen: Navigate to Home > Microscope > XYZ > Position > Z-Position > Set Zero > Auto to perform focus calibration Press OK to start the calibration procedure Wait a few seconds for the calibration to complete Lower the objective by pressing Home > Load Position Press Set Work Position to store this position If necessary, slightly adjust the focus upward to clear the Lower Z Limit Reached message displayed on the touch screen Navigate to Home > Microscope > XYZ > Position > XY-Position > Set Zero > Auto to perform a stage calibration Press OK to start the calibration procedure Wait a few seconds for the calibration to complete
XY and Z calibrations are now complete |
| UI Expand |
|---|
| title | Ensuring the calibration dialog is displayed at startup |
|---|
| In Zen: In the menu bar, navigate to Tools > Options Select Startup/Shutdown Under Stage/Focus Calibration, ensure Request Stage/Focus Calibration on Startup is checked Click OK to close the Options dialog
|
|
|
| UI Expand |
|---|
| | Warning |
|---|
| Ensure that the calibration has been completed beforehand. Calibration will significantly reduce the time required to locate and focus on your sample. |
On the microscope touch screen: - If not already done, select the lowest-magnification objective Home > Microscope > Control > Objectives > 5x
| Info |
|---|
The 5× objective is the safest to use due to its long working distance (12 mm). The sample will appear in sharp focus well before the objective approaches it. It is recommended to always focus using the safest objectives first. Since the objectives are parafocal, focusing with the safest objectives will facilitate locating the sample when switching to higher-magnification objectives.
|
Lower the objective by pressing Home > Load Position
|
|
|
|
Press Home>Load Position to lower the stage to its lowest positionPress Set Work Position to store this position If necessary,
|
|
|
|
move slightly adjust the focus
|
|
|
|
slightly up remove “Lower limit reached” Limit Reached message displayed on the
|
|
|
|
touchscreentouch screen - Place the test slide on the microscope stage with the coverslip
|
|
|
|
toward |
Always use the to perform the first focuswill significantly reduce the time required to set up the instrument. |
- If necessary,
|
|
|
|
move so - to ensure that the sample is
|
|
|
|
centered on computer- centred under the objective
|
|
|
|
On the Open the Zen
the - the Locate tab, select BF or the desired
|
|
|
|
fluorescence , - or mPlum) to activate the configuration
- Adjust the focus with the main dial while looking through the eyepieces until the image is perfectly sharp
|
|
|
|
note| Tip |
|---|
Once calibrated, the focus |
|
|
|
|
can be is typically found at Z = 1. |
|
|
|
|
6 mm). 7 mm for a microscope slide. The Z |
|
|
|
|
value found In the viewed on the microscope touch screen |
|
|
|
|
Home>Z-Positionunder Home > Z-Position, as well as in Zen within the Focus tab, located on the right side of the screen. |
- In the
|
|
|
|
- Locate tab, select Off to turn off the illumination
|
|
|
|
|
First focus with the safest lens before selecting another lens and continuing with secondary focusPerform the initial focus using the safest objective before switching to higher-magnification objectives. |
| UI Expand |
|---|
| title | Focusing with air objectives |
|---|
| After performing the |
|
|
|
|
first initial focus, on the microscope touch screen: - Press
|
|
|
|
|
Home>Microscope>Turret>- Home>Microscope>Control>Objectives
- Press 10x, 20x
|
|
|
|
|
or 40x - , 40x (0.95) to select the desired
|
|
|
|
|
lens 40x 40× (0.95) Air objective is the best |
|
|
|
|
|
Air air objective because it has the |
|
|
|
|
|
greatest highest number of optical corrections (Plan Apochromat) and the largest numerical aperture (0.95) |
|
|
|
|
|
., but it has a smaller field of view. The 20×(0.8) objective offers the best compromise between resolution and field of view. |
|
|
|
|
|
It offers a lateral resolution of 420nm at a wavelength of 550nm.note |
|---|
There are two (2) 40x objectives, make sure you select the |
|
|
|
|
|
Air oneIn Zen: - In the Locate tab, select BF or the desired fluorescence (DAPI, GFP or mPlum) to activate the configuration
- Adjust the focus with the precision dial while looking through the eyepieces until the image is perfectly sharp
- In the Locate tab,
|
|
|
|
|
select - select Off to turn off the illumination
|
|
|
|
|
off
Your sample is ready for acquisition! |
|
|
|
|
lenses first initial focus, on the microscope touch screen:
- Press Home>Microscope>Turret>Objectives
|
|
|
|
|
Press 63x Oil, 40x Oil - Press 40x Oil (1.4) or 63x Oil (1.4)to select the desired
|
|
|
|
|
lens- objective.The microscope will automatically
|
|
|
|
|
lower the stage so that the sample is accessible.| Info |
|---|
The 40x objective is the best oil objective because it has the most optical corrections (Plan Apochromat) and the largest numerical aperture (1.4). It offers a lateral resolution of 240nm at a wavelength of 550nm. |
- lower the stage so that the sample becomes accessible.
