Versions Compared

Key

  • This line was added.
  • This line was removed.
  • Formatting was changed.

Include Page
_Micro-head-en
_Micro-head-en

Button Hyperlink
iconbuild
titleFrançais
typeprimary
urlZeiss Z1 Colibri
Lire en Francais


Zeiss

Axio

Z1-

Observer Z1

Colibri

inverted microscope

Microscope

Desmarais Building, Room 2234
Simple Microscope usage price
Instrument awarded to Dr. Audrey Claing and Dr. Jean-Philippe Gratton by the Canadian Foundation for Innovation (CFI) in 2014

Transmitted light

 

Applications

  • Inverted microscope
  • Widefield imaging
    • Brightfield
    • Phase contrast
    • Fluorescence
  • Incubation
  • Timelapse imaging
  • Deconvolution

Zeiss Z1 ColibriImage Removed

  • Extended Depth of Focus

Image Added

Image Added


Instrument Tabs
Tabs Container
Tabs Container
id
directionhorizontal
Tabs Page
idDescription
titleDescription

Description

Light sources

  • LED lamp for transmitted light

  • Colibri 7 R[G/Y]B-UV (385/469/555-590/631) for fluorescence
Expand
titleComplete specifications

Emission peak (nm)

Nominal Power (mW)Measured Power (mW)
385/30150 
469/38110 
555/3031 
631/3350 

View file
nameZeiss_Colibri
-7_Product Information.pdf
height250

Objectives

  1. 2.5x/0.075 Air

  2. 10x/0.25 Air Ph1
  3. 20x/0.5 Air Ph2
  4. 40x/0.75 Air Ph2

  5. 63x/0.75 Air Ph2 Long Distance
  6. 63x/1.4 Oil
Expand
titleComplete specifications
PositionNameBrandFull nameIDMagnificationNumerical ApertureImmersionTypeWorking distance (mm)Transmittance
(% [nm])
TechniqueCover glass thickness (mm)
12.5x/0.075 AirZeiss2.5x/0.075
EC Plan-Neofluar

M27
420320-9901-0002.5x0.075AirPlan Neofluar9.5>80% [400-840]BF, Fluo0.17
210x/0.25 AirZeiss10x/0.25 Ph1
N-Achroplan
M27 
420941-9911-00010x0.25AirN-AchroPlan

6.5

Not AvailableBF, PhC, Fluo0.17
320x/0.5 AirZeiss20x/0.50 Ph2
EC Plan-Neofluar
M27

420351-9910-000

20x0.50AirPlan Neofluar2.0Not AvailableBF, PhC, Fluo0.17
4

40x/0.75 Air

Zeiss40x/0.75 Ph2
EC Plan-Neofluar
M27

420361-9910-000

40x

0.75

Air

Plan Neofluar0.71Not AvailableBF, PhC, Fluo

0.17

563x/0.75 Air
Zeiss

63x/0.75 Corr Ph2
LD Plan-Neofluar


M27

421381-9970-000

63x

0.75

AirLD Plan-Neofluar1.7 at cover glass 0.75Not AvailableBF, PhC, Fluo

0 - 1.5

663x/1.4 OilZeiss63x/1.4 DIC
 Plan-Apochromat Oil
M27
420782-9900-00063x1.4OilPlan Apochromat0.19>80% [400-700]BF, Fluo0.17

Filter

  1. DAPI
  2. GFP
  3. Rhodamine
  4. DHE (dihydroethidium)
  5. Cy5
  6. Quadruple DAPI/GFP/Cy3/Cy5
Expand
titleComplete specifications
PositionNameBrandIDExcitation filterDichroic mirrorEmission filterComments
1DAPI
Filter Set 49
Zeiss

488049-9901

365/50
[325-375]
395LP445/50
[420-470]

2GFP
Filter Set 13
Zeiss488013-0000

470/20
[460-480]

495LP

517/25
[505-530]

3DsRed
Filter Set 43
Zeiss

000000-1114-101

545/25
[533-567]
570LP605/70
[570-640]

