Zeiss Axioscan 7 Numériseur de lames
Pavillon Desmarais, Local 5308-2
Tarif d'utilisation Microscope Simple
Instrument en 2025
- Numériseur de lames
- Lumière transmise
- Champs clair
- Fluorescence
- Lumière transmise
Zeiss Axioscan 7
Light sources
- LED lamp for transmitted light
- Viluma 7 FR-R[G/Y]BV-UV for fluorescence
Channel | Emission peak (nm) | Nominal Power (mW) | Measured Power (mW) | Compatible Dyes |
|---|---|---|---|---|
| UV | 385/30 | DAPI, Hoechst 33342, Hoechst 33258, Alexa Fluor 350, Alexa Fluor 405, Indo-1, eBFP / BFP, eGFP (wt), True Blue | ||
| V | 423/44 | Pacific Blue, Lucifer Yellow, Alexa Fluor 433, eCFP, Cerulean | ||
| B | 469/38 | FM1-43, Cy2, eGFP, NBD, MitoTracker Green, Alexa Fluor 488, Fluo-4, BCECF, Calcein, DiO SNAFL, YO-Pro-1, Nissl, LysoSensor Green, mHoneydew, FITC / Fluorescein, Kaede (green / red), PerCP, YoYo-1, FuraRed | ||
| G | 555/30 | TRITC, 7-AAD, Cy3, tdTomato, Alexa Fluor 546, Alexa Fluor 555, DsRed, mOrange, TagRFP, SNARF, DyLight 549, Spectrum Orange | ||
| Y | 591/27 | MitoTracker RED FM/CMXRos, txRed, mCherry, mRFP1, Cy3.5, Rhodamine B, Alexa Fluor 568, Dylight 594, Alexa Fluor 594, Bodipy TR | ||
| R | 631/33 | Alexa Fluor 633, Alexa Fluor 647, Cy5, DRAQ5, ToTo-3, ATTO-655, MitoTracker DeepRed, APC, ATTO-647N | ||
| FR | 735/40 | Alexa Fluor 750, Alexa Fluor 790, Cy7, Cy7.5 |
Total output 1510 mW
Zeiss Viluma 7 Emission Filters
Comparison Relative normalized intensity between Viluma 7 and HXP light sources
The information above does not directly indicate the effective intensity provided by the instrument as Zeiss does not provide the nominal power of their light sources,excepted the total output being 1510 mW. The graph on the left only indicates the emission filter bandwidth while the graph on the right compared the normalized relative intensities between the Viluma and the historic HXP 120 mercury bulb lightsource. Finally both graphs are misleading since they present 8 individual channels which can't be provided within the same product. The Viluma 7 can only provide 7 individual channel. In this case, the model included in this instrument does not have the Cyan Channel.
Zeiss_Viluma_Product Info.pdf
Objectives
- 5x/0.25 Air
- 10x/0.45 Air
- 20x/0.8 Air
- 40x/0.95 Air
| Position | Name | Brand | Full name | ID | Magnification | Numerical Aperture | Immersion | Type | Working distance (mm) | Transmittance (% [nm]) | Technique | Cover glass thickness (mm) |
|---|---|---|---|---|---|---|---|---|---|---|---|---|
| 1 | 5x/0.25 Air | Zeiss | 5x/0.25 Fluar | 420130-9900-000 | 5x | 0.25 | Air | Fluar | 12.5 | >90% [410-800] | BF, Fluo | 0.17 |
| 2 | 10x/0.45 Air | Zeiss | 10x/0.45 Plan Apochromat | 420640-9900-000 | 10x | 0.45 | Air | Plan Apocrhomat | 2.0 | >80% [380-900] | BF, Fluo | 0.17 |
| 3 | 20x/0.8 Air | Zeiss | 20x/0.80 Plan Apochromat | 420650-9903-000 | 20x | 0.80 | Air | Plan Apochromat | 0.55 | >90% [400-800] | BF, Fluo | 0.17 |
| 4 | 40x/0.95 Air | Zeiss | 40x/0.95 Plan Apochromat | 420660-9970-000 | 40x | 0.95 | Air | Plan Apochromat | 0.25 | >90% [480-780] | BF, Fluo | 0.13 - 0.21 |
| 5 | Empty | |||||||||||
| 6 | Empty | |||||||||||
| 7 | Empty |
Dichroic