| Info |
|---|
The 40× and 63× oil objectives provide the same spatial resolution because they share the same numerical aperture (1.4). The 40× oil objective offers a larger field of view and transmits light slightly better beyond 700 nm. The 63× oil objective transmits light slightly better within the visible spectrum (440–710 nm) and has a higher Strehl ratio (90%). It is particularly well suited for super-resolution imaging, although its field of view is smaller. |
|
|
|
|
|
notePlace a |
|---|
There are two (2) 40x objectives, make sure you select the |
|
|
|
|
|
OIL one- drop of oil on the objective
- Press Done. The microscope will automatically return the
|
|
|
|
|
sample - objective to its original position
|
|
|
|
|
In software the - the Locate tab, select BF or the desired fluorescence (DAPI, GFP
|
|
|
|
|
, - or mPlum) to activate the configuration
- Adjust the focus with the precision dial while looking through the eyepieces
|
|
|
|
|
until the - until the image is perfectly sharp
- In the Locate tab,
|
|
|
|
|
select - select Off to turn off the illumination
|
|
|
|
|
off
Your sample is ready for acquisition! |
|
|
| UI Expand |
|---|
| - Files can be saved temporarily (during acquisition) on the local C: drive (desktop)
- At the end of each session, copy your data to your external drive and delete it from the local C: drive
- You can store your files on the D: drive (Data Storage). If you do, please create a folder per laboratory using the principal investigator last name. Within, create one folder per user (Firstname_Lastname).
| Note |
|---|
In any case, your files should be removed from the C: drive. |
|
|
|
Close the software - Close Zen
- Transfer your data to the D: drive (Data Storage) or to your external drive and delete it from the local C: drive
- If used, clean oil
|
|
|
lenses - objectives with lens cleaner and paper
- Select the 5x objective and press load position to place the objectives in a safe position
- Turn off
|
|
|
the - the laser key (#4) in the rack
|
|
|
on left of the System #2 and Components#3 switches on the left of - below the microscope
- Turn off
|
|
|
the the microscope- the instrument with the protective dust cover
|
|
|
title | Important Reminders |
|---|
- Take back your samples including ones in the microscope
|
|
|
|
Leave the microscope and the working area clean| Tabs Page |
|---|
| id | Light Path |
|---|
| title | Light Path |
|---|
|
The following diagrams allow you to follow the light path in transmitted light (bright field, DIC) and in reflected light (fluorescence).- Leave the microscope and the working area clean
|
|
|
|
ManualsManuals | |
| Tabs Page |
|---|
| Log| UI Expand |
|---|
| - Quality control for Illumination, Liquid light guide and filters quality.
- Revise Technical Datasheet
|
|
|
| id | Log |
|---|
title | LogTo do | - Quality control for Illumination, Liquid light guide and filters quality.
|
| 2025-10-09 20x and 40x Air objectives dirty |
| - 20x/0.8 and 40x/0.95 with immersion oil on it. Cleaning objectives.
- Added Troubleshooting
|
| UI Expand |
|---|
| title | 2024-09-19 Joystick replacement |
|---|
| |
| UI Expand |
|---|
| title | 2024-09-03 Airyscan upgrade |
|---|
| - Installation Airyscan upgrade. New workstation. New detector. New casing.
|
| UI Expand |
|---|
| title | 2024-07-08 Joystick issue |
|---|
| - Issue when the joystick is at the rest position. If it is slightly to the left, the stage will move while the image is acquired. Temporary fix: Make sure the joystick is slightly to the right when it is resting.