4DHECustom

Custom

500/50
[475-525]

540LP580/20
[570-590]
Undefined specifications
Best guess values
5Cy5
Filter Set 50
Zeiss488050-9901640/30
[625-655]
660LP

690/50
[665-715]


6Quadruple
DAPI/GFP/Cy3/Cy5
FS90 HE LED
Zeiss489090-9110-000


QBS
405+493+575+653

QBP 425/30+514/30+592/25+709/100

Excitation filters included in the light source FS90 HE LED

Detectors

  • PCO Edge 5.5Zeiss AxioCam MRm
    Expand
    titleComplete specifications
    Camera
Zeiss AxioCam MRm
  • PCO Edge 5.5

    Sensor Type

CCD
  • sCMOS

    Sensor Category

    Monochrome

    Nb Pixels

1
  • 5.
4
  • 5 M

    Pixel Layout

1388
  • 2560 x
1040
  • 2160

    Pixel size

     6.
45
  • 5 um

    Sensor size

8
  • 16.
9
  • 6 mm
x 6
  • x  14.
7
  • 0 mm

    Sensor diameter

11
  • 21.8 mm

    Bit depth

12
  • 16-bit
    Speed at full resolution
13
  • 100 images/s

    Max QE

 55
  •  60 % at 600 nm
Readout
  • Reading noise
 8
  •  1.0 e⁻

    Cooling

Pelletier
Forced air +7C

Dark Current

0.
7
  • 6 e⁻/pixel/sec

    Full well capacity

17
  • 30 000 e-

    Dynamic Range

    1:
2000
  • 30000

    Interface

FireWire (IEEE 1394a)
  • Dual Camera Link PCIe

    Mount

    C-mount

Image Removed

Zeiss_AxioCam MRm
  • Image Added

    View file
    namePCO_Edge 5.5_Datasheet.pdf

  • height250

  • PCO Edge 5.5
  • Zeiss Axiocam MRm (not installed)

Expand
titleComplete specifications
sCMOS55 2560 21605 16 mm x  140 21.8 16100
CameraPCO Edge 5.5Zeiss Axiocam MRm

Sensor Type

CCD

Sensor Category

Monochrome

Nb Pixels

1.4 M

Pixel Layout

1388 x 1040

Pixel size

 6.45 um

Sensor size

8.9 mm x 6.7 mm

Sensor diameter

11 mm

Bit depth

12-bit
Speed at full resolution13 images/s

Max QE

 60  55 % at 600 nm
Reading Readout noise 1.0  8 e⁻

Cooling

Forced air +7CPelletier

Dark Current

0.6 7 e⁻/pixel/sec

Full well capacity

30 17 000 e-

Dynamic Range

1:300002000

Interface

Dual Camera Link PCIe
FireWire (IEEE 1394a)

Mount

C-mount

Image Removed


Image Added

View file
nameZeiss_AxioCam MRm
PCO_Edge 5.5
_Datasheet.pdf
height250

Tabs Page
idUser Guide
titleUser Guide

User Guide

UI Expand
expandedtrue
titleStartup
  1. If not already done, turn on the computer (#1) and log in to Windows using your UdeM credentials
  2. Remove the dust cover from the microscope
  3. If incubation is required, turn on the incubation power bar (#2) on the desk near the computer and open the CO₂ tank cylinder (#2B) near the sink
    Note

    Make sure the humidifier is properly filled with distilled water.

  4. Turn on the microscope power bar (#3) on the desk near the computer

  5. When using the instrument for the first time, it is necessary to import the microscope configuration into before starting the software. See the the First Use section below.
  6. Start Zen
UI Expand
titleFirst Use

When using the instrument for the first time, it is necessary to import the microscope configuration into the software. This procedure is usually carried out during the training session. However, it can also be used to reset the software if it does not display correctly, for example. Running this procedure will erase all your experiment protocols and reset the software to its original settings (ask for support if you are not sure).

