If not already done, turn on the computer (#1) and log in to Windows using your UdeM credentials
Remove the dust cover from the microscope
Turn on the microscope power bar (#2) on the left of the computer
When using the instrument for the first time, it is necessary to import the microscope configurationbefore starting the software. See the First Use section below.
Start-up NIS-Elements
When using the instrument for the first time, it is necessary to import the microscope configuration into the software. This procedure is usually carried out during the training session. However, it can also be used to reset the software if it does not display correctly, for example.
Running this procedure will erase all your experiment protocols and reset the software to its original settings (ask for support if you are not sure).
If open, closeNIS-Elementsand wait until it is completely closed (up to 30 seconds)
On your Desktop open the Softwares folder
Open NIS Settings Utility
Click on theImporttab
Click onBrowse
Navigate to your C:\Users\Public\Documents
Select the fileNikon_Ti2-Fura_NIS Settings.bin
ClickSelect
Select all items
ClickImport
ClickOK
Close theNIS Settings Utility
You can now reopen NIS-Elements
During this procedure, you will:
Set the microscope in a safe configuration
Load your sample
Find and adjust the focus
Once completed, your sample will be ready for acquisition.
Save your data
Close NIS-Elementssoftware
Transfer your data to the D: drive (Data Storage) or to your external drive and delete it from the local C: drive
Turn off the computer
If the oil objective was used, clean it with lens cleaner and lens paper (not Kimwipes)
Wait until the Spectra has cooled down and turn off the microscope power bar (#2)
Cover the microscope
Important Reminders
Collect your samples, especially those in the microscope
Leave the microscope and workspace clean
Files can be saved temporarily (during acquisition) to local C: drive (desktop)
At the end of each session, copy your data to your external drive and delete it from local C: drive
You can store your files on the D: drive (Data Storage).If you do, please create one folder per laboratory using the principal investigator's last name.Inside, create a folder per user using the following nomenclature (First Name_Last Name).
Important
In any case, do not store your files on the C: drive.
When using the microscope for the first time, you need to import the microscope settings into the software. You will usually do this during the training session. This procedure can also be performed if something is not working properly and if you want to reset the software to its original settings.
This process will delete all experiment protocols and reset the software to the original settings for this specific microscope.
If open, close NIS-Elements and wait until it is completely closed (up to 30 seconds)
On your Desktop open the Softwares folder
Open NIS Settings Utility
Click on the Importtab
Click on Browse
Navigate to your Desktop
Select the file Nikon Ti2-Fura settings for NIS.bin