|
| UI Expand |
|---|
| title | 2024-02-14 Added to wiki |
|---|
| - User guide added to Wiki French and english version
|
| UI Expand |
|---|
| - Zen 2.6 updated hotfix 12
- Microsoft Windows updated
|
|
| Tabs Page |
|---|
| id | Technical Datasheet |
|---|
| title | Technical Datasheet |
|---|
| Technical datasheetStand |
|
LSM900 Serial with LSM900 confocal - Confocal
- Manual Aperture diaphragm
- Manual Fluorescence field diaphragm
- 6x motorized nosepiece
- 6x Motorized reflector changer
- 3x Motorized sideport turret (100% Visible, 100% Left (confocal), 100% right (empty)
- TL Motorized shutter
- RL Motorized shutter
Light sources- Transmitted LED light with ESID 423053-9071
- X-Cite 120LED
|
|
mini - mini Serial XT120LM Serial XT120LM-0279 RS232 COM Port 6
- Laser URGB 400102-9300-000 Serial
|
|
2631000466 - 2631000466 Toptica iChrome-ZLE-4_40407 v3.1 Aug 2016
CondenserObjectives5x/0.15 Ph1 Air WD 5.6 N-AchroPlan 420931-9911-000 - 10x/0.25 Ph1 WD 6.0 N-AchroPlan 420941-9911-000
20x/0.80 Air WD 0.61 Plan-Apochromat 420650-9901-000 - 40x/0.95 Air WD 0.25 Plan-Apochromat 420660-9970-000 Variable coverslip 0.13-0.21
- 40x/1.4 Oil WD 0.13 Plan-Apochromat 420762-9900-000
63x/1.40 Oil WD 0.19 Plan Apochromat 420782-9900-799
Stage- Motorized stage Marzhauser Scan IM 130x100 Serial 14032630
- Remote control joystick Marzhauser
|
|
2-Schsen-Joystick CZ Serial 2420142128 Article 90-76-200-0820 Zeiss Serial 432903-9011-000Inserts
- Slide combo
- 6-well plate
- 35 mm dish
- Multi-well plat
- 2-Schsen-Joystick CZ Serial 2420142128 Article 90-76-200-0820 Zeiss Serial 432903-9011-000
- Inserts
- Slide combo
- Multi-well plate
Filters6-Position motorized Filter Turret ##424947 - GFP Zeiss Filter Set 38 cube 424931 000000-1031-346
- DAPI Zeiss Filter Set 49 cube 424931 489064-0000-000 (in MTB 488049-0000-000)
- mPlum Zeiss Filter Set 64 HE cube 424931 489064-0000-000
- DIC Analyzer Zeiss 424937-9901
- Empty
- Empty
ScannerLSM 800 GaAsP-PMT 400102-9000-00000 confocal 2633000272 (in MTB LSM900 Scanhead)
- 8x Motorized Filter wheel 1
- Empty
- SP 470 000000-2090-296 410-470 93% 480-750 0.001%
- SP 545 000000-2090-297 410-546 93% 557-750 0.001%
- SP 620 000000-2090-298 410-617 93% 630-750 0.001%
- Empty
- Empty
- Empty
- Empty
- 8x Motorized Filter wheel 2
- Empty
- LBF 640 0000000-2090-299 410-617 93% 630-644 0.001%
- LP 575 0000000-2090-294 576-750 93% 410-565 0.001%
- LP 655 0000000-2090-295 656-750 93% 395-643
|
|
Filters
6-Position motorized Filter Turret 424947
- GFP Zeiss Filter Set 38 cube 424931 000000-1031-346 BP 470/40 450-490 FT 495 BP525/50 500-550
- DAPI Zeiss Filter Set 49 cube 424931 489064-0000-000 G365 FT 395 BP 445/50 335-383 395 420-470
- mPlum Zeiss Filter Set 64 HE cube 424931 489064-0000-000 BP587/25 FT605 BP647/70 574-599 605 612-682
- DIC Analyzer Zeiss 424937-9901
- Empty
- Empty
Scanner
LSM900 Scan head
- 8x Motorized Filter wheel 1
EmptySP 470 000000-2090-296 410-470 93% 480-750 0.001%SP 545 000000-2090-297 410-546 93% 557-750 - 0.001%
- SP 620 000000-2090-298 410-617 93% 630-750 0.001%
- Empty
- Empty
- Empty
|
|
Empty- 8x Motorized Filter wheel 2
- Empty
- LBF 640 0000000-2090-299 410-617 93% 630-644 0.001%
- LP 575 0000000-2090-294 576-750 93% 410-565 0.001%
- LP 655 0000000-2090-295 656-750 93% 395-643 0.001%
- SP 620 000000-2090-298 410-617 93% 630-750 0.001%
- Empty
- Empty
- Empty
Main Beam splitter
MBS 405+488+561+640 T10/R90
400102-9600-000
Transmission
- 420-471 93%
- 502-546 93%
- 575-618 93%
- 656-800 93%
- 631-642 10%
Reflection
- 399-410 99%
- 481-492 99%
- 558-564 99%
- 631-642 90%
Detector
- 2 GaASp PMT
- 1 Airyscan 2 63x Serial 2657000116
- 1 transmitted ESID Motorized Left TL ESID LED, Right ESID Detector
Workstation
- HP Z2 G9
I155440-CIB
- Motherboard HP 895C
- BIOS version 2024-09-03
- 1 x Intel Core i7-12700K 3.6 GHz
- RAM 128GB DDR5 4 x 32GB no ECC
- OS 500 GB NVMe 6 GB/s
- 7 TB HD Data Storage 213 MB/s
- VNvidia RTX A4000 16 GB GDDR6
- Monitor HP Z32x
- Software Zen 3.10
