| Tabs Page |
|---|
| DescriptionLight sourcesObjectives- 10x/0.45
- 20x/0.75
- 40x/0.6
- 60x/0.95
| Expand |
|---|
| title | Objective specifications |
|---|
| | Position | Name | Brand | Full name | Identifier | Working distance (mm) | Transmittance (% [nm]) | Techniques | Cover glass thickness (mm) |
|---|
| 1 | 10x/0.45 Air | Nikon | 10x/0.45 Air Plan Apo | MRD00101 | 4.0 |
| BF, p-PhC, p-DIC, Fluo | 0.17 |
|---|
| 2 | 20x/0.75 Air | Nikon | 20x/0.75 Air Plan Apo DIC N2 | MRD00201 | 1.0 |
| BF, p-PhC, p-DIC, Fluo | 0.17 |
|---|
| 3 | | Nikon | 40x/0.6 Plan Fluor ELWD | MRH08430 | 2.7-3.7 |
| BF, p-PhC, p-DIC, Fluo | 0.17 |
|---|
| 4 | 60x/0.95 Air | Nikon | 60x/0.95 Air Plan Apo | MRD00600 | 0.15 |
| BF, p-PhC, p-DIC, Fluo | 0.17 |
|---|
BF: Bright-field p-PhC: Pseudo-phase contrast p-DIC: Pseudo-Differetial interference contrast |
FiltersDAPI GFP Cy3 Cy5 - Cy5.5
| Expand |
|---|
| title | Complete specifications |
|---|
| | Position | Name | Brand | Identifier | Polychroic (Transmission) | Emission filter | Effective bandwith (nm) |
|---|
| 1 | DAPI | TBD | TBD | BP 420/20 [430-462] | 455/50 [430-480] | [430-480] |
|---|
| 2 | GFP | TBD | TBD | BP 446/32; 524/42; 600/36; 732/137 | 525/20 [515-535] | [515-535] |
|---|
| 3 | Cy3 | TBD | TBD | BP 446/32; 524/42; 600/36; 732/137 | 605/52 [579-631] | [582-618] |
|---|
| 4 | Cy5 | TBD | TBD | BP 446/32; 524/42; 600/36; 732/137 | 707/72 [671-743] | [671-743] |
|---|
| 5 | Cy5.5 | TBD | TBD | BP 446/32; 524/42; 600/36; 732/137 | 720/60 [690-750] | [690-750] |
|---|
|
Detector| Expand |
|---|
| title | Complete specifications |
|---|
| Camera | PCO Edge 5.5 |
|---|
Sensor Type | sCMOS |
|---|
Sensor Category
| Monochrome |
|---|
Nb Pixels
| 5.5 M |
|---|
Pixel Layout | 2560 x 2160 |
|---|
Pixel size | 6.5 um
|
|---|
Sensor size
| 16.6 mm x 14.0 mm |
|---|
Sensor diameter
| 21.8 mm
|
|---|
Bit depth
| 16-bit |
|---|
| Speed at full resolution | 100 images/s
|
|---|
Max QE
| 60 % at 600 nm |
|---|
| Reading noise | 1.0 e⁻ |
|---|
Cooling
| Forced air +7C |
|---|
Dark Current
| 0.6 e⁻/pixel/sec |
|---|
Full well capacity
| 30 000 e-
|
|---|
Dynamic Range
| 1:30000 |
|---|
Interface
| Dual Camera Link PCIe |
|---|
Mount
| C-mount |
|---|
|
|
| Tabs Page |
|---|
| User Guide| UI Expand |
|---|
| If not already done, turn on the microscope (#1) (the switch is not easily accessible and is located on the back of the right side of the device) If not already done, turn on the computer (#2) and log in to Windows using your UdeM credentials Start -up INCell Analyzer 6000
|
| UI Expand |
|---|
| title | Loading samplesUsage |
|---|
| During this procedure, you will: Once completed, your sample will be ready for acquisition. | linkText | | UI Expand |
|---|
| - Within the software, click Eject to open the access door
- Insert your sample in the space provided for this purpose, respecting the orientation indicated
Click Load to close the access doors | Note |
|---|
For use with the 60x objective it is absolutely necessary to use a multi-well plate with a glass bottom. The glass thickness should be 0.17mm. We recommend the following plates: |
|
| Link in New Window |
|---|
href | https://shop.gbo.com/en/row/products/bioscience/microplates/sensoplate-glass-bottom-plates/bs-sensoplate-plus/655891.htmllinkText | | Link in New Window | hrefhttps://www.thermofisher.com/order/catalog/product/164588 | use microscope instrument for the first time, it is necessary to define the type of plate used. | You will usually do this This procedure is usually carried out during the training session | .This procedure but can also be performed if you are using a different multi-well plate. During this procedure, you will: Create a new plate map Measure the A1 well offset Measure and compute the inter-well distance Measure the plate bottom height and thickness
| Tip |
|---|