- Filter Set 112 LED SBP 405/495/575/654/761
| Position | Name | Brand | ID | Excitation filter | Dichroic mirror | Emission filter | Comments |
|---|---|---|---|---|---|---|---|
| 1 | Filter Set 112 LED SBP | Zeiss | - | 405/495/575/654/761 | - | ||
| 2 | |||||||
| 3 | |||||||
| 4 | |||||||
| 5 | |||||||
| 6 |
Emission filters
- Empty
- DAPI
- FITC
- DsRed
- Cy5
- Cy7
| Position | Name | Brand | ID | Excitation filter | Dichroic mirror | Emission filter | Comments |
|---|---|---|---|---|---|---|---|
| 1 | Empty | ||||||
| 2 | DAPI | Zeiss | 425/30 [412-438 | ||||
| 3 | FITC | Zeiss | 514/30 [501-527 | ||||
| 4 | DsRed | Zeiss | 592/25 [583-601 | ||||
| 5 | Cy5 | Zeiss | [681/45 | ||||
| 6 | Cy7 | Zeiss | [788/38 |
List of compatible dyes
Detectors
- Zeiss AxioCam 705 Color
| Camera | Zeiss AxioCam 705 Color |
|---|---|
Sensor Type | CMOS |
Sensor Category | Color |
Nb Pixels | 5.07 M |
Pixel Layout | 2464 × 2056 |
Pixel size | 3.45 µm × 3.45 µm |
Sensor size | 8.5 mm x 7.1 mm |
Sensor diameter | 11.1 mm |
Bit depth | 14-bit |
| Speed at full resolution | 60 images/s |
Max QE | 80 % @ 540nm |
| Readout noise | 2.2 e⁻ |
Cooling | 18C Thermoelectric |
Dark Current | 0.5 e⁻/pixel/sec |
Full well capacity | 11 000 e- |
Dynamic Range | 1:5000 1: 25 000 HDR mode |
Interface | USB B 3.0 |
Mount | C-mount |
- Hamamatsu ORCA-Flash4.0 V3
| Camera | Hamamatsu ORCA-Flash4.0 V3 |
|---|---|
Sensor Type | sCMOS |
Sensor Category | Monochrome |
Nb Pixels | 4.2 M |
Pixel Layout | 2048 x 2048 |
Pixel size | 6.5 um |
Sensor size | 13.312 mm x 13.312 mm |
Sensor diameter | 18.826 mm |
Bit depth | 16-bit |
| Speed at full resolution | 30 images/s |
Max QE | 82 % at 560 nm |
| Reading noise | 0.8 e⁻ |
Cooling | Air cooled -10C |
Dark Current | 0.06 e⁻/pixel/sec |
Full well capacity | 30 000 e- |
Dynamic Range | 1:37000 |
Interface | USB 3.0 or Camera-Link |
Mount | C-mount |
This instrument is only offered with service. Please contact us.
- If not already done, turn on the computer (#1) and log in to Windows using your UdeM credentials
- Turn on the instrument power bar (#2) on the desk between the instrument and the computer
- Turn on the instrument (#3) on the left side of the instrument
- Start Zen
- Files can be saved temporarily (during acquisition) on the local C: drive (desktop)
- At the end of each session, copy your data to your external drive and delete it from the local C: drive
- You can store your files on the D: drive (Data Storage). If you do, please create a folder per laboratory using the principal investigator last name. Within, create one folder per user (Firstname_Lastname).
In any case, your files should be removed from the C: drive.
- Save your data
- Close Zen
- Transfer your data to the D: drive (Data Storage) or to your external drive and delete it from the local C: drive
- Turn off the microscope power bar (#1)
- Turn off the computer
Reminder
- Take back your samples including ones in the microscope
- Leave the microscope and the working area clean
Troubleshooting
Technical support is free and unlimited. Contact us !
FAQ
No. This is a slide scanner dedicated to digitize size mounted between a slide and a 0.17mm thick coverslip.