  1. If Zen is open, close it and wait until it has completely shut down (this may take up to 30 seconds)

  2. On the Desktop, open the Softwares folder

  3. Double-click Zen Settings for Axio-Observer Zeiss Z1-Colibri

  4. A script will run and a black window will appear briefly

  5. When the message Settings for Zen have been imported successfully appears, click OK to close it

  6. You can now open Zen

UI Expand
titleLoading samples

During this procedure, you will:

  • Set the microscope to a safe configuration
  • Perform a calibration
  • Load your sample
  • Find and adjust the focus

Once completed, your sample will be ready for acquisition.

UI Expand
titleCalibration

This step is required to calibrate the microscope in XY and Z. Performing this calibration will significantly reduce the time needed to locate and focus on your sample.

  • If not already done, select the lowest-magnification objective. On the microscope touch screen Home > Microscope > Control > Objectives > 2.5x
  • In Zen, once it has started a calibration dialog should appear. Simply click Calibrate Now.
    The microscope will lower the objectives, perform an XY calibration first, followed by a Z calibration, and then return to its original position.
Infotip

Once calibrated, the focus is typically found at Z = 1.1 mm for a microscope slide. For a multi-well plate the focus is generally around Z = 3 mm but this is highly dependent on your multi-well plate.
The Z position can be viewed on the microscope touch screen under Home > Z-Position, as well as in Zen within the Focus tab, located on the right side of the screen.

Expand
titleThe calibration dialog did not appear...

The system may already be calibrated. This can occur if a previous user calibrated the system and left it on.

UI Expand
titleVerifying if the system is already calibrated

In Zen, within the Focus tab, located on the right side of the screen:

  1. Click Load to lower the objective

  2. The Z position is indicated below Current
  3. If Z position value is less than 100 um the system is calibrated

On the microscope touch screen:

  1. Lower the objective by pressing Home > Load Position

  2. Press Set Work Position to store this position

  3. If necessary, slightly adjust the focus upward to clear the Lower Z Limit Reached message displayed on the touch screen

  4. The Z position value is indicated
  5. If the value is less than 0.1 mm then the system is calibrated
UI Expand
titleManual calibration with the software

In Zen, within the Stage tab, located on the right side of the screen:

  1. If not already done, check the Show All option
  2. At the bottom of the tab click Calibrate

  3. A warning message will show up click Continue

  4. The microscope will lower the objectives and perform a XY stage calibration and then return to its original position.

In Zen, within the Focus tab, located on the right side of the screen:

  1. If not already done, check the Show All option
  2. At the bottom of the tab click Calibrate

  3. A warning message will show up click Continue

  4. The microscope will then perform a Z calibration and then return to its original position

XY and Z calibrations are now complete

UI Expand
titleManual calibration with the microscope touch screen

On the microscope touch screen:

  1. Navigate to Home > Microscope > XYZ > Position > Z-Position > Set Zero > Auto to perform focus calibration

  2. Press OK to start the calibration procedure

  3. Wait a few seconds for the calibration to complete

  4. Lower the objective by pressing Home > Load Position

  5. Press Set Work Position to store this position

  6. If necessary, slightly adjust the focus upward to clear the Lower Z Limit Reached message displayed on the touch screen

  7. Navigate to Home > Microscope > XYZ > Position > XY-Position > Set Zero > Auto to perform a stage calibration

  8. Press OK to start the calibration procedure

  9. Wait a few seconds for the calibration to complete

XY and Z calibrations are now complete

UI Expand
titleEnsuring the calibration dialog is displayed at startup

In Zen:

  1. In the menu bar, navigate to Tools > Options

  2. Select Startup/Shutdown

  3. Under Stage/Focus Calibration, ensure Request Stage/Focus Calibration on Startup is checked

  4. Click OK to close the Options dialog

UI Expand
titleFirst Initial focus
Warning
titleImportant
Make sure to calibrate the focus before performing the first focusEnsure that the calibration has been completed beforehand. Calibration will significantly reduce the time required to locate and focus on your sample.