If you need help, feel free to ask for support. We are always happy to help. |
| UI Expand |
|---|
| title | Create a new plate map |
|---|
| - In the
| INCell Analyzer software:Select - software, select from the
| menu - menu Application>Plate\Slide Manager...
- Click on the New Plate\Slide icon
- Select the number of wells in your plate (6, 24, 96 etc...)
- Adjust the following settings:
- Name: Your-Name_Your-Plate
- Plate height, bottom thickness and well volume are usually provided in the product technical sheet by the manufacturer
- The material used plastic or glass
- If necessary adjust the number of rows and columns as well as the shape of the well (round or rectangular)
- Click OK
- Close the Plate/Slide Manager window
|
| UI Expand |
|---|
| title | Measure A1 well offset |
|---|
| - In the
| INCell Analyzer :- , if necessary, click on the Dashboard > Objective
| Lens- Lens> 10x to select the 10x objective
- In the channel list, click the + button and add a transmitted light channel (brightfield)
- To the right of the plate layout, click the Setup Preview Imaging button
- Draw a rectangle around well A1
- Click on preview (on the acquisition panel at the top right of the screen)
- Wait for the preview to be generated
- On the plate map, click on the real center of well A1 to center this well on the objective
In the Dashboard menu select Plate/Slide Click Edit Plate then Define Upper Left Well Click Yes - Check that the yellow circle matches the edges of the well
Click Yes
|
| UI Expand |
|---|
| title | Compute the inter-well distance |
|---|
| - In the
| INCell Analyzer :- , if necessary, click on the Dashboard > Objective
| Lens- Lens> 10x to select the 10x objective
- In the channel list, click the + button and add a transmitted light channel (brightfield)
- To the right of the plate layout, click the Setup Preview Imaging button
- Draw a rectangle around the bottom right well
- Click on preview (on the acquisition panel at the top right of the screen)
- Wait for the preview to be generated
- On the plate map, click on the real center of
| well A1 to - bottom right Well to center this well on the objective
In the Dashboard menu select Plate/Slide Click Edit Plate then Define bottom right Well Click Yes - The inter-well distance will then be automatically calculated
- Check that the yellow circles matches the edges of the wells
Click Yes
|
| UI Expand |
|---|
| title | Measure the bottom height and thickness of the plate bottom |
|---|
|
In the INCell Analyzer software: CThis diagram explains how the Laser Autofocus measure and estimate the position of the sample. | View file |
|---|
| name | Plate_Diagram.pdf |
|---|
| height | 250 |
|---|
|
- In the software, if necessary, click on the Dashboard
| In Objective Lens - > Objective Lens> 10x to select the 10x objective
- On the plate
| plan- map, click on the center of a well containing a sample to center this well on the objective
- In the Dashboard menu select Plate/Slide
- Click Verify LAF
- If the 2 detected peaks (blue vertical lines) correspond to the measured peaks (black curve)
- Click Apply Measured Parameters
- Click OK
- Click Yes
- If the 2 detected peaks (blue vertical lines) do not correspond to the measured peaks (black curve)
- Change the bottom thickness value and repeat the operation until the peak are detected successfully
- Close the Laser Autofocus Plate/Slide Verification window
|
|
|
|---|
| UI Expand |
|---|
| - Files can be saved saved temporarily (during acquisition) to local C: drive (desktop)
- At the end of each session, copy your data to your external drive and delete it from local C: drive
- You can store your files on the D: drive (Data Storage). If you do, please create one folder per laboratory using the principal investigator's last name. Inside, create a folder per user using the following nomenclature (First Name_Last Name).