Incubation
Consumables
Main Beam splitter : MBS 405+488+561+640 T10/R90 400102-9600-000 Transmission - 420-471 93%
- 502-546 93%
- 575-618 93%
- 656-800 93%
- 631-642 10%
Reflection - 399-410 99%
- 481-492 99%
- 558-564 99%
- 631-642 90%
Detector- 2 GaASp PMT
- 1 Airyscan 2 63x Serial 2657000116 2263-682
- 1 transmitted ESID Motorized Left TL ESID LED, Right ESID Detector
Workstation- HP Z2 G9 Serial CZC4027PF7 829W6EC#ABB
I155440-CIB - Motherboard HP 895C
- BIOS version 03.04.08 Rev.A 2025-07-02
- 1 x Intel Core i7-12700K 3.6 GHz
- RAM 128GB DDR5 4 x 32GB no ECC
- OS 500 GB NVMe 6 GB/s
- Data Storage 14 TB (2 x 7 TB) HD 14 TB (2 x 7 TB) HD 213 MB/s
- Nvidia RTX A4000 16 GB GDDR6
- Monitor HP Z32x CNC61804QR 3840 x 2160 60 Hz CNC61804QR 3840 x 2160 60 Hz
- Software Zen 3.10 SN=1122911743-424077 HASP=110522193
- Zen Module Arivis Cloud Basic, Arivis Cloud Advanced, Colocalization, Data Storage, Direct Processing, Manual extended Focus, Measurements Multi-Channel. Panorama, Time Series, Advance Processing, Airyscan, Airyscan 2 Basic, Ariscan Hoiunt Deconvolution Exteneded Focus, HDR Confocal Basic, Image Analysis, Sotfware Autofocus, Tiles and Positions, Z Stack
Incubation- Pecon stage top incubation
Consumables- CO2 Tank
- N2 Tank
- Oil
- Lens Cleaner
Manuals |
| Tabs Page |
|---|
| title | Troubleshooting & FAQ |
|---|
| Troubleshooting| UI Expand |
|---|
| title | I don't see any fluorescence! |
|---|
| Technical support is free and unlimited. Contact us ! |
| UI Expand |
|---|
| This can happen when objectives are dirty. this is especially true for the 20x/0.8 and 40x/0.95 which have a short working distance. If not careful immersion oil can make them dirty. If your image seems blurry with one of these objective please contact us and we will clean it for you ! Image AddedClick to enlarge
|
FAQ |
|
| Tabs Page |
|---|
| id | FAQ |
|---|
| title | Troubleshooting & FAQ |
|---|
|
TroubleshootingFAQI don't see any fluorescence! | The best way to solve a problem in Microscopy is to follow the light path. You will find in the Light path tab of this page, the diagrams which will allow you to follow the light all along its path through the microscope.- Open the light path file
- Starting from the light source and moving towards the detector, verify that there is indeed light after each component of the microscope
| Can I use this microscope to perform timelapse experiments? |
| Yes, but... This microscope has an incubation module to maintain temperature, humidity and atmosphere. Yet, it does NOT have a module which can maintain focus throughout time. Therefore, it is possible to loose the focus over long period. You can always do a software auto-focus at different time but be aware of photo-toxicity if you are using fluorescence. |
|
|
Can I use this microscope to observe cells in culture? | | Yes it is an inverted microscope. Point scanning confocal induce damage so care should be taken while imaging live cells. a coverslip (thickness 0.17mm). |
| What is Extended Depth of Focus? |
| Extended Focus is a method to flatten a Z-stack by keeping only the relevant information. In Zen: - Select the Processing Tab
- Select the Extended Depth of Focus Method
- Click Apply
- Wait until the processing is finished
- Save the processed image
Image AddedZ Stack animation Image AddedMaximum Intensity projection of a Z-Stack Image AddedExtended Depth of Focus projection of a Z-Stack
|
| UI Expand |
|---|
| title | Can I use this microscope to observe cells in culture? |
|---|
| | Yes. This is a inverted microscope designed to look at sample in a dish or a multi-well plate. For long timelapse, be aware of photo-toxicity. |
|
|
| Tabs Container |
|---|
| id | Demo Image |
|---|
| title | Zeiss LSM-800 |
|---|
| direction | horizontal |
|---|
|
| Tabs Page |
|---|
| id | Demo Image |
|---|
| title | Demo Image |
|---|
|
Image Removed| Include Page |
|---|
| Plateformes:_Plat-foot-en |
|---|
| Plateformes:_Plat-foot-en |
|---|
|