On the microscope touch screen:

  1. If not already done,
  2. press Home>Microscope>Turret>Objectives>5x to
  3. select the
  4. 5x objective
  5. lowest-magnification objective Home > Microscope > Control > Objectives > 2.5x
Info

The

5x objective is the safest because it has the longest

2.5× and 10× objectives are the safest to use due to their long working distance (

12mm

>6 mm). The sample will appear

perfectly

in sharp

long

focus well before the

lens

objective approaches it. It is recommended

to always first

to always focus

with

using the safest

lens

objectives first.

The

Since the objectives are parafocal, focusing with the safest objectives will facilitate locating the sample when switching to higher-magnification objectives.

  1. Lower the objective by pressing Home > Load Position

  2. Press Set Work Position to

  3. objective will then allow you to easily find your sample with another objective. The 10x objective is also safe because its working distance is 6.5 mm.If not done already, press Home>Load Position to lower the objective to the lowest position and press Set Work Position to
  4. store this position

  5. If necessary,

  6. move
  7. slightly adjust the focus

  8. slightly up
  9. upward to

  10. remove
  11. clear the Lower Z

  12. limit reached
  13. Limit Reached message displayed on the

  14. touchscreen
  15. touch screen

  16. Place the test slide on the microscope stage with the coverslip
  17. toward
  18. facing the objective
    Note
    titleImportant
  19. Always use the
  20. Using a test slide

  21. to perform the first focus
  22. will significantly reduce the time required to set up the instrument.

  23. If necessary,
  24. move
  25. adjust the stage
  26. so
  27. to ensure that the sample is
  28. centered on
  29. centred under the objective

On the computerIn Zen:

  1. Open Zen
  2. In
  3. the
  4. the Locate tab, select BF or the desired fluorescence (Fluo, DAPI, GFP,
  5. mPlum
  6. Cy3, Cy5) to activate the configuration
  7. Adjust the focus with the main dial while looking through the eyepieces until the image is perfectly sharp
  8. note
Info

Once calibrated, the focus

can be

is typically found at Z = 1.1

.7 mm). The Z value can be found

mm for a microscope slide. For a multi-well plate the focus is generally around Z = 3 mm but this is highly dependent on your multi-well plate.
The Z position can be viewed on the microscope touch screen

Home>Z

under Home > Z-Position, as well as in Zen within the Focus tab, located on the right side of the screen.

  1. In the Locate tab, select Off to turn off the
  2. In the Locate tab, select Off to turn off the
  3. illumination
UI Expand
titleSeconday focus
Warning
titleImportant

First Perform the initial focus with using the   safest objective before selecting another lens and continuing with secondary focusobjective before switching to higher-magnification objectives.

There are two (2) 40x objectives, make sure you select the Air 40x
UI Expand
titleFocusing with air objectives

After performing the first initial focus, on the microscope touch screen:

  1. Press Home>Microscope>Control>Objectives, press
  2. Press 10x, 20x
  3. or 40x
  4. 40x or 63x (0.75) to select the desired objective
Info
The 40x objective is the best Air objective because it has the
greatest number of optical corrections (Plan Apochromat) and the largest numerical
largest Numerical aperture (0.
95
75)
but has
together with a
smaller
wider field of view
.
The 20x/0.8 objective offers the best compromise between Resolution and Field of View
Note
(40x).

In Zen:

  1. In
  2. the
  3. the Locate tab, select BF or the desired fluorescence (Fluo, DAPI, GFP, Cy3,
  4. mPlum
  5. DsRed, Cy5) to activate the configuration
  6. Adjust the
  7. focus 
  8. focus with the precision dial while looking through the eyepieces until the image is perfectly sharp
  9. In the Locate tab,
  10. select
  11. select Off to turn off the illumination
  12. off

Your sample is ready for acquisition!

UI Expand
titleFocusing with oil objectives

After performing the first initial focus, on the microscope touch screen:

  1. Press Home>Microscope>Turret>Objectives
  2. Press 63x Oil, 40x Oil (1.4)to select the desired objective. The microscope will automatically lower the lower the stage so that the sample is becomes accessible.