| Note |
|---|
| In any case, do not store your files on the C: drive. |
|
| UI Expand |
|---|
| - Click Eject to open the access door
- Collect your samples
- Click Load to close the access door
- Select the 10x objective
- Close the INCell Analyzer software
- Transfer your data to
disk - the D: drive (Data Storage) or to your external
hard - drive and delete it from the local
disk - C: drive
- Turn off
the - the computer
| Note |
|---|
| - Collect your samples, especially those in the microscope
- Leave the microscope and workspace clean
|
|
|
| Tabs Page |
|---|
id | Light path |
|---|
| The following diagrams allow you to follow the light path in transmitted light (bright field) and in reflected light (fluorescence). These diagrams will be available soon. | | View file |
|---|
| name | Plate_Diagram.pdf |
|---|
| height | 250 |
|---|
|
| Tabs Page |
|---|
| ManualsAvailable manuals |
| Tabs Page |
|---|
| Log
| UI Expand |
|---|
| title | To do | 2025-11-23 Migrated to windows 11 |
|---|
| Migrated to windows 11 |
| UI Expand |
|---|
| title | 2022-12 Liquid handling removal |
|---|
| Because of an issue with the liquid handling motor, the liquid handling arm as been replaced by a standard fixed arm for the transmitted light LED. Therefore liquid handling is no longer available. |
| UI Expand |
|---|
| title | 2021-10-26 Y motor not resting |
|---|
| Y motor not resting when reaching the transmitted light position: This time the issue is slightly different: Y motor is homing properly during the initialization procedure but instead of resting it keeps pushing toward the home position and after a brief instant the instrument stops and the red light shows up. If the liquid handler Y motor is disabled the initialization proceed properly and the system is usable at the exception of the brightfield of course |
| UI Expand |
|---|
| title | 2021-09-27 Added to Wiki |
|---|
| Added to wiki |
|
| Tabs Page |
|---|
| idtitle | Technical Datasheet | title |
|---|
| Technical DatasheetInstrument Serial W24318-1493815 BK02029 Service Password apiAWL Workstation |
| Tabs Page |
|---|
| Tabs Page |
|---|
id | FAQ |
|---|
| title | Troubleshooting & FAQ |
|---|
| Troubleshooting| UI Expand |
|---|
| title | The microscope does not show the usual green light, what should I do? |
|---|
| Usually, this microscope displays a green light when operational and an orange light when in use. A red light is on ON means the microscope is not functional. In this case please contact the platform manager. |
FAQ| UI Expand |
|---|
| title | Can I use this microscope to observe cells in culture? |
|---|
| | Yes. This microscope is designed for high-throughput acquisition of specimens in multi-well plates. The microscope can control temperature but does not have an environmental chamber.You can also acquire images of specimen mounted between a slide and a coverslip (thickness 0.17mm). |
| UI Expand |
|---|
| title | How to turn off the microscope? |
|---|
| Usually the microscope remains on. However, it is best to turn it off if not in use for a long time (weeks). To do this: - Navigate to the menu Application > Hardware >
- Click Shutdown Instrument
- Wait for the microscope to turn off
- The software closes automatically
|
|
|