Zeiss Axio-Imager Z2 fully motorized upright microscope

Roger Gaudry B-301


  • X-Cite Liquid light guide transmittance 75% remaining (37mW at 100% for GFP 488nm using 10x objective)
  • Fluorescence Light bulb 2593 hours should be replaced
  • Objective parafocality adjusted
  • Objective Focus speed was changed 10x: 7; 20x: 6; 4-x: 5; 63x: 3; 100x-1.3: 3; 100x-1.4: 3.
  • Condenser lens cleaned (oil)
  • Data storage OK
  • Added to wiki


Can I use this microscope to look at cell in a dish?

  • Yes. This is an inverted microscope designed to look at specimen in a dish or a multi-well plate
  • The objectives are optimized to image through thin glass bottom multi-well plates
  • You may also image specimen mounted between a slide and a 0.17mm thick coverslip



Core Facilities CIB               Structural Biology    |    Cytometry    |    Microscopy    |    Histology

  • Aucune étiquette