    Info

    The 40x and 63x oil objectives provide the same spatial resolution because they have the same numerical aperture (1.4).
    The 40x oil objective offers a larger field of view and transmits light slightly better beyond 700nm.
    The 63x oil objective transmits light slightly better in the visible spectrum (440-710 nm) and has a better Strehl ratio (90%). It is particularly suited for super-resolution imaging, but its field of view is smaller.

    Note

    There are two (2) 40x objectives, make sure you select the 40x Oil

  3. Place a single drop of oil on the objective
  4. Press Done. The microscope will automatically return the objective to its original position

In Zen :

  1. Place a single drop of oil on the objective
  2. Press Done. The microscope will automatically return the objective to its original position

In Zen :

  1. In the Locate tab, select BF or the desired fluorescence (Fluo, DAPI, GFP, Cy3, DsRed, Cy5
  2. In the Locate tab, select BF or the desired fluorescence (DAPI, GFP, mPlum
  3. ) to activate the configuration
  4. Adjust the focus with the precision dial while looking through the eyepieces
  5. until the
  6. until the image is perfectly sharp
  7. In the Locate tab,
  8. select
  9. select Off to turn off the illumination
  10. off

Your sample is ready for acquisition!

This procedure puts the microscope in a safe configuration and performs a focus calibration. At the end of this procedure the microscope will be ready for acquisition.

UI Expand
titleLoading samples
UI Expand
titleFocus Calibration Z

On the microscope touch screen:

  1. Press Home>Load Position to lower the objectives to its lowest position
  2. Press Set Work Position to store this position
  3. If necessary, move the focus slightly up to remove the Lower Z limit reached message displayed on the touchscreen
  4. Press Home>Microscope>Control>Objectives>2.5x to select the 2.5x objective
  5. If asked, tap Done to remove the oil lens cleaning warning
  6. Press Home>Microscope>XYZ>Position>Z-Position>Set zero>Auto to perform focus calibration
  7. Press OK to start the focus calibration procedure
  8. Wait a few seconds for the calibration to be completed
Note

Once calibrated, the focus can be found at Z = 1.1 mm). The Z value can be found on the microscope touch screen Home>Z-Position

UI Expand
titleFirst focus
Warning
titleImportant

Make sure to calibrate the focus before performing the first focus.

On the microscope touch screen:

  1. Press Home>Microscope>Turret>Objectives
  2. Press 2.5x to select the 2.5x lens
    Info

    The 2.5x objective is the safest because it has the longest working distance (9 mm). The sample will appear perfectly sharp long before the lens approaches it. It is recommended to always first focus with the safest lens. The objectives are para-focal, focusing with the safest objective will then allow you to easily find your sample with another objective.

  3. Press Home>Load Position to lower the objectives to its lowest position
  4. Press Set Work Position to store this position
  5. If necessary, move the focus slightly up to remove the Lower Z limit reached message displayed on the touchscreen
  6. Place the test slide on the microscope stage with the coverslip toward the objective
    Note
    titleImportant

    Always use the test slide to perform the first focus.

  7. If necessary, move the stage so that the sample is centered on the objective

On the computer:

  • Open Zen
  • In the Locate tab, select BF or the desired fluorescence (DAPI, GFP, Rhodamine, etc…) to activate the configuration
  • Adjust the focus with the main dial while looking through the eyepieces until the image is perfectly sharp
    Note

    Once calibrated, the focus can be found at Z =  1.1mm). The Z value can be found on the microscope touch screen Home>Z-Position

  • In the Locate tab, select Off to turn off the illumination
  • UI Expand
    titleSecondary focus
    Warning
    titleImportant

    First focus with the safest lens before selecting another lens and continuing with secondary focus.

    UI Expand
    titleFocusing with air objectives

    After performing the first focus, on the microscope touch screen:

  • Press Home>Microscope>Control>Objectives
  • Press 20x or 40x or 63x to select the desired lens
    Info

    The 40x objective is the best Air objective because it has the largest numerical aperture (0.75).

  • Adjust the focus with the precision dial while looking through the eyepieces until the image is perfectly sharp
  • Your sample is ready for acquisition!
  • UI Expand
    titleFocusing with oil lenses

    After performing the first focus, on the microscope touch screen:

    1. Press Home>Microscope>Control>Objectives
    2. Press 63x Oil. The microscope will automatically lower the objective so that the sample is accessible

    3. Place a drop of oil on your sample
    4. Press Done. The microscope will automatically return the sample to its original position

    In Zen:

    1. In the Locate tab, select BF or the desired fluorescence (DAPI, GFP, Rhodamine, etc…) to activate the configuration
    2. Adjust the focus with the precision dial while looking through the eyepieces until the image is perfectly sharp
    3. In the Locate tab, select Off to turn the illumination off
    4. Your sample is ready for acquisition!
    UI Expand
    titleStorage management
    • Files can be saved temporarily (during acquisition) on the local C: drive (desktop)
    • At the end of each session, copy your data to your external drive and delete it from the local C: drive
    • You can store your files on the D: drive (Data Storage). If you do, please create a folder per laboratory using the principal investigator last name. Within, create one folder per user (Firstname_Lastname).
    Note

    In any case, your files should be removed from the C: drive.

    UI Expand
    titleShutdown
    1. Save your data
    2. Close Zen
    3. Transfer your data to the D: drive (Data Storage) or to your external drive and delete it from the local C: drive
    4. If used, turn off the incubation module power strip (#2A) and close the CO2 cylinder (#2B)

    5. Select the 2.5x objective and press load position to bring the objectives to the bottom position

    6. Turn off the microscope power bar (#3)

    7. Turn off the computer
    8. Cover the instrument with the protective dust cover
    Note
    titleImportant Reminders
    • Take back your samples including ones in the microscope
    • Leave the microscope and the working area clean
    Tabs Page
    idLightpath
    titleLightpath

    The following schematics depict the light path for transmitted (bright-field and Phase Contrast) and reflected (fluorescence) lights.

    View file
    nameZeiss_Z1-Colibri_LightPath.pdf
    height250

    Tabs Page
    idManuals
    titleManuals

    Available manuals

    Storage management
    • Files can be saved temporarily (during acquisition) on the local C: drive (desktop)
    • At the end of each session, copy your data to your external drive and delete it from the local C: drive
    • You can store your files on the D: drive (Data Storage). If you do, please create a folder per laboratory using the principal investigator last name. Within, create one folder per user (Firstname_Lastname).
    Note

    In any case, your files should be removed from the C: drive.

    UI Expand
    titleShutdown
    1. Save your data
    2. Close Zen
    3. Transfer your data to the D: drive (Data Storage) or to your external drive and delete it from the local C: drive
    4. If used, clean oil objectives with lens cleaner and paper
    5. Select the lowest magnification objective and press load position to place the objectives in a safe position
    6. If used, turn off the incubation module power strip (#2A) and close the CO2 cylinder (#2B)

    7. Turn off the microscope power bar (#3)

    8. Turn off the computer
    9. Cover the instrument with the protective dust cover
    Note
    titleReminder
    • Take back your samples including ones in the microscope
    • Leave the microscope and the working area clean

    Tabs Page
    titleLog
    Tabs Page
    idLog
    titleLog

    Log

    UI Expand
    titleTo be done
    • Measure Power output
    • Modify camera power supply cable
    UI Expand
    title2025-10-22 Windows 11 and more
    • Installation of Windows 11
    • Added CO2 connection to incubation module
    • Cable management
    • Replaced video card
    UI Expand
    title2025-09-01 Camera upgrade
    • Added compressed air connexion to the anti-vibration table
    UI Expand
    title2025-09-01 Camera upgrade
    • AxioCam replacement with PCO Edge 5.5
    UI Expand
    title2025-08-21 Wiki Updated
    • Updated wiki with complete specifications
    UI Expand
    title2024-02-22 Microscope Firmware update
    • Microscope Firmware update to add Colibri to the touchscreen
    • Zen 3.5 HotFix 10
    UI Expand
    title2022-05-09
    • Computer replacement
    • Added Colibri
    • Added startup procedure
    • Parafocality and paracentrality
    UI Expand
    title2022-03-17
    • Added complete description
    UI Expand
    title2021-09-27
    • Added to wiki
    Tabs Page
    title
    idtitleTechnical Datasheet

    Technical Datasheet

    Stand

    • Zeiss Axio-Observer Z1 inverted  inverted Serial: 3851001242 Part Number: 431007-9902-000
      System ID: 1024979772
    • Camera adapter Model 60N-C, 1", 1x, Model: 426114

    Light sources

    • Transmitted LED light
    • Colibri 7 R(G/Y)B-UV 423052-9730-000 Serial 5440000661

    Condenser

    • Manual condenser Product number: TBD, Serial: TBD
    • Lens NA 0.35 WD 70 mm Part Number: 424241

    • Filter turret 6 positions manual

      1. H

      2. Ph0
      3. Ph1

      4. Ph2

      5. DIC

      6. DUC

    Objectives

    1. 2.5x/0.075 Air 420320-9901-000

    2. 10x/0.25 Air Ph1 420941-9911-000

    3. 20x/0.5 Air Ph2 420351-9910-000

    4. 40x/0.75 Air Ph2 420361-9910-000

    5. 63x/0.75 Air Ph2 Long Distance 421381-9970-000

    6. 63x/1.4 Oil 420782-9900-000

    Stage

    • Motorized stage Marzhauser Sensotech, Part number 432903-9011-000, #14 07 132052; 90-76-200-0820
    • Remote control joystick
    • Inserts
      • Slide combo
      • 6-well plate
      • 35 mm dish
      • Multi-well plat

    Filters

    1. DAPI Filter Set 49 488049-9901
    2. GFP Filter Set 13 488013-0000
    3. Rhodamine Filter Set 43 000000-1114-101
    4. DHE (dihydroethidium) 424931
    5. Cy5 Filter Set 50 488050-9901
    6. Multiband FS90 HE LED 489090-9110-000

    Detector

    • Zeiss AxioCam MR R3 Model: r3.1 Part Number: 426509-9901-000. Serial: 1 22 12 5537

    Workstation

      1. Ph0
      2. Ph1

      3. Ph2

      4. DIC

      5. DUC

    Objectives

    1. 2.5x/0.075 Air 420320-9901-000

    2. 10x/0.25 Air Ph1 420941-9911-000

    3. 20x/0.5 Air Ph2 420351-9910-000

    4. 40x/0.75 Air Ph2 420361-9910-000

    5. 63x/0.75 Air Ph2 Long Distance 421381-9970-000

    6. 63x/1.4 Oil 420782-9900-000

    Stage

    • Motorized stage Marzhauser Scan IM 130x100-2mm 90-24-550-0000 Serial 16053038
      Stage Controller 90-76-024-1803 Serial 14 04 1 2040 SMC 2009 432929-9011-000
    • Remote control joystick 2-Axis 90-76-200-0820 Zeiss Article 432903-9011-000 Serial 1615142059
    • Inserts
      • Slide combo
      • Multi-well plate

    Filters

    1. DAPI Filter Set 49 488049-9901
    2. GFP Filter Set 13 488013-0000
    3. Rhodamine Filter Set 43 000000-1114-101
    4. DHE (dihydroethidium) 424931
    5. Cy5 Filter Set 50 488050-9901
    6. Multiband FS90 HE LED 489090-9110-000

    Detector

    • Zeiss AxioCam MRm Model: r3.1 Part Number: 426509-9901-000. Serial: 1 22 12 5537
    • PCO Edge 5.5 Model pco.API.Air.BX Serial 6000001404

    Workstation

    • Computer
      • Fujitsu Esprimo P920 E90+ Model MI5W Serial YLPS036541
    • Motherboard
      • Fujistsu D3222-A1 (1 PCIe Gen3 x16, 1 PCIe Gen 2 x16 (x4 lanes), 2x PCIe Gen2 x1
      • Chipset Intel Q87
      • BIOS UEFI compatible American Trend v4.6.5.4 R1.47.0 for D3222-A1x 2019-08-26 
    • Fujitsu Esprimo P920 E90+
    • Processor
        • Intel Core i5-4670 @ 3.4 GHz
      • RAM 32 GB DDR3 1600 MHz ECC (4
        • 24 GB DDR3 800 MHz (3 x 8 GB)
      • OS 500 GB Drive
        • 1 TB SSD 550 MB/s
        • Win 11 23H2
      • Storage
        • 2 TB HD
        Data Storage (2 x 1 TB spanned volume)
        • 110 MB/s
      • Video Card AMD Card
        • AMD FirePro V4900 1 GB
        DDR5 dedicated memory
        • DDR5 0.768 TFLOPS 1x PCIe gen2 x16
      • Monitor
        • LG Flatron E2711 27' 1920 x 1080
        • LG W2442 1920 x1080
      • Software
        • Zen Blue 3.
        5 SN=1121159628-524292 HASP=1798977001
        • 5
        • Serial 1121159628-524292
        • HASP 1798977001
        • Extended Focus
        • Measurements
        • Multi-Channel
        • Panorama
        • Software autofocus
        • Time Series
        • Z-stack 

      Incubation

      • Pecon stage top incubation

      Anti-vibration table

      • TMC Model 63 512 Serial 933240
      • Plain table top

      Consumables

      • CO2 Tank
      • N2 Tank
      • Oil
      • Lens Cleaner
      • Oil
      • Lens Cleaner

      Manuals

    Tabs Page
    Tabs Page
    idFAQ
    titleTroubleshooting & FAQ

    Troubleshooting

    UI Expand
    titleI don't see any fluorescence!

    Technical support is free and unlimited. Contact us !

    UI Expand
    titleI see a high background in fluorescence

    This happens in the following conditions at

    • At 2.5x using the DAPI or GFP channel
    • At 10x using the GFP channel
    • At 20x using the GFP channel

     To solve this you can use any of the following optionThe fluoresncece light source is a Colibri while the transmitted light is a LED. What happens is that the fluorescence illumination reflects and into the LED and give a high background. To solve this:

    • Tilt the transmitted light arm backwardorStop the light from entering the transmitted LED by using a cardboard
    • Close the manual shutter between the condenser and the transmitted light
    • Use a higher magnification objective

    FAQ

    UI Expand
    titleCan I use this microscope to look at cell in a dish?

    Yes. This is an inverted microscope designed to look at specimen in a dish or a multi-well plate

    The objectives are optimized to image through thin glass bottom multi-well plates

    You may also image specimen mounted between a slide and a 0.17mm thick coverslip

    For long timelapse, be aware of photo-toxicity.

    UI Expand
    titleCan I use this microscope to perform timelapse experiments?

    Yes, but... This microscope has an incubation module to maintain temperature, humidity and gas. Yet it does not have a Definite focus which can maintain

    focus throughout time. Therefore, it is possible to loose the focus over long period.

    focus throughout time. Therefore, it is possible to loose the focus over long period. You can always do a software autofocus at different time but be aware of photo-toxicity if you are using fluorescence.

    UI Expand
    titleWhat is Extended Depth of Focus?

    Extended Focus is a method to flatten a Z-stack by keeping only the relevant information.

    In Zen:

    • Select the Processing Tab
    • Select the Extended Depth of Focus Method
    • Click Apply
    • Wait until the processing is finished
    • Save the processed image

    Z Stack animationImage AddedMaximum Intensity projection of a Z-StackImage AddedExtended Depth of Focus projection of a Z-StackImage Added

    Include Page
    Plateformes:_Plat-foot-en
    Plateformes:_Plat-foot-en